Regularity of corneal topography after penetrating keratoplasty - Nonmechanical trephination with the excimer laser 193 nm vs mechanical trephination

被引:50
|
作者
Langenbucher, A
Seitz, B
Kus, MM
Vilchis, E
Naumann, GOH
机构
关键词
penetrating keratoplasty; corneal topography; astigmatism; excimer laser trephination; motor trephination;
D O I
10.1055/s-2008-1035263
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background ''Vertical tilt'' of the graft caused by incongruent cut angles, ''horizontal torsion'' caused by a asymmetric suturing and decentration oi: the wound seem to be the most important reasons for high and/or irregular postoperative astigmatism and deterioration of visual outcome after penetrating keratoplasty. We studied the time course of qual il alive and quantitative topographic criteria after nonmechanical and mechanical trephination in correlation with visual outcome. Patients and methods Fifty patients each (30 keratconus, 20 Fuchs' dystrophies) underwent nonmechanical trephination (excimer laser MEL60, Aesculap-Meditec, Heroldsberg) and mechanical motor-trephination in penetrating keratoplasty. All procedures (7.5 mm trephination diameter in Fuchs, 8.0 mm in keratoconus, double-running 10-0 nylon suture) were done by one surgeon. Pre-, intra- and postoperative treatment were identical. At the follow-up examinations, the keratometric astigmatism, qualitative and quantitative criteria of the automatic videokeratography, visual acuity and refraction were assessed. In addition, vector-corrected astigmatism (Jaffe-model) was evaluated in the postoperative time course, particularly after suture removal Results After a mean follow-up of 1.1 +/- 0.8 years, keratometric net astigmatism was 4.0 and 4.2 diopters after nonmechanical and mechanical trephination. Corneal topography analysis showed a higher orthogonality of the bow-tie shape and less asymmetry between opposite hemimeridians. Vector-corrected changes of corneal astigmatism were less after complete suture removal in nonmechanical trephination. After nonmechanical trephination. visual acuity was 2 decimal lines better. Discussion Nonmechanical trephination with the excimer laser 193 nm has the potential to improve the visual outcome by creating smooth and congruent cut edges parallel to the optical axis and the application of ''orientation teeth''. Since corneal net astigmatism does not differ significantly, a higher regularity of corneal topography after excimer laser trephination is supposed to be a main reason for this finding.
引用
收藏
页码:450 / 458
页数:9
相关论文
共 18 条
  • [1] Interpretation of corneal topography with wave-front parameters after penetrating keratoplasty - Comparison between nonmechanical trephination with the excimer laser and motor trephination
    Langenbucher, A
    Seitz, B
    Kus, MM
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1998, 212 (06) : 433 - 443
  • [2] Vertical tilt after penetrating keratoplasty - Comparison between nonmechanical trephination with the excimer laser and motor trephination
    Langenbucher, A
    Seitz, B
    Kus, MM
    Naumann, GOH
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1998, 212 (03) : 129 - 140
  • [3] 193 NM EXCIMER-LASER TREPHINATION IN PENETRATING KERATOPLASTY
    NAUMANN, GOH
    SEITZ, B
    LANG, GK
    LANGENBUCHER, A
    KUS, MM
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1993, 203 (04) : 252 - 261
  • [4] Experimental 193-nm excimer laser trephination with divergent cut angles in penetrating keratoplasty
    Seitz, B
    Behrens, A
    Langenbucher, A
    Kus, MM
    Naumann, GOH
    CORNEA, 1998, 17 (04) : 410 - 416
  • [5] Donor decentration of mechanical and excimer laser trephination for penetrating keratoplasty
    Seitz, B
    Langenbucher, A
    Meiller, R
    Kus, MM
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2000, 217 (03) : 144 - 151
  • [6] Immunologic graft reactions after nonmechanical corneal trephination using the excimer laser
    Seitz, B
    Langenbucher, A
    Kus, MM
    Kuchle, M
    Naumann, GOH
    OPHTHALMOLOGE, 1998, 95 (09): : 607 - 618
  • [7] Corneal wound healing after experimental penetrating keratoplasty after excimer laser trephination in free form
    Schmitz, K.
    Lang, G. K.
    Behrens-Baumann, W.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2006, 223 (12) : 957 - 965
  • [8] Immunological graft reactions after penetrating keratoplasty -: A prospective randomized trial comparing corneal excimer laser and motor trephination
    Seitz, B
    Langenbucher, A
    Diamantis, A
    Cursiefen, C
    Küchle, M
    Naumann, GOH
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2001, 218 (11) : 710 - 719
  • [9] Comparison of Excimer Laser Versus Femtosecond Laser Assisted Trephination in Penetrating Keratoplasty: A Retrospective Study
    Gábor Tóth
    Nóra Szentmáry
    Achim Langenbucher
    Elina Akhmedova
    Moatasem El-Husseiny
    Berthold Seitz
    Advances in Therapy, 2019, 36 : 3471 - 3482
  • [10] Design of computer-generated hologram with ring focus for nonmechanical corneal trephination with Er:YAG laser in penetrating keratoplasty
    Langenbucher, A
    Seitz, B
    Kus, MM
    van der Heyd, G
    Köchle, M
    Naumann, GOH
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 1999, 43 (06) : 453 - 457