Origin of a false positive urine pregnancy test in a patient with membranoproliferative glomerulonephritis type I

被引:10
作者
Marzinke, Mark A. [1 ]
Jarrar, Pamela [1 ]
Atkinson, Meredith [2 ]
Humphrey, Richard L. [1 ]
Detrick, Barbara [1 ]
Sokoll, Lori J. [1 ]
机构
[1] Johns Hopkins Med Inst, Dept Pathol, Baltimore, MD 21287 USA
[2] Johns Hopkins Med Inst, Dept Pediat, Baltimore, MD 21287 USA
关键词
False positive; Human chorionic gonadotropin; Interference; Pregnancy test; Urine; INTERFERENCES;
D O I
10.1016/j.cca.2011.10.011
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Discrepant qualitative urine and quantitative serum hCG results were observed in a patient with a history of membranoproliferative glomerulonephritis. Further studies were performed to investigate this discrepancy. Methods: Urine and serum specimens were analyzed using alternate methods and heterophilic antibody blocking and dilution studies were performed. The potentially interfering substance was identified and subsequent specimens from the patient's hospital stay were analyzed for hCG and the effect of the interferent was assessed. Results: Initially, the patient's urine specimen was positive for the presence of hCG based on the qualitative QuickVue + hCG point-of-care device. However, quantitative serum testing was negative for pregnancy on the Tosoh AIA-1800 analyzer (<2 mIU/ml). The serum specimen was also positive on the QuickVue + device, but upon incubating urine or diluted serum with heterophilic antibody binding reagent, hCG results shifted from positive to negative. Although the patient had a non-elevated human anti-mouse antibody (HAMA) level, the patient exhibited an elevated serum rheumatoid factor concentration (270 IU/ml; reference interval: <= 20 IU/ml). Application of a calibrator solution containing rheumatoid factor at levels above 175 IU/ml resulted in a positive reading on the QuickVue + point-of-care card. Conclusion: The discrepancy between the qualitative urine and quantitative hCG results was due to an interfering substance causing a false positive on the QuickVue + urine point-of-care device. Subsequent studies identified rheumatoid factor as the potential interferent in this case. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:361 / 363
页数:3
相关论文
共 12 条
[1]   Membranoproliferative glomerulonephritis [J].
Alchi, Bassam ;
Jayne, David .
PEDIATRIC NEPHROLOGY, 2010, 25 (08) :1409-1418
[2]   Pathological aspects of membranoproliferative glomerulonephritis (MPGN) and haemolytic uraemic syndrome (HUS) thrombocytic thrombopenic purpura (TTP) [J].
Benz, Kerstin ;
Amann, Kerstin .
THROMBOSIS AND HAEMOSTASIS, 2009, 101 (02) :265-270
[3]  
CARSON DA, 1987, ANNU REV IMMUNOL, V5, P109, DOI 10.1146/annurev.immunol.5.1.109
[4]   False-positive urine pregnancy test in a woman with adenomysosis [J].
Er, Tze Kiong ;
Chiang, Chein-Hua ;
Cheng, Bin Hong ;
Hong, Fu Jung ;
Lee, Ching Ping ;
Ruiz Gines, Miguel Angel .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2009, 27 (08) :1019-1019
[5]   Pregnancy tests with end-stage renal disease [J].
Fahy, Brenda G. ;
Gouzd, Valerie A. ;
Ataltah, Joseph N. .
JOURNAL OF CLINICAL ANESTHESIA, 2008, 20 (08) :609-613
[6]   Interferences in hormone immunoassays [J].
Klee, GG .
CLINICS IN LABORATORY MEDICINE, 2004, 24 (01) :1-+
[7]  
KOUNTZ DS, 1989, NEW ENGL J MED, V321, P1416
[8]  
Kricka LJ, 1999, CLIN CHEM, V45, P942
[9]   Membranoproliferative glomerulonephritis [J].
Nakopoulou, L .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2001, 16 :71-73
[10]   DOUBLING TIMES OF HUMAN CHORIONIC-GONADOTROPIN INCREASE IN EARLY VIABLE INTRAUTERINE PREGNANCIES [J].
PITTAWAY, DE ;
REISH, RL ;
WENTZ, AC .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1985, 152 (03) :299-302