共 50 条
Rapid method for evaluating reversed-phase high-performance liquid chromatography column stability
被引:28
|作者:
Fonseca, DA
Gutiérrez, HR
Collins, KE
Collins, CH
机构:
[1] Univ Estadual Campinas, Inst Quim, BR-13084971 Campinas, SP, Brazil
[2] Univ Estadual Campinas, Inst Fis, BR-13084971 Campinas, SP, Brazil
关键词:
stationary phases;
LC;
stability studies;
D O I:
10.1016/j.chroma.2003.12.033
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
A procedure is presented for the rapid evaluation of HPLC stationary phase stability at pH 8.4 or 10.1 using a temperature of 60degreesC. Mobile phase (MeOH-0.1 mol 1(-1) aqueous NaHCO3, 50:50, v/v) is continuously passed through the column with periodic injections of a test solution until the several chromatographic parameters of the resulting chromatograms are degraded. The tests were applied to several commercial and laboratory-made stationary phases. After degradation two of these phases, one commercial and one laboratory-made, were examined by elemental analysis and scanning electron microscopy to elucidate the degradation process. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:149 / 155
页数:7
相关论文