PPG-Based Systolic Blood Pressure Estimation Method Using PLS and Level-Crossing Feature

被引:37
作者
Fujita, Daisuke [1 ]
Suzuki, Arata [1 ]
Ryu, Kazuteru [2 ]
机构
[1] Wakayama Univ, Fac Syst Engn, Sakaedani 930, Wakayama 6408510, Japan
[2] Kanai Hosp, Fushimi Ku, 612-12 Yodokizu Cho, Kyoto, Japan
来源
APPLIED SCIENCES-BASEL | 2019年 / 9卷 / 02期
关键词
blood pressure estimation; photoplethysmography; partial least-squares regression; PHOTOPLETHYSMOGRAM;
D O I
10.3390/app9020304
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This paper proposes a cuff-less systolic blood pressure (SBP) estimation method using partial least-squares (PLS) regression. Level-crossing features (LCFs) were used in this method, which were extracted from the contour lines arbitrarily drawn on the second-derivative photoplethysmography waveform. Unlike conventional height ratio features (HRFs), which are extracted on the basis of the peaks in the waveform, LCFs can be reliably extracted even if there are missing peaks in the waveform. However, the features extracted from adjacent contour lines show similar trends; thus, there is a strong correlation between the features, which leads to multicollinearity when conventional multiple regression analysis (MRA) is used. Hence, we developed a multivariate estimation method based on PLS regression to address this issue and estimate the SBP on the basis of the LCFs. Two-hundred-and-sixty-five subjects (95 males and 170 females [(Mean +/- Standard Deviation) SBP: 133.1 +/- 18.4 mmHg; age: 62.8 +/- 16.8 years] participated in the experiments. Of the total number of subjects, 180 were considered as learning data, while 85 were considered as testing data. The values of the correlation coefficient between the measured and estimated values were found to be 0.78 for the proposed method (LCFs + PLS), 0.58 for comparison method 1 (HRFs + MRA), and 0.62 for comparison method 2 (HRFs + MRA). The proposed method was therefore found to demonstrate the highest accuracy among the three methods being compared.
引用
收藏
页数:11
相关论文
共 21 条
[1]   NEW LOOK AT STATISTICAL-MODEL IDENTIFICATION [J].
AKAIKE, H .
IEEE TRANSACTIONS ON AUTOMATIC CONTROL, 1974, AC19 (06) :716-723
[2]   Photoplethysmography and its application in clinical physiological measurement [J].
Allen, John .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) :R1-R39
[3]  
Baek H.J., 2007, P 6 INT SPEC TOP C I, P70
[4]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[5]  
Choudhury AD, 2014, IEEE ENG MED BIO, P4567, DOI 10.1109/EMBC.2014.6944640
[6]   Detection of c, d, and e waves in the acceleration photoplethysmogram [J].
Elgendi, Mohamed .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2014, 117 (02) :125-136
[7]   On the Analysis of Fingertip Photoplethysmogram Signals [J].
Elgendi, Mohamed .
CURRENT CARDIOLOGY REVIEWS, 2012, 8 (01) :14-25
[8]   Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio [J].
Gholamhosseini, Hamid ;
Baig, Mirza ;
Rastegar, Solmaz ;
Linden, Maria .
PROCEEDINGS OF THE 15TH INTERNATIONAL CONFERENCE ON WEARABLE MICRO AND NANO TECHNOLOGIES FOR PERSONALIZED HEALTH (PHEALTH 2018), 2018, 249 :77-83
[9]   Pulse wave velocity and the second derivative of the finger photoplethysmogram in treated hypertensive patients: their relationship and associating factors [J].
Hashimoto, J ;
Chonan, K ;
Aoki, Y ;
Nishimura, T ;
Ohkubo, T ;
Hozawa, A ;
Suzuki, M ;
Matsubara, M ;
Michimata, M ;
Araki, T ;
Imai, Y .
JOURNAL OF HYPERTENSION, 2002, 20 (12) :2415-2422
[10]   Multiclass MTS for Saxophone Timbre Quality Inspection Using Waveform-shape-based Features [J].
Hsiao, Yu-Hsiang ;
Su, Chao-Ton .
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS PART B-CYBERNETICS, 2009, 39 (03) :690-704