Impedimetric blood pH sensor based on MoS2-Nafion coated microelectrode

被引:20
作者
Awasthij, Prasoon [1 ]
Mukherjee, Ranjan [1 ]
Kare, Siva Prakasam O. [1 ]
Das, Soumen [1 ]
机构
[1] IIT, Sch Med Sci & Technol, BioMEMS Lab, Kharagpur 721302, W Bengal, India
关键词
GAS-ANALYSIS; CAPILLARY BLOOD; VENOUS PH; ACID-BASE; ARTERIAL; MOS2; NAFION; TRANSPORT; MEMBRANES; STATE;
D O I
10.1039/c6ra17786g
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Ion selective electrodes are a good alternative to glass-based pH sensors to replace traditional bulky pH sensors with compact and robust point of care analyzers. This is especially important for healthcare applications, as the pH-sensing element of hand-held blood gas analyzers to aid blood gas analysis in remote or mobile medical set ups. In this work, we have presented an impedimetric pH sensor consisting of parallel coplanar electrodes of aluminum coated with molybdenum disulfide (MoS2) nanoparticles and Nafion as a permselective layer. This sensor showed a sensitivity of 6193 Omega pH(-1) and a response time of 4 s on testing with Sorensen's phosphate buffers at different pH values, used as a physiological buffer. A preliminary study of the sensor with capillary blood shows that its estimated pH lies close to that of the venous blood pH, thus showcasing good characteristics to be a potential candidate for the pH-sensing element of a point of care analyzer.
引用
收藏
页码:102088 / 102095
页数:8
相关论文
共 47 条
[31]   Characteristics of Self-Assembled Ultrathin Nafion Films [J].
Paul, Devproshad K. ;
Karan, Kunal ;
Docoslis, Aristides ;
Giorgi, Javier B. ;
Pearce, Joshua .
MACROMOLECULES, 2013, 46 (09) :3461-3475
[32]   A method for calculation of arterial acid-base and blood gas status from measurements in the peripheral venous blood [J].
Rees, SE ;
Toftegaard, M ;
Andreassen, S .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2006, 81 (01) :18-25
[33]   Monolayer MoS2 Bandgap Modulation by Dielectric Environments and Tunable Bandgap Transistors [J].
Ryou, Junga ;
Kim, Yong-Sung ;
Santosh, K. C. ;
Cho, Kyeongjae .
SCIENTIFIC REPORTS, 2016, 6
[34]   WO3 Nanoparticle-Based Conformable pH Sensor [J].
Santos, Lidia ;
Neto, Joana P. ;
Crespo, Ana ;
Nunes, Daniela ;
Costa, Nuno ;
Fonseca, Isabel M. ;
Barquinha, Pedro ;
Pereira, Luis ;
Silva, Jorge ;
Martins, Rodrigo ;
Fortunato, Elvira .
ACS APPLIED MATERIALS & INTERFACES, 2014, 6 (15) :12226-12234
[35]   MoS2 Field-Effect Transistor for Next-Generation Label-Free Biosensors [J].
Sarkar, Deblina ;
Liu, Wei ;
Xie, Xuejun ;
Anselmo, Aaron C. ;
Mitragotri, Samir ;
Banerjee, Kaustav .
ACS NANO, 2014, 8 (04) :3992-4003
[36]   VARIABILITY OF ARTERIAL BLOOD-GAS VALUES OVER TIME IN STABLE MEDICAL ICU PATIENTS [J].
SASSE, SA ;
CHEN, PA ;
MAHUTTE, CK .
CHEST, 1994, 106 (01) :187-193
[37]   Soil pH Mapping with an On-The-Go Sensor [J].
Schirrmann, Michael ;
Gebbers, Robin ;
Kramer, Eckart ;
Seidel, Jan .
SENSORS, 2011, 11 (01) :573-598
[38]   The invention and development of blood gas analysis apparatus [J].
Severinghaus, JW .
ANESTHESIOLOGY, 2002, 97 (01) :253-256
[39]   CLINICAL-EVALUATION CONTINUOUS REAL-TIME INTRAARTERIAL BLOOD-GAS MONITORING DURING ANESTHESIA AND SURGERY BY FIBER OPTIC SENSOR [J].
SMITH, BE ;
KING, PH ;
SCHLAIN, L .
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1992, 9 (01) :45-52
[40]  
Steed J.W., 2012, Supramolecular chemistry: from molecules to nanomaterials, DOI [10.1002/9780470661345.smc002, DOI 10.1002/9780470661345]