Comparison of performance of high-performance liquid chromatography columns packed with superficially and fully porous 2.5 μm particles using kinetic plots

被引:29
作者
Fanigliulo, Ameriga [2 ]
Cabooter, Deirdre [1 ]
Bellazzi, Guido [2 ]
Tramarin, Davide [3 ]
Allieri, Brigida [2 ]
Rottigni, Alessandro [2 ]
Desmet, Gert [1 ]
机构
[1] Vrije Univ Brussel, Dept Chem Engn, B-1050 Brussels, Belgium
[2] GlaxoSmithKline R&D, Pharmaceut Dev, Verona, Italy
[3] GlaxoSmithKline R&D, Chem Dev, Verona, Italy
关键词
Kinetic plot method; Pharmaceutical analysis; Porous shell HPLC columns; SIZE DISTRIBUTION; DIFFUSION-COEFFICIENTS; SILICA MICROSPHERES; IONIZABLE COMPOUNDS; SHELL PARTICLES; PEAK-CAPACITY; FUSED-CORE; FAST-HPLC; PACKINGS; SEPARATION;
D O I
10.1002/jssc.201000463
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A recently introduced 2.5 mu m fully porous support (Kromasil Eternity) is compared with three different brands of superficially porous material (Kinetex, Halo and Poroshell 120) by means of the kinetic plot method using pharmaceutical compounds from Glaxo-SmithKline as probe molecules. The kinetic plot method immediately shows the range of plate numbers wherein a support performs better than another. Results from experiments carried out at pH 4.5 and 8.0 are presented in order to assess the pH stability of the tested phases. Moreover, since all supports are able to withstand pressures higher than 400 bar, they have been evaluated both on HPLC and UHPLC instrumentation. True average particle sizes were determined by SEM images taken from loose stationary phases. Kinetex outperforms the other columns in HPLC conditions for practically relevant efficiencies, but shows poor packing quality in the 100 x 2.1-mm format. Kromasil is advantageous for simple and fast separations on short columns both in HPLC and in UHPLC conditions. Halo achieves the highest efficiencies of all columns at the lowest pressure cost and shows a noticeable lower axial diffusion. Poroshell 120 has the best packing quality reproducibility across the tested formats. All columns preserve their performance at high pH.
引用
收藏
页码:3655 / 3665
页数:11
相关论文
共 48 条
[1]   Relation between the particle size distribution and the kinetic performance of packed columns -: Application to a commercial sub-2 μm particle material [J].
Billen, Jeroen ;
Guillarme, Davy ;
Rudaz, Serge ;
Veuthey, Jean-Luc ;
Ritchie, Harald ;
Grady, Brian ;
Desmet, Gert .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1161 (1-2) :224-233
[2]   Fused-core, sub-2 μm packings, and monolithic HPLC columns: a comparative evaluation [J].
Brice, Richard W. ;
Zhang, Xin ;
Colon, Luis A. .
JOURNAL OF SEPARATION SCIENCE, 2009, 32 (15-16) :2723-2731
[3]   Relationship between the particle size distribution of commercial fully porous and superficially porous high-performance liquid chromatography column packings and their chromatographic performance [J].
Cabooter, D. ;
Fanigliulo, A. ;
Bellazzi, G. ;
Allieri, B. ;
Rottigni, A. ;
Desmet, G. .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (45) :7074-7081
[4]   Kinetic plot and particle size distribution analysis to discuss the performance limits of sub-2 μm and supra-2 μm particle columns [J].
Cabooter, Deirdre ;
Billen, Jeroen ;
Terryn, Herman ;
Lynen, Frederic ;
Sandra, Pat ;
Desmet, Gert .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1204 (01) :1-10
[5]   Method to predict and compare the influence of the particle size on the isocratic peak capacity of high-performance liquid chromatography columns [J].
Cabooter, Deirdre ;
de Villiers, Andre ;
Clicq, David ;
Szucs, Roman ;
Sandra, Pat ;
Desmet, Gert .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1147 (02) :183-191
[6]   Kinetic plot method as a tool to design coupled column systems producing 100,000 theoretical plates in the shortest possible time [J].
Cabooter, Deirdre ;
Lestremau, Francois ;
Lynen, Frederic ;
Sandra, Pat ;
Desmet, Gert .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1212 (1-2) :23-34
[7]   The role of UHPLC in pharmaceutical development [J].
Chesnut, Stephen M. ;
Salisbury, John J. .
JOURNAL OF SEPARATION SCIENCE, 2007, 30 (08) :1183-1190
[8]   Fused-core particle technology as an alternative to sub-2-αm particles to achieve high separation efficiency with low backpressure [J].
Cunliffe, Jennifer M. ;
Maloney, Todd D. .
JOURNAL OF SEPARATION SCIENCE, 2007, 30 (18) :3104-3109
[9]   A new approach to the determination of column overload characteristics in reversed-phase liquid chromatography [J].
Dai, Jun ;
Carr, Peter W. ;
McCalley, David V. .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (12) :2474-2482
[10]   Analysis of basic compounds by reversed-phase high-performance liquid chromatography using hybrid inorganic/organic phases at high pH [J].
Davies, Nicola H. ;
Euerby, Melvin R. ;
McCalley, David V. .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1178 (1-2) :71-78