Scatter correction for full-fan volumetric CT using a stationary beam blocker in a single full scan

被引:97
作者
Niu, Tianye
Zhu, Lei [1 ]
机构
[1] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Nucl & Radiol Engn Program, Atlanta, GA 30332 USA
关键词
scatter correction; single scan; stationary beam blocker; volumetric CT; cone-beam CT; X-RAY SCATTER; COMPUTED-TOMOGRAPHY; IMAGE-RECONSTRUCTION; PRIMARY MODULATION; DIAGNOSTIC-RADIOLOGY; CORRECTION ALGORITHM; SHADING CORRECTION; RADIATION-THERAPY; NOISE-REDUCTION; SIMULATION;
D O I
10.1118/1.3651619
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Applications of volumetric CT (VCT) are hampered by shading and streaking artifacts in the reconstructed images. These artifacts are mainly due to strong x-ray scatter signals accompanied with the large illumination area within one projection, which lead to CT number inaccuracy, image contrast loss and spatial nonuniformity. Although different scatter correction algorithms have been proposed in literature, a standard solution still remains unclear. Measurement-based methods use a beam blocker to acquire scatter samples. These techniques have unrivaled advantages over other existing algorithms in that they are simple and efficient, and achieve high scatter estimation accuracy without prior knowledge of the imaged object. Nevertheless, primary signal loss is inevitable in the scatter measurement, and multiple scans or moving the beam blocker during data acquisition are typically employed to compensate for the missing primary data. In this paper, we propose a new measurement-based scatter correction algorithm without primary compensation for full-fan VCT. An accurate reconstruction is obtained with one single-scan and a stationary x-ray beam blocker, two seemingly incompatible features which enable simple and efficient scatter correction without increase of scan time or patient dose. Methods: Based on the CT reconstruction theory, we distribute the blocked data over the projection area where primary signals are considered approximately redundant in a full scan, such that the CT image quality is not degraded even with primary loss. Scatter is then accurately estimated by interpolation and scatter-corrected CT images are obtained using an FDK-based reconstruction algorithm. Results: The proposed method is evaluated using two phantom studies on a tabletop CBCT system. On the Catphan (c) 600 phantom, our approach reduces the reconstruction error from 207 Hounsfield unit (HU) to 9 HU in the selected region of interest, and improves the image contrast by a factor of 2.0 in the high-contrast regions. On an anthropomorphic head phantom, the reconstruction error is reduced from 97 HU to 6 HU in the soft tissue region and image spatial nonuniformity decreases from 27% to 5% after correction. Conclusions: Our method inherits the main advantages of measurement-based methods while avoiding their shortcomings. It has the potential to become a practical scatter correction solution widely implementable on different VCT systems. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3651619]
引用
收藏
页码:6027 / 6038
页数:12
相关论文
共 43 条
[1]   Image reconstruction from the Compton scattering of X-ray fan beams in thick/dense objects [J].
Arsenault, Paul J. ;
Hussein, Esam M. A. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) :1622-1633
[2]   Cone beam X-ray scatter removal via image frequency modulation and filtering [J].
Bani-Hashemi, A ;
Blanz, E ;
Maltz, J ;
Hristov, D ;
Svatos, M .
MEDICAL PHYSICS, 2005, 32 (06) :2093-2093
[3]   AN ANALYTICAL MODEL OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :721-725
[4]   Accelerated simulation of cone beam X-ray scatter projections [J].
Colijn, AP ;
Beekman, FJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2004, 23 (05) :584-590
[5]   Implementation and characterization of a 320-slice volumetric CT scanner for simulation in radiation oncology [J].
Coolens, C. ;
Breen, S. ;
Purdie, T. G. ;
Owrangi, A. ;
Publicover, J. ;
Bartolac, S. ;
Jaffray, D. A. .
MEDICAL PHYSICS, 2009, 36 (11) :5120-5127
[6]   Effect of scattered radiation on image noise in cone beam CT [J].
Endo, M ;
Tsunoo, T ;
Nakamori, N ;
Yoshida, K .
MEDICAL PHYSICS, 2001, 28 (04) :469-474
[7]   Magnitude and effects of x-ray scatter in a 256-slice CT scanner [J].
Endo, Masahiro ;
Mori, Shinichiro ;
Tsunoo, Takanori ;
Miyazaki, Hiroaki .
MEDICAL PHYSICS, 2006, 33 (09) :3359-3368
[8]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[9]   Modulator design for x-ray scatter correction using primary modulation: Material selection [J].
Gao, Hewei ;
Zhu, Lei ;
Fahrig, Rebecca .
MEDICAL PHYSICS, 2010, 37 (08) :4029-4037
[10]   Scatter correction method for x-ray CT using primary modulation: Phantom studies [J].
Gao, Hewei ;
Fahrig, Rebecca ;
Bennett, N. Robert ;
Sun, Mingshan ;
Star-Lack, Josh ;
Zhu, Lei .
MEDICAL PHYSICS, 2010, 37 (02) :934-946