Field-Based Stable Isotope Analysis of Carbon Dioxide by Mid-Infrared Laser Spectroscopy for Carbon Capture and Storage Monitoring

被引:46
|
作者
van Geldern, Robert [1 ]
Nowak, Martin E. [1 ,3 ]
Zimmer, Martin [2 ]
Szizybalski, Alexandra [2 ]
Myrttinen, Anssi [1 ]
Barth, Johannes A. C. [1 ]
Jost, Hans-Juerg [4 ]
机构
[1] Univ Erlangen Nurnberg, Dept Geog & Geosci, GeoZentrum Nordbayern, D-91054 Erlangen, Germany
[2] GFZ German Res Ctr Geosci, D-14473 Potsdam, Germany
[3] Max Planck Inst Biogeochem, D-07745 Jena, Germany
[4] Thermo Fisher Sci, D-28199 Bremen, Germany
关键词
DISSOLVED INORGANIC CARBON; KETZIN PILOT SITE; RATIO INFRARED-SPECTROSCOPY; CO2; STORAGE; PIPELINE LEAKS; WATER; DELTA-C-13; DELTA-O-18; OXYGEN; CALIBRATION;
D O I
10.1021/ac5031732
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A newly developed isotope ratio laser spectrometer for CO2 analyses has been tested during a tracer experiment at the Ketzin pilot site (northern Germany) for CO2 storage. For the experiment, 500 tons of CO2 from a natural CO2 reservoir was injected in supercritical state into the reservoir. The carbon stable isotope value (delta C-13) of injected CO2 was significantly different from background values. In order to observe the breakthrough of the isotope tracer continuously, the new instruments were connected to a stainless steel riser tube that was installed in an observation well. The laser instrument is based on tunable laser direct absorption in the mid-infrared. The instrument recorded a continuous 10 day carbon stable isotope data set with 30 min resolution directly on-site in a field-based laboratory container during a tracer experiment. To test the instruments performance and accuracy the monitoring campaign was accompanied by daily CO2 sampling for laboratory analyses with isotope ratio mass spectrometry (IRMS). The carbon stable isotope ratios measured by conventional IRMS technique and by the new mid-infrared laser spectrometer agree remarkably well within analytical precision. This proves the capability of the new mid-infrared direct absorption technique to measure high precision and accurate real-time stable isotope data directly in the field. The laser spectroscopy data revealed for the first time a prior to this experiment unknown, intensive dynamic with fast changing delta C-13 values. The arrival pattern of the tracer suggest that the observed fluctuations were probably caused by migration along separate and distinct preferential flow paths between injection well and observation well. The short-term variances as observed in this study might have been missed during previous works that applied laboratory-based IRMS analysis. The new technique could contribute to a better tracing of the migration of the underground CO2 plume and help to ensure the long-term integrity of the reservoir.
引用
收藏
页码:12191 / 12198
页数:8
相关论文
共 50 条
  • [1] Mid-infrared polarization spectroscopy of carbon dioxide
    S. Roy
    R.P. Lucht
    A. McIlroy
    Applied Physics B, 2002, 75 : 875 - 882
  • [2] Mid-infrared polarization spectroscopy of carbon dioxide
    Roy, S
    Lucht, RP
    Mcilroy, A
    APPLIED PHYSICS B-LASERS AND OPTICS, 2002, 75 (08): : 875 - 882
  • [3] Monitoring of carbon dioxide exhaust emissions using mid-infrared spectroscopy
    Mulrooney, Jim
    Clifford, John
    Fitzpatrick, Colin
    Chambers, Paul
    Lewis, Elfed
    JOURNAL OF OPTICS A-PURE AND APPLIED OPTICS, 2007, 9 (06): : S87 - S91
  • [4] The Study of Atmospheric Carbon Isotope with Laser Absorption Spectroscopy at the Mid-Infrared Wavelength
    Xia Hua
    Dong Feng-zhong
    Han Luo
    Wu Bian
    Sun Peng-shuai
    Zhang Zhi-rong
    Cui Xiao-juan
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37 (11) : 3365 - 3369
  • [5] Monitoring Microbial Mineralization Using Reverse Stable Isotope Labeling Analysis by Mid-Infrared Laser Spectroscopy
    Dong, Xiyang
    Jochmann, Maik A.
    Elsner, Martin
    Meyer, Armin H.
    Baecker, Leonard E.
    Rahmatullah, Mona
    Schunk, Daniel
    Lens, Guido
    Meckenstock, Rainer U.
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2017, 51 (20) : 11876 - 11883
  • [6] A mid-infrared exhaled carbon dioxide isotope detection system based on 4.35 μm quantum cascade laser
    Li, Guolin
    Zhang, Xuena
    Zhang, Zecheng
    Wu, Yunhui
    Ma, Kun
    Jiao, Yue
    Li, Jiarui
    Liu, Yajing
    Song, Yimeng
    Zhao, Hao
    Zhai, Shenqiang
    Li, Qiang
    OPTICS AND LASER TECHNOLOGY, 2022, 152
  • [7] Carbon Dioxide Mid-Infrared Laser Spectroscopy with A Circular Multi-reflection (CMR) Cell
    Thuc Phan
    Dat Tran
    Esper, Jaime
    Nixon, Conor
    Nehmetallah, George
    CHEMICAL, BIOLOGICAL, RADIOLOGICAL, NUCLEAR, AND EXPLOSIVES (CBRNE) SENSING XXIII, 2022, 12116
  • [8] Compound specific stable carbon isotope analysis of aromatic organic contaminants in water using gas chromatography coupled to mid-infrared laser spectroscopy
    Zhang, Jiyun
    Shi, Zhe
    Zhu, Di
    Wang, Qiang
    Zhang, Gan
    Jin, Biao
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (06) : 1186 - 1192
  • [9] Carbon dioxide absorption spectroscopy with a mid-infrared silicon photonic waveguide
    Ottonello-Briano, Floria
    Errando-Herranz, Carlos
    Rodjegard, Henrik
    Martin, Hans
    Sohlstrom, Hans
    Gylfason, Kristinn B.
    OPTICS LETTERS, 2020, 45 (01) : 109 - 112
  • [10] Applying reverse stable isotope labeling analysis by mid-infrared laser spectroscopy to monitor BDOC in recycled wastewater
    Schulte, S. M.
    Koester, D.
    Jochmann, M. A.
    Meckenstock, R. U.
    SCIENCE OF THE TOTAL ENVIRONMENT, 2019, 665 : 1064 - 1072