Comparison of Magnetic Resonance Imaging of Aortic Valve Stenosis and Aortic Root to Multimodality Imaging for Selection of Transcatheter Aortic Valve Implantation Candidates

被引:39
作者
Paelinck, Bernard P. [1 ]
Van Herck, Paul L. [1 ]
Rodrigus, Inez [2 ]
Claeys, Marc J. [1 ]
Laborde, Jean-Claude [4 ]
Parizel, Paul M. [3 ]
Vrints, Christiaan J. [1 ]
Bosmans, Johan M. [1 ]
机构
[1] Univ Antwerp Hosp, Dept Cardiol, Antwerp, Belgium
[2] Univ Antwerp Hosp, Dept Cardiac Surg, Antwerp, Belgium
[3] Univ Antwerp Hosp, Dept Radiol, Antwerp, Belgium
[4] Royal Brompton Hosp, London SW3 6LY, England
关键词
TRANSESOPHAGEAL ECHOCARDIOGRAPHY; COMPUTED-TOMOGRAPHY; AREA; DIMENSIONS; COMPLEX;
D O I
10.1016/j.amjcard.2011.02.348
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of the present study was to compare the aortic valve area, aortic valve annulus, and aortic root dimensions measured using magnetic resonance imaging (MRI) with catheterization, transthoracic echocardiography (TTE), and transesophageal echocardiography (TEE). An optimal prosthesis aortic root match is an essential goal when evaluating patients for transcatheter aortic valve implantation. Comparisons between MRI and the other imaging techniques are rare and need validation. In 24 consecutive, high-risk, symptomatic patients with severe aortic stenosis, aortic valve area was prospectively determined using MRI and direct planimetry using three-dimensional TTE and calculated by catheterization using the Gorlin equation and by Doppler echocardiography using the continuity equation. Aortic valve annulus and the aortic root dimensions were prospectively measured using MRI, 2-dimensional TTE, and invasive aortography. In addition, aortic valve annulus was measured using TEE. No differences in aortic valve area were found among MRI, Doppler echocardiography, and 3-dimensional TTE compared with catheterization (p = NS). Invasive angiography underestimated aortic valve annulus compared with MRI (p < 0.001), TEE (p < 0.001), and 2-dimensional TTE (p < 0.001). Two-dimensional TTE tended to underestimate the aortic valve annulus diameters compared to TEE and MRI. In contrast to 2-dimensional TTE, 3 patients had aortic valve annulus beyond the transcatheter aortic valve implantation range using TEE and MRI. In conclusion, MRI planimetry, Doppler, and 3-dimensional TTE provided an accurate estimate of the aortic valve area compared to catheterization. MRI and TEE provided similar and essential assessment of the aortic valve annulus dimensions, especially at the limits of the transcatheter aortic valve implantation range. (C) 2011 Elsevier Inc. All rights reserved. (Am J Cardiol 2011;108:92-98)
引用
收藏
页码:92 / 98
页数:7
相关论文
共 19 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Aortic Root Measurement by Cardiovascular Magnetic Resonance Specification of Planes and Lines of Measurement and Corresponding Normal Values [J].
Burman, Elisabeth D. ;
Keegan, Jennifer ;
Kilner, Philip J. .
CIRCULATION-CARDIOVASCULAR IMAGING, 2008, 1 (02) :104-113
[3]   Echocardiography for transcatheter aortic valve implantation [J].
Chin, Derek .
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2009, 10 (01) :I21-I29
[4]   Assessment of aortic stenosis by three-dimensional echocardiography: an accurate and novel approach [J].
Goland, Sorel ;
Trento, Alfredo ;
Iida, Kiyoshi ;
Czer, Lawrence S. C. ;
De Robertis, Michele ;
Naqvi, Tasneem Z. ;
Tolstrup, Kirsten ;
Akima, Takashi ;
Luo, Huai ;
Siegel, Robert J. .
HEART, 2007, 93 (07) :801-807
[5]   HYDRAULIC FORMULA FOR CALCULATION OF THE AREA OF THE STENOTIC MITRAL VALVE, OTHER CARDIAC VALVES, AND CENTRAL CIRCULATORY SHUNTS .1. [J].
GORLIN, R ;
GORLIN, SG .
AMERICAN HEART JOURNAL, 1951, 41 (01) :1-29
[6]   Magnetic resonance to assess the aortic valve area in aortic stenosis - How does it compare to current diagnostic standards? [J].
John, AS ;
Dill, T ;
Brandt, RR ;
Rau, M ;
Ricken, W ;
Bachmann, G ;
Hamm, CW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (03) :519-526
[7]   Pre-Procedural Imaging of Aortic Root Orientation and Dimensions Comparison Between X-Ray Angiographic Planar Imaging and 3-Dimensional Multidetector Row Computed Tomography [J].
Kurra, Vikram ;
Kapadia, Samir R. ;
Tuzcu, E. Murat ;
Halliburton, Sandra S. ;
Svensson, Lars ;
Roselli, Eric E. ;
Schoenhagen, Paul .
JACC-CARDIOVASCULAR INTERVENTIONS, 2010, 3 (01) :105-113
[8]   Multimodal Assessment of the Aortic Annulus Diameter Implications for Transcatheter Aortic Valve Implantation [J].
Messika-Zeitoun, David ;
Serfaty, Jean-Michel ;
Brochet, Eric ;
Ducrocq, Gregory ;
Lepage, Laurent ;
Detaint, Delphine ;
Hyafil, Fabien ;
Himbert, Dominique ;
Pasi, Nicoletta ;
Laissy, Jean-Pierre ;
Iung, Bernard ;
Vahanian, Alec .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (03) :186-194
[9]   Role of Echocardiography in Percutaneous Aortic Valve Implantation [J].
Moss, Robert R. ;
Ivens, Emma ;
Pasupati, Sanjeevan ;
Humphries, Karin ;
Thompson, Christopher R. ;
Munt, Brad ;
Sinhal, Ajay ;
Webb, John G. .
JACC-CARDIOVASCULAR IMAGING, 2008, 1 (01) :15-24
[10]   Comparison of Aortic Root Dimensions and Geometries Before and After Transcatheter Aortic Valve Implantation by 2-and 3-Dimensional Transesophageal Echocardiography and Multislice Computed Tomography [J].
Ng, Arnold C. T. ;
Delgado, Victoria ;
van der Kley, Frank ;
Shanks, Miriam ;
van de Veire, Nico R. L. ;
Bertini, Matteo ;
Nucifora, Gaetano ;
van Bommel, Rutger J. ;
Tops, Laurens F. ;
de Weger, Arend ;
Tavilla, Giuseppe ;
de Roos, Albert ;
Kroft, Lucia J. ;
Leung, Dominic Y. ;
Schuijf, Joanne ;
Schalij, Martin J. ;
Bax, Jeroen J. .
CIRCULATION-CARDIOVASCULAR IMAGING, 2010, 3 (01) :94-102