Lead explantation late after atrial perforation

被引:8
作者
Trigano, AJ [1 ]
Caus, T [1 ]
机构
[1] UNIV MARSEILLE,CTR HOSP LA TIMONE,SCH MED,DEPT CARDIAC SURG,F-13915 MARSEILLE 20,FRANCE
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1996年 / 19卷 / 08期
关键词
atrial perforation; Accufix(TM) atrial lead;
D O I
10.1111/j.1540-8159.1996.tb04200.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This report describes the case of a patient in whom atrial perforation with penetration of the thoracic wall was diagnosed 2 years after the implantation of an Accufix lead. Despite this complication, atrial detection in the bipolar mode and ventricular pacing were normal. Digital fluoroscopy detected a fracture with extrusion of a short segment of the retention wire. The rupture of the retention wire might have been the result, but was not the cause of the perforation.
引用
收藏
页码:1268 / 1269
页数:2
相关论文
共 5 条
  • [1] PERCUTANEOUS REMOVAL OF TRANSVENOUS PACING LEAD PERFORATING THE HEART, PERICARDIUM, AND PLEURA
    DEUTSCH, LS
    DANG, H
    BRANDON, JC
    OTT, R
    ALLEN, B
    FUTERMAN, C
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 156 (03) : 471 - 473
  • [2] CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .1. THE ATRIAL LEAD
    GLIKSON, M
    VONFELDT, LK
    SUMAN, VJ
    HAYES, DL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (08): : 1399 - 1404
  • [3] ACUTE PERICARDITIS RESULTING FROM AN ENDOCARDIAL ACTIVE FIXATION SCREW-IN ATRIAL LEAD
    GREENE, TO
    PORTNOW, AS
    HUANG, SKS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01): : 21 - 25
  • [4] THE RETENTION WIRE FIX
    PARSONNET, V
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05): : 955 - 957
  • [5] VANACKER R, 1993, EUR JCPE, V1, P67