A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study

被引:31
作者
Zanella, Alberto [1 ]
Cressoni, Massimo [1 ]
Epp, Myra [1 ]
Stylianou, Mario [2 ]
Kolobow, Theodor [1 ]
机构
[1] NHLBI, NIH, Pulm & Crit Care Med Branch, Bethesda, MD 20892 USA
[2] NHLBI, NIH, Off Biostat Res, OD, Bethesda, MD 20892 USA
关键词
tube; endotracheal; cuff; leakage; fluid; aspiration; ventilator-associated pneumonia;
D O I
10.1007/s00134-008-1016-9
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: We designed a new endotracheal tube (ETT) cuff that does not form the folds that cause leakage of colonized subglottic secretions upon inflation within the trachea: a standard high-volume low-pressure cuff was draped with a second, highly elastic cuff made of a low-protein guayule natural latex rubber with 0.5 ml gel between the cuffs. We compared this prototype ETT cuff with four commercially available ETTs for efficacy in the prevention of fluid leakage across the cuff. Design: In vitro study. Measurements and results: We compared fluid leakage in our prototype cuff with that in four commercially available ETTs. Three cylindrical glass tubes 16, 20, and 22 mm in diameter were used as model tracheas, and five different intracuff pressures ( 20, 25, 30, 40, and 50 cmH(2)O) were tested. Each test was repeated three times with new ETTs. The guayule latex ETT cuff showed an average fluid leakage of 0.0007 +/- 0.002 ml/min which was significantly lower than that in any of the other ETTs ( Microcuff 0.07 +/- 0.09, Mallinckrodt/Hi-Lo Evac 5 +/- 5, Euromedical 7 +/- 4, Sheridan/CF 41 +/- 69). Conclusions: Our prototype ETT cuff significantly reduced fluid leakage in this bench-top study. In vivo testing and evaluation is to follow.
引用
收藏
页码:1145 / 1149
页数:5
相关论文
共 17 条
[1]  
BERNHARD WN, 1982, ANESTH ANALG, V61, P36
[2]  
CAREY AB, 1995, J ALLERGY CLIN IMMUN, V98, P895
[3]   EVOLUTION OF TRACHEAL INJURY DUE TO VENTILATORY ASSISTANCE THROUGH CUFFED TUBES - A PATHOLOGIC STUDY [J].
COOPER, JD ;
GRILLO, HC .
ANNALS OF SURGERY, 1969, 169 (03) :334-&
[4]  
Cornish K, 1999, J BIOMED MATER RES, V47, P434, DOI 10.1002/(SICI)1097-4636(19991205)47:3<434::AID-JBM20>3.0.CO
[5]  
2-N
[6]  
CORNISH K, 2006, GLOBAL HDB DIRECTORY, P71
[7]  
CORNISH K, 2005, P 168 TECHN M RUBB D
[8]   Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube [J].
Dullenkopf, A ;
Gerber, A ;
Weiss, M .
INTENSIVE CARE MEDICINE, 2003, 29 (10) :1849-1853
[9]   HIGH-VOLUME, LOW-PRESSURE CUFFS - ARE THEY ALWAYS LOW-PRESSURE [J].
GUYTON, D ;
BANNER, MJ ;
KIRBY, RR .
CHEST, 1991, 100 (04) :1076-1081
[10]   Pneumonia in intubated patients: Role of respiratory airway care [J].
Rello, J ;
Sonora, R ;
Jubert, P ;
Artigas, A ;
Rue, M ;
Valles, J .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 154 (01) :111-115