Direct determination of total sulfur in wine using a continuum-source atomic-absorption spectrometer and an air-acetylene flame

被引:54
作者
Huang, MD
Becker-Ross, H
Florek, S
Heitmann, U
Okruss, M
机构
[1] ISAS, Inst Analyt Sci, D-12489 Berlin, Germany
[2] Gesell Forder Angew Opt, D-12489 Berlin, Germany
关键词
continuum source; atomic-absorption spectrometry; sulfur determination; wine analysis;
D O I
10.1007/s00216-005-3333-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Determination of sulfur in wine is an important analytical task, particularly with regard to food safety legislation, wine trade, and oenology. Hitherto existing methods for sulfur determination all have specific drawbacks, for example high cost and time consumption, poor precision or selectivity, or matrix effects. In this paper a new method, with low running costs, is introduced for direct, reliable, rapid, and accurate determination of the total sulfur content of wine samples. The method is based on measurement of the molecular absorption of carbon monosulfide (CS) in an ordinary air-acetylene flame by using a high-resolution continuum-source atomic-absorption spectrometer including a novel high-intensity short-arc xenon lamp. First results for total sulfur concentrations in different wine samples were compared with data from comparative ICP-MS measurements. Very good agreement within a few percent was obtained.
引用
收藏
页码:1877 / 1881
页数:5
相关论文
共 26 条
[1]   DETERMINATION OF TOTAL AND FREE SULFUR-DIOXIDE IN WINE BY FLOW-INJECTION ANALYSIS AND GAS-DIFFUSION USING PARA-AMINOAZOBENZENE AS THE COLORIMETRIC-REAGENT [J].
BARTROLI, J ;
ESCALADA, M ;
JORQUERA, CJ ;
ALONSO, J .
ANALYTICAL CHEMISTRY, 1991, 63 (21) :2532-2535
[2]   Echelle-spectrograph as a tool for studies of structured background in flame atomic absorption spectrometry [J].
Becker-Ross, H ;
Okruss, M ;
Florek, S ;
Heitmann, U ;
Huang, MD .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (10) :1493-1504
[3]  
BeckerRoss H, 1996, FRESEN J ANAL CHEM, V355, P300
[4]   FAST AND SIMPLE POLAROGRAPHIC METHOD FOR THE DETERMINATION OF FREE AND TOTAL SULFUR-DIOXIDE IN WINES AND OTHER COMMON BEVERAGES [J].
BRUNO, P ;
CASELLI, M ;
DIFANO, A ;
TRAINI, A .
ANALYST, 1979, 104 (1244) :1083-1087
[5]   Determination of sulfur dioxide and ascorbic acid in beverages using a dual channel flow injection electrochemical detection system [J].
Cardwell, TJ ;
Christophersen, MJ .
ANALYTICA CHIMICA ACTA, 2000, 416 (01) :105-110
[6]   DETERMINATION OF FREE SULFUR-DIOXIDE IN RED WINE BY ALTERNATING-CURRENT VOLTAMMETRY [J].
CARDWELL, TJ ;
CATTRALL, RW ;
CHEN, GN ;
SCOLLARY, GR ;
HAMILTON, IC .
ANALYST, 1991, 116 (03) :253-256
[7]   Determination of sulphite in wines by gas-diffusion flow injection analysis utilizing spectrophotometric pH-detection [J].
DecnopWeever, LG ;
Kraak, JC .
ANALYTICA CHIMICA ACTA, 1997, 337 (02) :125-131
[8]  
DEIBNER L, 1970, CHIM ANAL-PARIS, V52, P412
[9]   Determination of sulfur dioxide in wine using a quartz crystal microbalance [J].
Gomes, MT ;
Rocha, TA ;
Duarte, AC ;
Oliveira, JP .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1561-1564
[10]   Measurements on the Zeeman-splitting of analytical lines by means of a continuum source graphite furnace atomic absorption spectrometer with a linear charge coupled device array [J].
Heitmann, U ;
Schutz, M ;
BeckerRoss, H ;
Florek, S .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (9-10) :1095-1105