Analytical evaluation of the new Beckman Coulter Access high sensitivity cardiac troponin I immunoassay

被引:5
|
作者
Lippi, Giuseppe [3 ]
Ferrari, Anna [1 ]
Gandini, Giorgio [2 ]
Gelati, Matteo [1 ]
Lo Cascio, Claudia [1 ]
Salvagno, Gian Luca [1 ]
机构
[1] Univ Verona, Sect Clin Biochem, Verona, Italy
[2] Univ Hosp Verona, Serv Transfus Med, Verona, Italy
[3] Univ Hosp Verona, Sect Clin Biochem, Ple Scuro 10, I-37314 Verona, Italy
关键词
immunoassay; method evaluation; myocardial infarction; troponin; ACUTE CORONARY SYNDROME; 99TH PERCENTILE; ASSAY; GENERATION; IMPRECISION; PERFORMANCE; ELEVATION; COHORT;
D O I
10.1515/cclm-2017-0350
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: This study was aimed to evaluate the analytical performance of the novel chemiluminescent and fully-automated Beckman Coulter Access hsTnI high-sensitivity immunoassay for measurement of cardiac troponin I (cTnI). Methods: The study, using lithium heparin samples, included assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, linearity, imprecision (within run, between-run and total), calculation of 99th percentile upper reference limit (URL) in 175 healthy blood donors (mean age, 36 +/- 12 years; 47% women) and comparison with two other commercial cTnI immunoassays. Results: The LOB, LOD and functional sensitivity of Access hsTnI were 0.14, 0.34 and 1.35 ng/L, respectively. The within-run, between-run and total imprecision was 2.2%-2.9%, 4.6%-5.4%, and 5.4%-6.1%, respectively. The linearity was excellent in the range of cTnI values between 0.95 and 4195 ng/L (r = 1.00). The 99th percentile URL was 15.8 ng/L. Measurable cTnI values were found in 173/175 healthy subjects (98.9%). Good agreement of cTnI values was found with AccuTnI+3 (r = 0.97; mean bias, -9.3%), whereas less satisfactory agreement was found with Siemens Dimension Vista cTnI (r = 0.95; mean bias, -55%). Conclusions: The results of our evaluation of the Beckman Coulter Access hsTnI indicate that the analytical performance of this fully-automated immunoassay is excellent.
引用
收藏
页码:157 / 161
页数:5
相关论文
共 50 条
  • [1] Analytical Evaluation of the New Beckman Coulter Access Procalcitonin (PCT) Chemiluminescent Immunoassay
    Lippi, Giuseppe
    Salvagno, Gian Luca
    Gelati, Matteo
    Pucci, Mairi
    Demonte, Davide
    Faggian, Diego
    Plebani, Mario
    DIAGNOSTICS, 2020, 10 (03)
  • [2] CytogeneticsComparability of Beckman Coulter Access Cardiac Troponin I Assays
    Stanek, Caroline
    Martin, Deirdra
    Lima, Jose
    Cao, Liyun
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2023, 160 : S140 - S141
  • [3] CytogeneticsComparability of Beckman Coulter Access Cardiac Troponin I Assays
    Stanek, Caroline
    Martin, Deirdra
    Lima, Jose
    Cao, Liyun
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2023, 160 : S140 - S141
  • [4] Evaluation of analytical performance of a new high-sensitivity immunoassay for cardiac troponin I
    Masotti, Silvia
    Prontera, Concetta
    Musetti, Veronica
    Storti, Simona
    Ndreu, Rudina
    Zucchelli, Gian Carlo
    Passino, Claudio
    Clerico, Aldo
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (03) : 492 - 501
  • [5] Analytical performance of Beckman Coulter's Access® AccuTnI™ (troponin I) in a multicenter evaluation.
    Uettwiller-Geiger, D
    Wu, AH
    Apple, FS
    Venge, P
    Jevans, AW
    Darte, C
    Roberts, S
    Yu, Y
    Olson, M
    CLINICAL CHEMISTRY, 2001, 47 (06) : A204 - A205
  • [6] Analytical performance characteristics of the new Beckman Coulter access PCT immunoassay
    Paulsen, K.
    Ley, K.
    Demirdjian, G.
    Cicirelli, E.
    Vigdorovich, I.
    Patzlaff, J.
    Ghimire, E.
    CLINICA CHIMICA ACTA, 2019, 493 : S553 - S554
  • [7] Analytical evaluation of the access AccuTnI cardiac troponin I immunoassay.
    Bainbridge, KL
    Collinson, PO
    CLINICAL CHEMISTRY, 2003, 49 (06) : A32 - A32
  • [8] Comparability of Beckman Coulter Cardiac Troponin I Assays
    Stanek, Caroline G.
    Lima, Jose
    Cao, Liyun
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2024, 54 (03): : 408 - 412
  • [9] Suspected myocardial infarction in the emergency department: An evaluation of clinical thresholds for the Beckman Coulter Access hsTnI high-sensitivity cardiac troponin I assay
    Meek, Robert
    Cullen, Louise
    Lu, Zhong
    Nasis, Arthur
    Kuhn, Lisa
    Sorace, Laurence
    EMERGENCY MEDICINE AUSTRALASIA, 2023, 35 (06) : 1005 - 1012
  • [10] High-sensitivity cardiac troponin I (HSTNI) immunoassay: Will the new Siemens Atellica® IM analyzer perform better than Beckman Coulter DXI 800?
    Yong, C. H.
    Tan, S. Y.
    Pajarillaga, A.
    Sahabdeen, N.
    Khoo, C.
    Wang, W. X.
    Saw, S.
    Sethi, S.
    CLINICA CHIMICA ACTA, 2022, 530 : S92 - S92