Contrast sensitivity and higher order aberrations in eyes implanted with AcrySof IQ SN60WF and AcrySof SN60AT interocular lenses

被引:39
作者
Awwad, Shady T. [1 ,2 ]
Warmerdam, Dorotheea [1 ]
Bowman, R. Wayne [1 ]
Dwarakanathan, Surendar [1 ]
Cavanagh, H. Dwight [1 ]
McCulley, James P. [1 ]
机构
[1] Univ Texas SW Med Ctr Dallas, Dept Ophthalmol, Dallas, TX 75390 USA
[2] Amer Univ Beirut, Med Ctr, Dept Ophthalmol, Beirut, Lebanon
关键词
D O I
10.3928/1081597X-20080601-12
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the efficacy of the aspheric optic of the AcrySof IQ SN60WF intraocular lens (IOL) in decreasing spherical and total higher order aberrations and improving contrast sensitivity after cataract surgery compared to the spheric AcrySof SN60AT. METHODS: Fifty-two eyes of 36 patients were evaluated at the University of Texas Southwestern Medical Center at Dallas 3 to 6 months after phacoemulsification with implantation of the AcrySof IQ SN60WF (27 eyes) or AcrySof SN60AT (25 eyes) IOL. Eyes with previous surgery or ocular pathology were excluded. RESULTS: ayes implanted with the AcrySof IQ SN60WF had less total higher order and spherical aberrations than AcrySof SN60AT-implanted eyes at pupils of 4 mm (0.14 +/- 60.06 vs 0.18 +/- 0.04 mu m and -0.01 +/- 0.03 vs 0.09 +/- 0.03 mu m, respectively), 5 mm (0.24 +/- 0.07 vs 0.34 +/- 0.08 mu m and 0.03 +/- 0.02 vs 0.23 +/- 0.08 mu m, respectively), and 6 mm (0.42 +/- 0.13 vs 0.67 +/- 0.12 mu m and 0.08 +/- 0.04 and 0.46 +/- 0.09 mu m respectively), (P<.05 for all), with similar mean coma, trefoil, and 5th order aberrations, AcrySof SN60WF eyes showed better contrast sensitivity than AcrySof SN60AT eyes at 12 and 18 cycles per degree (cpd) under mesopic conditions (1.65 +/- 0.23 vs 1.51 +/- 0.25 and 1.22 +/- 0.21 vs 1.01 +/- 0.19, respectively), and at 18 cpd under glare (1.12 +/- 0.30 vs 0.87 +/- 0.31, respectively) (P<.05). Photopic contrast sensitivity was similar in both groups. The two groups had comparable mean age, axial length, IOL power, and postoperative mesopic and photopic pupil sizes. CONCLUSIONS: Compared to a spheric optic, the aspheric design of the AcrySof IQ SN60WF reduces spherical aberration, especially under larger pupil sizes, and improves mesopic contrast sensitivity at higher frequencies with and without glare.
引用
收藏
页码:619 / 625
页数:7
相关论文
共 36 条
  • [1] Optical performance of 3 intraocular lens designs in the presence of decentration
    Altmann, GE
    Nichamin, LD
    Lane, SS
    Pepose, JS
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (03) : 574 - 585
  • [2] Age-related changes in corneal and ocular higher-order wavefront aberrations
    Amano, S
    Amano, Y
    Yamagami, S
    Miyai, T
    Miyata, K
    Samejima, T
    Oshika, T
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 137 (06) : 988 - 992
  • [3] Applegate RA, 2000, J REFRACT SURG, V16, P507
  • [4] Contribution of the cornea and internal surfaces to the change of ocular aberrations with age
    Artal, P
    Berrio, E
    Guirao, A
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2002, 19 (01): : 137 - 143
  • [5] EFFECTS OF AGING IN RETINAL IMAGE QUALITY
    ARTAL, P
    FERRO, M
    MIRANDA, I
    NAVARRO, R
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1993, 10 (07): : 1656 - 1662
  • [6] Bellucci R, 2004, J REFRACT SURG, V20, P297
  • [7] Visual acuity and contrast sensitivity comparison between Tecnis and AcrySof SA60AT intraocular lenses: A multicenter randomized study
    Bellucci, R
    Scialdone, A
    Buratto, L
    Morselli, S
    Chierego, C
    Criscuoli, A
    Moretti, G
    Piers, P
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04) : 712 - 717
  • [8] Glare disability and spherical aberration with five foldable intraocular lenses: a prospective randomized study
    Casprini, F
    Balestrazzi, A
    Tosi, GM
    Miracco, F
    Martone, G
    Cevenini, G
    Caporossi, A
    [J]. ACTA OPHTHALMOLOGICA SCANDINAVICA, 2005, 83 (01): : 20 - 25
  • [9] Wavefront analysis in post-LASIK eyes and its correlation with visual symptoms, refraction, and topography
    Chalita, MR
    Chavala, S
    Xu, M
    Krueger, RR
    [J]. OPHTHALMOLOGY, 2004, 111 (03) : 447 - 453
  • [10] Chalita MR, 2003, J REFRACT SURG, V19, pS682