Selective differential photoacoustic method for trace gas analysis

被引:29
作者
Uotila, J [1 ]
Koskinen, V [1 ]
Kauppinen, J [1 ]
机构
[1] Turku Univ, Dept Phys, Lab Opt & Spect, FIN-20014 Turku, Finland
关键词
photoacoustics; selectivity; gas detection; black body radiation; infrared;
D O I
10.1016/j.vibspec.2005.02.002
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A selective differential photoacoustic method for trace gas measurements is introduced. The method combines long path absorption spectroscopy with photoacoustic spectroscopy. A photoacoustic cell, equipped with a highly sensitive cantilever pressure sensor, was used as a selective detector for broadband IR-radiation passed through the sample gas. Therefore, an additional optical filter is not required. The detection limit at level of 10 ppb for methane was obtained (0.37 s measurement time and approximately I m effective absorption path). The sensitivity of the conventional photoacoustic method was enhanced by factor 2.8. The selectivity was tested by measuring the cross-interference signal between methane, carbon dioxide and butane in different conditions. The differential method enabled perfect separation of methane and carbon dioxide. Butane caused only one fifth of the cross-interference signal with methane compared to a cross-interference measured with a conventional filter system. (c) 2005 Published by Elsevier B.V.
引用
收藏
页码:3 / 9
页数:7
相关论文
共 16 条
[1]   CO-LASER PHOTOACOUSTIC-SPECTROSCOPY OF GASES AND VAPORS FOR TRACE GAS-ANALYSIS [J].
BERNEGGER, S ;
SIGRIST, MW .
INFRARED PHYSICS, 1990, 30 (05) :375-429
[2]  
Harren F.J., 2000, ENCY ANAL CHEM, V3, P2203, DOI [10.1002/9780470027318.A0718, DOI 10.1002/9780470027318.A0718]
[3]   High sensitivity in gas analysis with photoacoustic detection [J].
Kauppinen, J ;
Wilcken, K ;
Kauppinen, I ;
Koskinen, V .
MICROCHEMICAL JOURNAL, 2004, 76 (1-2) :151-159
[4]  
KAUPPINEN JK, 1997, SPECTROMETERS INFRAR, V19, P265
[5]   Resonant optoacoustic detector in nondispersive gas analyzer scheme [J].
Kovalyov, AA ;
Klebleyev, NR .
INFRARED PHYSICS & TECHNOLOGY, 1997, 38 (07) :415-421
[6]   Partial pressure measurement by means of infrared laser absorption spectroscopy [J].
Lanzinger, E ;
Jousten, K ;
Kuhne, M .
VACUUM, 1998, 51 (01) :47-51
[7]   ATMOSPHERIC-POLLUTION MONITORING USING CO2-LASER PHOTOACOUSTIC-SPECTROSCOPY AND OTHER TECHNIQUES [J].
MEYER, PL ;
SIGRIST, MW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (07) :1779-1807
[8]  
Michaelian K.H., 2003, PHOTOACOUSTIC INFRAR, V159
[9]   Application of acoustic resonators in photoacoustic trace gas analysis and metrology [J].
Miklós, A ;
Hess, P ;
Bozóki, Z .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (04) :1937-1955
[10]  
MURTZ M, 2004, ENVIRON SCI POLLUT R, V4, P61