A simple, versatile, and automated pulse-diffusion-focusing strategy for sensitive milliliter-level-injection HILIC-MS/MS analysis of hydrophilic toxins

被引:10
作者
Long, Fei [1 ]
Zhang, Mei [1 ]
Yuan, Jiahao [1 ]
Du, Juan [1 ]
Ma, Ande [1 ]
Pan, Jialiang [1 ]
机构
[1] Southern Med Univ, Guangdong Prov Key Lab Trop Dis Res, Dept Hyg Detect Ctr, Sch Publ Hlth, Guangzhou, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
Pulse diffusion focusing; Milliliter-level injection; Hydrophilic interaction chromatography-tandemmass spectrometry (HILIC-MS/MS); Tetrodotoxin; INTERACTION LIQUID-CHROMATOGRAPHY; TANDEM MASS-SPECTROMETRY; SOLID-PHASE EXTRACTION; LARGE-VOLUME INJECTION; STATIONARY PHASES; HUMAN URINE; GRADIENT ELUTION; POLAR; TETRODOTOXIN; SEPARATION;
D O I
10.1016/j.jhazmat.2020.122318
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The measurement of trace hydrophilic toxins in complex aqueous-rich matrices is a daunting challenge. To address this analytical bottleneck, pulse diffusion focusing (PDF), a novel sample injection technique for hydrophilic interaction chromatography-tandem mass spectrometry (HILIC-MS/MS), was developed. Theoretical and experimental investigations of the mechanism and key parameters revealed that the pulse-injection approach, assisted by solvent diffusion, efficiently solved the volume overload problem. This milliliter-level-injection HILIC-MS/MS technique was reported for the first time herein, and provided a significant enhancement in sensitivity compared to the conventional injection method, in addition to being an efficient approach to address the solvent incompatibility of off-line sample preparation and HILIC. The automated PDF-HILIC-MS/MS system was obtained by slightly modifying a commercial LC-MS/MS instrument in an easy and economical manner. The efficiency of the system was demonstrated through the detection of trace tetrodotoxin contents in plasma and urine samples. Low limits of detection (i.e., 0.65 and 2.2 ng.mL(-1)) were obtained using the simplified sample preparation method. The recoveries were in the range 91-113.3 % with infra-day and inter-day precisions of <= 9.6 %. Further experimental results proved the method to be versatile for various hydrophilic toxins.
引用
收藏
页数:10
相关论文
共 51 条
[21]   Hydrophilic interaction chromatography [J].
Gama, Mariana Roberto ;
da Costa Silva, Raquel Gomes ;
Collins, Carol H. ;
Bottoli, Carla B. G. .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2012, 37 :48-60
[22]   "Dilute-and-inject" multi-target screening assay for highly polar doping agents using hydrophilic interaction liquid chromatography high resolution/high accuracy mass spectrometry for sports drug testing [J].
Goergens, Christian ;
Guddat, Sven ;
Orlovius, Anne-Katrin ;
Sigmund, Gerd ;
Thomas, Andreas ;
Thevis, Mario ;
Schaenzer, Wilhelm .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (18) :5365-5379
[23]   Determination of melamine in milk powder using zwitterionic HILIC stationary phase with UV detection [J].
Ihunegbo, Felicia Nkem ;
Tesfalidet, Solomon ;
Jiang, Wen .
JOURNAL OF SEPARATION SCIENCE, 2010, 33 (6-7) :988-995
[24]   Gradient elution in normal-phase high-performance liquid chromatographic systems [J].
Jandera, P .
JOURNAL OF CHROMATOGRAPHY A, 2002, 965 (1-2) :239-261
[25]   Recent advances in stationary phases and understanding of retention in hydrophilic interaction chromatography. A review [J].
Jandera, Pavel ;
Janas, Petr .
ANALYTICA CHIMICA ACTA, 2017, 967 :12-32
[26]   Hydrophilic interaction chromatography coupled to tandem mass spectrometry as a method for simultaneous determination of guanidinoacetate and creatine [J].
Jovanov, Pavle ;
Vranes, Milan ;
Sakac, Marijana ;
Gadzuric, Slobodan ;
Panic, Jovana ;
Maric, Aleksandar ;
Ostojic, Sergej .
ANALYTICA CHIMICA ACTA, 2018, 1028 :96-103
[27]   Fabric phase sorptive extraction for the fast isolation of sulfonamides residues from raw milk followed by high performance liquid chromatography with ultraviolet detection [J].
Karageorgou, Eftychia ;
Manousi, Natalia ;
Samanidou, Victoria ;
Kabir, Abuzar ;
Furton, Kenneth G. .
FOOD CHEMISTRY, 2016, 196 :428-436
[28]   Analysis of atropine, its degradation products and related substances of natural origin by means of reversed-phase high-performance liquid chromatography [J].
Kirchhoff, C ;
Bitar, Y ;
Ebel, S ;
Holzgrabe, U .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1046 (1-2) :115-120
[29]   Effect of eluent on the ionization process in liquid chromatography-mass spectrometry [J].
Kostiainen, Risto ;
Kauppila, Tiina J. .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (04) :685-699
[30]   Analytical Challenges: Determination of Tetrodotoxin in Human Urine and Plasma by LC-MS/MS [J].
Leung, Kelvin Sze-Yin ;
Fong, Bonnie Mei-Wah ;
Tsoi, Yeuk-Ki .
MARINE DRUGS, 2011, 9 (11) :2291-2303