Lead explantation late after atrial perforation

被引:8
作者
Trigano, AJ [1 ]
Caus, T [1 ]
机构
[1] UNIV MARSEILLE,CTR HOSP LA TIMONE,SCH MED,DEPT CARDIAC SURG,F-13915 MARSEILLE 20,FRANCE
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1996年 / 19卷 / 08期
关键词
atrial perforation; Accufix(TM) atrial lead;
D O I
10.1111/j.1540-8159.1996.tb04200.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This report describes the case of a patient in whom atrial perforation with penetration of the thoracic wall was diagnosed 2 years after the implantation of an Accufix lead. Despite this complication, atrial detection in the bipolar mode and ventricular pacing were normal. Digital fluoroscopy detected a fracture with extrusion of a short segment of the retention wire. The rupture of the retention wire might have been the result, but was not the cause of the perforation.
引用
收藏
页码:1268 / 1269
页数:2
相关论文
共 5 条
[1]   PERCUTANEOUS REMOVAL OF TRANSVENOUS PACING LEAD PERFORATING THE HEART, PERICARDIUM, AND PLEURA [J].
DEUTSCH, LS ;
DANG, H ;
BRANDON, JC ;
OTT, R ;
ALLEN, B ;
FUTERMAN, C .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 156 (03) :471-473
[2]   CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .1. THE ATRIAL LEAD [J].
GLIKSON, M ;
VONFELDT, LK ;
SUMAN, VJ ;
HAYES, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (08) :1399-1404
[3]   ACUTE PERICARDITIS RESULTING FROM AN ENDOCARDIAL ACTIVE FIXATION SCREW-IN ATRIAL LEAD [J].
GREENE, TO ;
PORTNOW, AS ;
HUANG, SKS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01) :21-25
[4]   THE RETENTION WIRE FIX [J].
PARSONNET, V .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05) :955-957
[5]  
VANACKER R, 1993, EUR JCPE, V1, P67