Noise characteristics of virtual monoenergetic images from a novel detector-based spectral CT scanner

被引:59
作者
Kalisz, Kevin [1 ]
Rassouli, Negin [1 ]
Dhanantwari, Amar [2 ]
Jordan, David [1 ]
Rajiah, Prabhakar [1 ,3 ]
机构
[1] Univ Hosp Cleveland, Dept Radiol, Med Ctr, 2074 Abington Rd, Cleveland, OH 44106 USA
[2] Philips Healthcare, Cleveland, OH USA
[3] UT Southwestern Med Ctr, Dept Radiol, Cardiothorac Imaging, Dallas, TX USA
关键词
Spectral; Dual energy; CT; Noise; Abdomen; DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; PULMONARY ANGIOGRAPHY; CONTRAST; REDUCTION; ALGORITHM; CONSPICUOUSNESS; OPTIMIZATION; ENHANCEMENT; VESSELS;
D O I
10.1016/j.ejrad.2017.11.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim: To evaluate the noise characteristics of virtual monoenergetic images (VMI) obtained from a recently introduced dual-layer detector-based spectral CT (SDCT), both in a phantom and patients. Materials and methods: A cylindrical Catphan (R) 600 phantom (The Phantom Library, Salem NY, USA) was scanned using the SDCT. Image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured in VMI from 40 to 200 keV as well as conventional 120 kVp images. One hundred consecutive patients who had an abdominal CT on the SDCT were then recruited in the study. Noise, SNR and CNR were measured in the liver, pancreas, spleen, kidney, abdominal aorta, portal vein, muscle, bone, and fat, both in VMI (40-200 keV) and conventional 120 kVp images. Qualitative image analysis was performed by an independent reader for vascular enhancement and image quality on a 5 point scale (1-worst, 5-best). Results: On phantom studies, noise was low at all energies of VMI. Noise was highest at 40 keV (5.3 +/- 0.2 HU), gradually decreased up to 70 keV (3.6 +/- 0.2 HU), after which it remained constant up to 200 keV (3.5 +/- 0.2 HU). In the patient cohort, noise was low (< 25 HU) at all the energy levels of VMI for all the regions, with the exception of bone. For example, noise in the liver was highest at 40 keV (13.2 +/- 4.6 HU), steadily decreased up to 70 keV (12.0 +/- 4.4 HU) and then remained constantly low up to 200 keV (11.6 +/- 4.3 HU). For liver, pancreas, portal vein, aorta, muscle and fat, noise at all levels of VMI was lower than of conventional images (p < 0.01). For all organs, SNR, and CNR were highest at 40 keV (6.8-34.9; 18.3-44.9, respectively) after which they gradually decreased up to 120 keV (3.4-6.5; 9.5-13.0) and then remained constant to 200 keV (2.6-5.5; 8.5-12.5). Qualitative scores of VMI up to 70 keV were significantly higher than the conventional images (p <= 0.01), whereas for VMI >= 80 keV, they were lower than conventional images (p < 0.001). Conclusion: VMI obtained from the novel SDCT scanner have low noise across the entire spectrum of energies. There are significant SNR and CNR improvements compared to conventional 120 kVp images.
引用
收藏
页码:118 / 125
页数:8
相关论文
共 36 条
[11]   Dual-energy MDCT: Comparison of pulmonary artery enhancement on dedicated CT pulmonary angiography, routine and low contrast volume studies [J].
Godoy, Myrna C. B. ;
Heller, Samantha L. ;
Naidich, David P. ;
Assadourian, Bernard ;
Leidecker, Christianne ;
Schmidt, Bernhard ;
Vlahos, Ioannis .
EUROPEAN JOURNAL OF RADIOLOGY, 2011, 79 (02) :E11-E17
[12]   Assessment of an Advanced Image-Based Technique to Calculate Virtual Monoenergetic Computed Tomographic Images From a Dual-Energy Examination to Improve Contrast-To-Noise Ratio in Examinations Using Iodinated Contrast Media [J].
Grant, Katharine L. ;
Flohr, Thomas G. ;
Krauss, Bernhard ;
Sedlmair, Martin ;
Thomas, Christoph ;
Schmidt, Bernhard .
INVESTIGATIVE RADIOLOGY, 2014, 49 (09) :586-592
[13]   Monoenergetic reconstructions for imaging of coronary artery stents using spectral detector CT: In-vitro experience and comparison to conventional images [J].
Hickethier, Tilman ;
Baessler, Bettina ;
Kroeger, Jan Robert ;
Doerner, Jonas ;
Pahn, Gregor ;
Maintz, David ;
Michels, Guido ;
Bunck, Alexander C. .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2017, 11 (01) :33-39
[14]   Determination of the optimal energy level in spectral CT imaging for displaying abdominal vessels in pediatric patients [J].
Hu, Di ;
Yu, Tong ;
Duan, Xiaomin ;
Peng, Yun ;
Zhai, Renyou .
EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (03) :589-594
[15]   AN ALGORITHM FOR NOISE SUPPRESSION IN DUAL ENERGY CT MATERIAL DENSITY IMAGES [J].
KALENDER, WA ;
KLOTZ, E ;
KOSTARIDOU, L .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (03) :218-224
[16]   Application- and patient size-dependent optimization of x-ray spectra for CT [J].
Kalender, Willi A. ;
Deak, Paul ;
Kellermeier, Markus ;
van Straten, Marcel ;
Vollmar, Sabrina V. .
MEDICAL PHYSICS, 2009, 36 (03) :993-1007
[17]   Maximizing Iodine Contrast-to-Noise Ratios in Abdominal CT Imaging through Use of Energy Domain Noise Reduction and Virtual Monoenergetic Dual-Energy CT [J].
Leng, Shuai ;
Yu, Lifeng ;
Fletcher, Joel G. ;
McCollough, Cynthia H. .
RADIOLOGY, 2015, 276 (02) :562-570
[18]   Effect of a Noise-Optimized Second-Generation Monoenergetic Algorithm on Image Noise and Conspicuity of Hypervascular Liver Tumors: An In Vitro and In Vivo Study [J].
Marin, Daniele ;
Ramirez-Giraldo, Juan Carlos ;
Gupta, Sonia ;
Fu, Wanyi ;
Stinnett, Sandra S. ;
Mileto, Achille ;
Bellini, Davide ;
Patel, Bhavik ;
Samei, Ehsan ;
Nelson, Rendon C. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 206 (06) :1222-1232
[19]   Virtual Monochromatic Spectral Imaging with Fast Kilovoltage Switching: Improved Image Quality as Compared with That Obtained with Conventional 120-kVp CT [J].
Matsumoto, Kazuhiro ;
Jinzaki, Masahiro ;
Tanami, Yutaka ;
Ueno, Akihisa ;
Yamada, Minoru ;
Kuribayashi, Sachio .
RADIOLOGY, 2011, 259 (01) :257-262
[20]   Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications [J].
McCollough, Cynthia H. ;
Leng, Shuai ;
Yu, Lifeng ;
Fletcher, Joel G. .
RADIOLOGY, 2015, 276 (03) :637-653