Comparison of bone-implant attachment strength between the implants with hydroxyapatite-coating and tricalciumphosphate-coating on titanium arc sprayed titanium

被引:0
|
作者
Hayashi, K [1 ]
Hara, T [1 ]
Inamura, T [1 ]
Iwamoto, Y [1 ]
机构
[1] Kyushu Univ, Fac Med, Dept Orthopaed, Higashi Ku, Fukuoka 812, Japan
来源
关键词
D O I
10.1016/B978-008042692-1/50069-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We compared the bone-implant attachment strength between are sprayed titanium without ceramic coating(Proarc), hydroxyapatite(HA)-coated Proarc(Proarc HA),and tricalciumphosphate(TCP)-coated Proarc(Proarc TCP) by push-out test in dog femurs,Push-out strength of Proarc-HA was significantly greater than that of Proarc(p<0.05) 4 weeks postimplantation. There was no significant difference between Proarc and Proarc TCP. There was no significant difference between Proarc HA and Proarc TCP,Combination with significant difference between three materials was not found.
引用
收藏
页码:291 / 294
页数:4
相关论文
共 50 条
  • [21] THE EFFECT OF HYDROXYAPATITE COATING ON BONE-GROWTH INTO POROUS TITANIUM-ALLOY IMPLANTS
    OONISHI, H
    YAMAMOTO, M
    ISHIMARU, H
    TSUJI, E
    KUSHITANI, S
    AONO, M
    UKON, Y
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1989, 71 (02): : 213 - 216
  • [22] Effect of hydroxyapatite/tricalcium-phosphate coating on osseointegration of plasma-sprayed titanium alloy implants
    Stewart, M
    Welter, JF
    Goldberg, VM
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2004, 69A (01) : 1 - 10
  • [23] Sand-blasting treatment as a way to improve the adhesion strength of hydroxyapatite coating on titanium implant
    Grubova, I.
    Priamushko, T.
    Surmeneva, M.
    Korneva, O.
    Epple, M.
    Prymak, O.
    Surmenev, R.
    5TH INTERNATIONAL CONGRESS ON ENERGY FLUXES AND RADIATION EFFECTS 2016, 2017, 830
  • [24] The correlation between osseointegration and bonding strength at the bone-implant interface: In-vivo & ex-vivo investigations on hydroxyapatite and hydroxyapatite/titanium coatings
    Ghadami, Farhad
    Hamedani, Mohsen Amani
    Rouhi, Gholamreza
    Saber-Samandari, Saeed -
    Dehghan, Mohammad Mehdi
    Farzad-Mohajeri, Saeed
    Abbas, Fatemeh Mashhadi-
    JOURNAL OF BIOMECHANICS, 2022, 144
  • [25] Ceramic hydroxyapatite coating on titanium implants drives selective bone marrow stromal cell adhesion
    Torensma, R
    ter Brugge, PJ
    Jansen, JA
    Figdor, CG
    CLINICAL ORAL IMPLANTS RESEARCH, 2003, 14 (05) : 569 - 577
  • [26] Effects of hydroxyapatite tricalcium phosphate coating and intracancellous placement on bone ingrowth in titanium fibermetal implants
    Dean, JC
    Tisdel, CL
    Goldberg, VM
    Parr, J
    Davy, D
    Stevenson, S
    JOURNAL OF ARTHROPLASTY, 1995, 10 (06): : 830 - 838
  • [27] In vitro and in vivo mechanical evaluations of plasma-sprayed hydroxyapatite coatings on titanium implants: The effect of coating characteristics
    Yang, CY
    Lin, RM
    Wang, BC
    Lee, TM
    Chang, E
    Hang, YS
    Chen, PQ
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1997, 37 (03): : 335 - 345
  • [28] Effect of bone sialoprotein and collagen coating on cell attachment to TICER® and pure titanium implant surfaces
    Graf, H. -L.
    Stoeva, S.
    Armbruster, F. P.
    Neuhaus, J.
    Hilbig, H.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2008, 37 (07) : 634 - 640
  • [29] EFFECT OF SURFACE-ROUGHNESS OF HYDROXYAPATITE-COATED TITANIUM ON THE BONE-IMPLANT INTERFACE SHEAR-STRENGTH
    HAYASHI, K
    INADOME, T
    TSUMURA, H
    NAKASHIMA, Y
    SUGIOKA, Y
    BIOMATERIALS, 1994, 15 (14) : 1187 - 1191
  • [30] Effect of immobilized bone morphogenic protein 2 coating of titanium implants on peri-implant bone formation
    Schliephake, H
    Aref, A
    Scharnweber, D
    Bierbaum, S
    Roessler, S
    Sewing, A
    CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (05) : 563 - 569