Underestimation of low-dose radiation in treatment planning of intensity-modulated radiotherapy

被引:37
作者
Jang, Si Young [1 ]
Liu, H. Helen [1 ]
Mohan, Radhe [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 94, Houston, TX 77030 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2008年 / 71卷 / 05期
关键词
IMRT dosimetry; underestimation of low dose; pulmonary complications; Monte Carlo simulation; thoracic cancers;
D O I
10.1016/j.ijrobp.2008.04.014
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate potential dose calculation errors in the low-dose regions and identify causes of such errors for intensity-modulated radiotherapy (IMRT). Methods and Materials: The IMRT treatment plans of 23 patients with lung cancer and mesothelioma were reviewed. Of these patients, 15 had severe pulmonary complications after radiotherapy. Two commercial treatment-planning systems (TPSs) and a Monte Carlo system were used to calculate and compare dose distributions and dose-volume parameters of the target volumes and critical structures. The effect of tissue heterogeneity, multi-leaf collimator (MLC) modeling, beam modeling, and other factors that could contribute to the differences in IMRT dose calculations were analyzed. Results: In the commercial TPS-generated IMRT plans, dose calculation errors primarily occurred in the low-dose regions of IMRT plans (<50% of the radiation dose prescribed for the tumor). Although errors in the dose-volume histograms of the normal lung were small (<5%) above 10 Gy, underestimation of dose <10 Gy was found to be up to 25% in patients with mesothelioma or large target volumes. These errors were found to be caused by inadequate modeling of MLC transmission and leaf scatter in commercial TPSs. The degree of low-dose errors depends on the target volumes and the degree of intensity modulation. Conclusions: Secondary radiation from MLCs contributes a significant portion of low dose in IMRT plans. Dose underestimation could occur in conventional IMRT dose calculations if such low-dose radiation is not properly accounted for. (C) 2008 Elsevier Inc.
引用
收藏
页码:1537 / 1546
页数:10
相关论文
共 21 条
[1]   Fatal pneumonitis associated with intensity-modulated radiation therapy for mesothelioma [J].
Allen, AM ;
Czerminska, M ;
Jänne, PA ;
Sugarbaker, DJ ;
Bueno, R ;
Harris, JR ;
Court, L ;
Baldini, EH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (03) :640-645
[2]   IMRT head and neck treatment planning with a commercially available Monte Carlo based planning system [J].
Boudreau, C ;
Heath, E ;
Seuntjens, J ;
Ballivy, O ;
Parker, W .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05) :879-890
[3]   Feasibility of using intensity-modulated radiotherapy to improve lung sparing in treatment planning for distal esophageal cancer [J].
Chandra, A ;
Guerrero, TM ;
Liu, HH ;
Tucker, SL ;
Liao, ZX ;
Wang, XC ;
Murshed, H ;
Bonnen, MD ;
Garg, AK ;
Stevens, CW ;
Chang, JY ;
Jeter, MD ;
Mohan, R ;
Cox, JD ;
Komaki, R .
RADIOTHERAPY AND ONCOLOGY, 2005, 77 (03) :247-253
[4]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[5]   Variations in energy spectra and water-to-material stopping-power ratios in three-dimensional conformal and intensity-modulated photon fields [J].
Jang, Si Young ;
Liu, H. Helen ;
Mohan, Radhe ;
Siebers, Jeffrey V. .
MEDICAL PHYSICS, 2007, 34 (04) :1388-1397
[6]   Dosimetric verification for intensity-modulated radiotherapy of thoracic cancers using experimental and Monte Carlo approaches [J].
Jang, Si Young ;
Liu, H. Helen ;
Wang, Xiaochun ;
Vassiliev, Oleg N. ;
Siebers, Jeffrey V. ;
Dong, Lei ;
Mohan, Radhe .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03) :939-948
[7]   Development and commissioning of a multileaf collimator model in Monte Carlo dose calculations for intensity-modulated radiation therapy [J].
Jang, SY ;
Vassiliev, ON ;
Liu, HH ;
Mohan, R ;
Siebers, JV .
MEDICAL PHYSICS, 2006, 33 (03) :770-781
[8]   A Monte Carlo study of radiation transport through multileaf collimators [J].
Kim, JO ;
Siebers, JV ;
Keall, PJ ;
Arnfield, MR ;
Mohan, R .
MEDICAL PHYSICS, 2001, 28 (12) :2497-2506
[9]   Fatal pneumonitis associated with intensity-modulated radiation therapy for mesothelioma:: In regard to Allen et al. (Int J Radiat Oncol Biol Phys 2006;65:640-645) [J].
Komaki, Ritsuko ;
Liao, Zhongxing ;
Liu, Helen ;
Tucker, Susan ;
Rice, David .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (05) :1595-1596
[10]   Monte Carlo verification of IMRT dose distributions from a commercial treatment planning optimization system [J].
Ma, CM ;
Pawlicki, T ;
Jiang, SB ;
Li, JS ;
Deng, J ;
Mok, E ;
Kapur, A ;
Xing, L ;
Ma, L ;
Boyer, AL .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (09) :2483-2495