Novel Method of Measuring Corneal Viscoelasticity Using the Corvis ST Tonometer

被引:5
作者
Boszczyk, Agnieszka [1 ]
Kasprzak, Henryk [1 ]
Przezdziecka-Dolyk, Joanna [1 ,2 ]
机构
[1] Wroclaw Univ Sci & Technol, Dept Opt & Photon, PL-50370 Wroclaw, Poland
[2] Wroclaw Med Univ, Dept Ophthalmol, PL-50556 Wroclaw, Poland
关键词
non-contact tonometry; Corvis ST; corneal applanation; corneal buckling; corneal viscoelasticity; BIOMECHANICAL PROPERTIES; INTRAOCULAR-PRESSURE;
D O I
10.3390/jcm11010261
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The process of rapid propagation of the corneal deformation in air puff tonometer depends not only on intraocular pressure, but also on the biomechanical properties of the cornea and anterior eye. One of the biomechanical properties of the cornea is viscoelasticity, which is the most visible in its high-speed deformations. It seems reasonable to link the corneal viscoelasticity parameter to two moments of the highest speed of corneal deformations, when the cornea buckles. The aim of this work is to present a method of determining the time and place of occurrence of corneal buckling, examine spatial and temporal dependencies between two corneal applanations and bucklings in the Corvis ST tonometer, and correlate these dependencies with corneal viscoelastic properties. Methods: Images of the horizontal cross section of the Corvis ST deformed cornea from the air puff tonometer Corvis ST were used. 14 volunteers participated in the study, each of them had one eye measured eight times. Mutual changes in the profile slopes of the deformed corneas were numerically determined. They describe pure corneal deformation, eliminating the influence of rotation, and displacement of the entire eyeball. For each point in the central area of the corneal profile, the maximum velocities of mutual slope changes accompanying the applanations were estimated. The times of their occurrence were adopted as buckling times. Results: The propagation of buckling along the corneal profile is presented, as well as the repeatability and mutual correlations between the buckling parameters and intraocular pressure. Based on the relationship between them, a new parameter describing corneal hysteresis: Corvis Viscoelasticity (CVE) is introduced. It is characterized by high repeatability: ICC = 0.82 (0.69-0.93 CI) and low and insignificant correlation with intraocular pressure: r = 0.25 (p-value = 0.38). Conclusion: The results show for the first time how to measure the corneal buckling and viscoelastic effects with Corvis ST. CVE is a new proposed biomechanical parameter related to the viscoelastic properties of the cornea, which has high repeatability for the examined subject. The distribution of its values is planned to be tested on different groups of patients in order to investigate its clinical applicability.
引用
收藏
页数:16
相关论文
共 31 条
[1]  
Abass A, 2020, J REFRACT SURG, V36, P346, DOI 10.3928/1081597X-20200212-04
[2]  
[Anonymous], 2016, US GUID CORV ST G 72
[3]   Localization in spherical shell buckling [J].
Audoly, Basile ;
Hutchinson, John W. .
JOURNAL OF THE MECHANICS AND PHYSICS OF SOLIDS, 2020, 136
[4]   Clinical Evaluation of Methods to Correct Intraocular Pressure Measurements by the Goldmann Applanation Tonometer, Ocular Response Analyzer, and Corvis ST Tonometer for the Effects of Corneal Stiffness Parameters [J].
Bao, FangJun ;
Huang, ZiXu ;
Huang, JinHai ;
Wang, JunJie ;
Deng, ManLi ;
Li, LinNa ;
Yu, AYong ;
Wang, QinMei ;
Elsheikh, Ahmed .
JOURNAL OF GLAUCOMA, 2016, 25 (06) :510-519
[5]  
Bland JM, 1996, BRIT MED J, V313, P744
[6]   CALCULATING CORRELATION-COEFFICIENTS WITH REPEATED OBSERVATIONS .2. CORRELATION BETWEEN SUBJECTS [J].
BLAND, JM ;
ALTMAN, DG .
BRITISH MEDICAL JOURNAL, 1995, 310 (6980) :633-633
[7]  
Cicchetti DV., 1994, PSYCHOL ASSESSMENTS, V6, P284, DOI [DOI 10.1037/1040-3590.6.4.284, 10.1037/1040-3590.6.4.284]
[8]   Biomechanics and wound healing in the cornea [J].
Dupps, William J., Jr. ;
Wilson, Steven E. .
EXPERIMENTAL EYE RESEARCH, 2006, 83 (04) :709-720
[9]   Determination of Corneal Biomechanical Behavior in-vivo for Healthy Eyes Using CorVis ST Tonometry: Stress-Strain Index [J].
Eliasy, Ashkan ;
Chen, Kai-Jung ;
Vinciguerra, Riccardo ;
Lopes, Bernardo T. ;
Abass, Ahmed ;
Vinciguerra, Paolo ;
Ambrosio, Renato, Jr. ;
Roberts, Cynthia J. ;
Elsheikh, Ahmed .
FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2019, 7 (MAY)
[10]   Ex-vivo experimental validation of biomechanically-corrected intraocular pressure measurements on human eyes using the CorVis ST [J].
Eliasy, Ashkan ;
Chen, Kai-Jung ;
Vinciguerra, Riccardo ;
Maklad, Osama ;
Vinciguerra, Paolo ;
Ambrosio Jr, Renato ;
Roberts, Cynthia J. ;
Elsheikh, Ahmed .
EXPERIMENTAL EYE RESEARCH, 2018, 175 :98-102