Three-dimensionally corrective external fixator system for proximal femoral osteotomy

被引:18
作者
Ito, H [1 ]
Minami, A [1 ]
Suzuki, K [1 ]
Matsuno, T [1 ]
机构
[1] Hokkaido Univ, Sch Med, Dept Orthopaed Surg, Kita Ku, Sapporo, Hokkaido 0608638, Japan
关键词
external fixator system; Legg-Calve-Perthes disease; proximal femoral osteotomy; slipped capital femoral epiphysis;
D O I
10.1097/00004694-200109000-00019
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The authors introduce a technique of proximal femoral osteotomy using three-dimensionally corrective external fixator systems for treatment of Legg-Calve-Perthes disease and slipped capital femoral epiphysis, and describe the results of 47 hips in 44 children. Postoperative casting immobilization was not performed. All osteotomies gained the correction angle as planned. All osteotomies healed with no nonunions, malunions, device failures, or avascular necrosis. Complications occurred iri 7 (15%) of 47 hips. Pin tract infection occurred in six (13%) hips; one of these (2%) needed debridement. The bulky external fixator caused slight inconvenience, but the current results indicate that the proximal femoral osteotomy using the external fixator system provided an easily repeatable intraoperative varus/valgus and flexion/extension correction compared with other internal fixation devices, and sufficient fixation for children without significant complications.
引用
收藏
页码:652 / 656
页数:5