Rapid method for evaluating reversed-phase high-performance liquid chromatography column stability

被引:28
|
作者
Fonseca, DA
Gutiérrez, HR
Collins, KE
Collins, CH
机构
[1] Univ Estadual Campinas, Inst Quim, BR-13084971 Campinas, SP, Brazil
[2] Univ Estadual Campinas, Inst Fis, BR-13084971 Campinas, SP, Brazil
关键词
stationary phases; LC; stability studies;
D O I
10.1016/j.chroma.2003.12.033
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A procedure is presented for the rapid evaluation of HPLC stationary phase stability at pH 8.4 or 10.1 using a temperature of 60degreesC. Mobile phase (MeOH-0.1 mol 1(-1) aqueous NaHCO3, 50:50, v/v) is continuously passed through the column with periodic injections of a test solution until the several chromatographic parameters of the resulting chromatograms are degraded. The tests were applied to several commercial and laboratory-made stationary phases. After degradation two of these phases, one commercial and one laboratory-made, were examined by elemental analysis and scanning electron microscopy to elucidate the degradation process. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:149 / 155
页数:7
相关论文
共 50 条
  • [1] Effect of buffers on silica-based column stability in reversed-phase high-performance liquid chromatography
    Claessens, HA
    vanStraten, MA
    Kirkland, JJ
    JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) : 259 - 270
  • [2] Characterization of an avidin-bonded column for direct injection in reversed-phase high-performance liquid chromatography
    Tanaka, H
    Takahashi, K
    Ohira, M
    JOURNAL OF CHROMATOGRAPHY A, 2000, 869 (1-2) : 151 - 157
  • [3] Evaluation and modulation of selectivity in reversed-phase high-performance liquid chromatography
    Sándi, A
    Szepesy, L
    JOURNAL OF CHROMATOGRAPHY A, 1999, 845 (1-2) : 113 - 131
  • [4] Organic monoliths for high-performance reversed-phase liquid chromatography
    Liu, Kun
    Aggarwal, Pankaj
    Lawson, John S.
    Tolley, H. Dennis
    Lee, Milton L.
    JOURNAL OF SEPARATION SCIENCE, 2013, 36 (17) : 2767 - 2781
  • [5] Reversed-phase liquid chromatography on a microspherical carbon column at high temperature
    Yamaki, S
    Isobe, T
    Okuyama, T
    Shinoda, T
    JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) : 189 - 194
  • [6] Microwave-immobilized polybutadiene stationary phase for reversed-phase high-performance liquid chromatography
    Lopes, NP
    Collins, KE
    Jardim, ICSF
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1030 (1-2) : 225 - 229
  • [7] Preparation of a polybutadiene stationary phase immobilized by γ radiation for reversed-phase high-performance liquid chromatography
    Lopes, NP
    Collins, KE
    Jardim, ICSF
    JOURNAL OF CHROMATOGRAPHY A, 2003, 987 (1-2) : 77 - 85
  • [8] A comparison and column selection of Hydrophilic Interaction Liquid Chromatography and Reversed-Phase High-Performance Liquid Chromatography for detection of DNA methylation
    Rastegar, Leila
    Mighani, Hossein
    Ghassempour, Alireza
    ANALYTICAL BIOCHEMISTRY, 2018, 557 : 123 - 130
  • [9] Comparative study of test methods for reversed-phase columns for high-performance liquid chromatography
    Claessens, HA
    van Straten, MA
    Cramers, CA
    Jezierska, M
    Buszewski, B
    JOURNAL OF CHROMATOGRAPHY A, 1998, 826 (02) : 135 - 156
  • [10] Separation of polyesters by gradient reversed-phase high-performance liquid chromatography on a 1.5 μm non-porous column
    Rissler, K