Determinants of aerosolized albuterol delivery to mechanically ventilated infants

被引:23
|
作者
Coleman, DM
Kelly, HW
McWilliams, BC
机构
[1] UNIV NEW MEXICO,HLTH SCI CTR,SCH MED,COLL PHARM,ALBUQUERQUE,NM 87131
[2] UNIV NEW MEXICO,HLTH SCI CTR,SCH MED,DEPT PEDIAT,ALBUQUERQUE,NM 87131
关键词
aerosol delivery; albuterol; infant; mechanical ventilation; metered-dose inhaler; nebulizer; spacer;
D O I
10.1378/chest.109.6.1607
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
An in vitro lung model and a volume ventilator were used to evaluate the delivery of aerosolized albuterol through an infant ventilator circuit, We compared the following: continuous nebulization (CNA) and intermittent nebulization (INA); various nebulizer gas flows, 5.0, 6.5, and 8.0 L/min; and duty cycle of 33% and 50%. The efficiency and consistency of aerosol delivery by metered-dose inhaler (MDI) with four different spacer devices and by nebulizer positioned at the manifold and at the same position as the MDI were also evaluated. A volume ventilator (Servo 900B) was used with settings selected to reflect those of a moderately to severely ill 4-kg infant. A 3.5-mm endotracheal tube,vas used in all experiments, A specific type of nebulizer used (Airlife Misty Neb; Baxter; Valencia, Calif) and several spacers were studied (Aerochamber and Aerovent, Diemolding Healthcare Div, Canastota, NY; ACE, Monaghan Medical Corp, Plattsburgh, NY; and an in-line MDI adapter, Instrumentation Industries Inc, Pittsburgh). CNA delivered significantly more aerosol to the lung model (4.8+/-0.6% of the starting dose) than INA (3.8+/-0.3%; p<0.01), There was a significant stepwise decrease in aerosol delivery with increasing nebulizer flow (4.8+/-1.3% at 5.0 L/min; 3.7+/-1.1% at 6.5 L/min; and 2.7+/-1.1% at 8.0 L/min). Increasing duty cycle did not significantly affect delivery. Overall the spacers with MDI were more efficient than the nebulizer in either position delivering about twice the percentage of the starting dose than the nebulizers. All modes of delivery, except the Aerochamber, demonstrated a marked degree of variability. Most of the starting dose of albuterol either remained in the nebulizer (30.4+/-6.0% at 5.0 L/min and 25.3+/-4.1% at 8.0 L/min) or was deposited in the inspiratory tubing (34.7+/-0.7% at 5.0 L/min and 43.7+/-4.9% at 8.0 L/min) in our system. In conclusion, we have confirmed that aerosol delivery depends on the mode of delivery and the operating conditions. Although delivery with an MDI and spacer is more efficient than a nebulizer, both methods may produce high variability depending on the method or spacer used.
引用
收藏
页码:1607 / 1613
页数:7
相关论文
共 50 条
  • [41] Delivery of albuterol in a pediatric emergency department
    Williams, JR
    Bothner, JP
    Swanton, D
    PEDIATRIC EMERGENCY CARE, 1996, 12 (04) : 263 - 267
  • [42] Measuring the effects of airway clearance in mechanically ventilated infants and children: A systematic review
    McAlinden, Bronagh M.
    Hough, Judith L.
    Kuys, Suzanne
    PHYSIOTHERAPY, 2022, 117 : 47 - 62
  • [43] Performance of different add-on devices in dual limb non-invasive mechanically ventilated circuit
    Rashad, Amal E.
    Mohammad, Mohammad F.
    Eckes, Myriam
    Poree, Thierry
    Sarhan, Rania M.
    Madney, Yasmin M.
    Abdelrahim, Mohamed EA.
    JOURNAL OF DRUG DELIVERY SCIENCE AND TECHNOLOGY, 2021, 66
  • [44] Bronchodilator delivery with metered-dose inhalers in mechanically-ventilated patients
    Dhand, R
    Tobin, MJ
    EUROPEAN RESPIRATORY JOURNAL, 1996, 9 (03) : 585 - 595
  • [45] Albuterol Delivery by 4 Different Nebulizers Placed in 4 Different Positions in a Pediatric Ventilator In Vitro Model
    Berlinski, Ariel
    Willis, J. Randy
    RESPIRATORY CARE, 2013, 58 (07) : 1124 - 1133
  • [46] Development of a New Technique for the Efficient Delivery of Aerosolized Medications to Infants on Mechanical Ventilation
    Longest, P. Worth
    Tian, Geng
    PHARMACEUTICAL RESEARCH, 2015, 32 (01) : 321 - 336
  • [47] Pulsed dosed delivery of oxygen in mechanically ventilated pigs with acute lung injury
    Gustafson, Joshua D.
    Yang, Sung
    Blakeman, Thomas Chris
    Dorlac, Warren C.
    Branson, Richard
    JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2013, 75 (05) : 775 - 779
  • [48] In Vitro Measurements of Spiriva Respimat Dose Delivery in Mechanically Ventilated Tracheostomy Patients
    Mehri, Rym
    Alatrash, Abubakar
    Ogrodnik, Nicholas
    Matida, Edgar A.
    Fiorenza, Frank
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2021, 34 (04) : 242 - 250
  • [49] Delivery options and devices for aerosolized therapeutics
    Anderson, PJ
    CHEST, 2001, 120 (03) : 89S - 93S
  • [50] Determinants of point-of-care ultrasound lung sliding amplitude in mechanically ventilated patients
    Briganti, David N.
    Choi, Christine E.
    Nguyen, Julien
    Lanks, Charles W.
    ULTRASOUND JOURNAL, 2023, 15 (01)