A phantom study on the positioning accuracy of the Novalis Body system

被引:90
作者
Yan, H [1 ]
Yin, FF [1 ]
Kim, JH [1 ]
机构
[1] Henry Ford Hosp, Dept Radiat Oncol, Detroit, MI 48202 USA
关键词
image-guided therapy; patient positioning; Novalis Body system; extracranial radiosurgery;
D O I
10.1118/1.1626122
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A phantom study was conducted to investigate inherent positioning accuracy of an image-guided patient positioning system-the Novalis Body system for three-dimensional (3-D) conformal radiotherapy. This positioning system consists of two infrared (IR) cameras and one video camera and two kV x-ray imaging devices. The initial patient setup was guided by the IR camera system and the target localization was accomplished using the kV x-ray imaging system. In this study, the IR marker shift and phantom rotation were simulated and their effects on the positioning accuracy were examined by a Rando phantom. The effects of CT slice thickness and treatment sites on the positioning accuracy were tested. In addition, the internal target shift was simulated and its effect on the positioning accuracy was examined by a water tank. With the application of the Novalis Body system, the positioning error of the planned isocenter was significantly reduced. The experimental results with the simulated IR marker shifts indicated that the positioning errors of the planned isocenter were 0.6 +/- 0.3, 0.5 +/- 0.2, and 0.7 +/- 0.2 mm along the lateral, longitudinal, and vertical axes, respectively. The experimental results with the simulated phantom rotations indicated that the positioning errors of the planned isocenter were 0.6 +/- 0.3, 0.7 +/- 0.2, and 0.5 +/- 0.2 mm along the three axes, respectively. The experimental results with the simulated target shifts indicated that the positioning errors of the planned isocenter were 0.6 +/- 0.3, 0.7 +/- 0.2, and 0.5 +/- 0.2 mm along the three axes, respectively. On average, the positioning accuracy of I mm for the planned isocenter was achieved using the Novalis Body system. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:3052 / 3060
页数:9
相关论文
共 13 条
[1]   Three-dimensional ultrasound image guidance for high-precision extracranial radiosurgery and radiotherapy [J].
Bouchet, LG ;
Meeks, SL ;
Bova, FJ ;
Buatti, JM ;
Friedman, WA .
RADIOSURGERY, VOL 4, 2002, 4 :262-278
[2]   The university of Florida frameless high-precision stereotactic radiotherapy system [J].
Bova, FJ ;
Buatti, JM ;
Friedman, WA ;
Mendenhall, WM ;
Yang, CC ;
Liu, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :875-882
[3]   Ultrasound-based stereotactic guidance of precision conformal external beam radiation therapy in clinically localized prostate cancer [J].
Lattanzi, J ;
McNeeley, S ;
Hanlon, A ;
Schultheiss, TE ;
Hanks, GE .
UROLOGY, 2000, 55 (01) :73-78
[4]   Daily prostate targeting using implanted radiopaque markers [J].
Litzenberg, D ;
Dawson, LA ;
Sandler, H ;
Sanda, MG ;
McShan, D ;
Ten Haken, RK ;
Lam, KL ;
Brock, KK ;
Balter, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (03) :699-703
[5]   Fiducial-based targeting accuracy for external-beam radiotherapy [J].
Murphy, MJ .
MEDICAL PHYSICS, 2002, 29 (03) :334-344
[6]   Commissioning an image-guided localization system for radiotherapy [J].
Phillips, MH ;
Singer, K ;
Miller, E ;
Stelzer, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (01) :267-276
[7]   Initial clinical experience with frameless stereotactic radiosurgery: Analysis of accuracy and feasibility [J].
Ryken, TC ;
Meeks, SL ;
Pennington, EC ;
Hitchon, P ;
Traynelis, V ;
Mayr, NA ;
Bova, FJ ;
Friedman, WA ;
Buatti, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :1152-1158
[8]   Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancers [J].
Shimizu, S ;
Shirato, H ;
Kitamura, K ;
Shinohara, N ;
Harabayashi, T ;
Tsukamoto, T ;
Koyanagi, T ;
Miyasaka, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (05) :1591-1597
[9]   Clinical use of stereoscopic X-ray positioning of patients treated with conformal radiotherapy for prostate cancer [J].
Soete, G ;
Verellen, D ;
Michielsen, D ;
Vinh-Hung, V ;
Van de Steene, J ;
Van den Berge, D ;
De Roover, P ;
Keuppens, F ;
Storme, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (03) :948-952
[10]  
THEVENAZ P, 1996, P IEEE INT C IM PROC, V1, P265