Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners

被引:27
|
作者
Bowden, AE [1 ]
Kurtz, SM
Edidin, AA
机构
[1] Drexel Univ, Implant Res Ctr, Philadelphia, PA 19104 USA
[2] Exponent Inc, Philadelphia, PA USA
关键词
polyethylene; wear; retrieval; microCT; registration;
D O I
10.1002/jbm.b.30318
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, a novel micro-CT-based technique for evaluating wear in retrieved acetabular liners was introduced and validated. Six UHMWPE acetabular components ranging in implantation time from 2.7 to 14.4 years were collected and evaluated with the use of a high-resolution micro-CT scanner. The components were scanned with a uniform volumetric resolution of 74 microns (16-bit precision) with the use of a 1024 X 1024 in-plane image matrix. Manual rigid 3D image registration of the interior hemispherical portion of the acetabular cup with geometric primitives by trained observers allowed for isolation, visualization, and measurement of the wear volume. Results for these six components indicated an average wear rate of 65 mm(3)/year. Overall scanner error was quantified gravimetrically and associated with a maximum uncertainty of 0.6%. Intra- and interobserver uncertainty analysis showed the method to be both accurate and repeatable. (c) 2005 Wiley Periodicals, Inc.
引用
收藏
页码:205 / 209
页数:5
相关论文
共 42 条
  • [1] Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants
    Adam, P
    Farizon, F
    Fessy, MH
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2005, 91 (07): : 627 - 636
  • [2] Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners
    Knowlton, Christopher B.
    Bhutani, Priyanka
    Wimmer, Markus A.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2017, 105 (07) : 2053 - 2059
  • [3] Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE
    Schroder, David T.
    Kelly, Natalie H.
    Wright, Timothy M.
    Parks, Michael L.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) : 387 - 394
  • [4] Novel Heterodyne Fringe Projection Technique for Measuring Volumetric Wear in Acetabular UHMWPE Retrievals: A Pilot Study
    Medel, Francisco J.
    Povar, Maria
    Santolaria, Jorge
    BEYOND THE IMPLANT: RETRIEVAL ANALYSIS METHODS FOR IMPLANT SURVEILLANCE, 2018, 1606 : 93 - 105
  • [5] Development of the 3D volumetric micro-CT scanner for preclinical animals
    Kim, Kyong-Woo
    Kim, Kyu-Gyeom
    Kim, Jae-Hee
    Min, Jong-Hwan
    Lee, Hee-Sin
    Lee, Joon-Whoan
    3D RESEARCH, 2011, 2 (02) : 1 - 6
  • [6] A COMPARISON OF OPTICAL AND MICRO-CT TECHNIQUE FOR POROSITY ANALYSIS OF PORITES
    Listiyowati, Lina Nur
    RISET GEOLOGI DAN PERTAMBANGAN, 2018, 28 (01): : 91 - 100
  • [7] Coronary bifurcation stent morphology in dual-source CT: validation with micro-CT
    Eom, Hye-Joung
    Yang, Dong Hyun
    Kim, Young-Hak
    Roh, Jae-Hyung
    Kweon, Jihoon
    Kang, Soo-Jin
    Kim, Namkug
    Kang, Joon-Won
    Lim, Tae-Hwan
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2016, 32 (11) : 1659 - 1665
  • [8] Validation of finite element models of liver tissue using Micro-CT
    Shi, Hongjian
    Farag, Aly A.
    Fahmi, Rachid
    Chen, Dongqing
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (03) : 978 - 984
  • [9] Coronary bifurcation stent morphology in dual-source CT: validation with micro-CT
    Hye-Joung Eom
    Dong Hyun Yang
    Young-Hak Kim
    Jae-Hyung Roh
    Jihoon Kweon
    Soo-Jin Kang
    Namkug Kim
    Joon-Won Kang
    Tae-Hwan Lim
    The International Journal of Cardiovascular Imaging, 2016, 32 : 1659 - 1665
  • [10] Volumetric change and gap formation in class V composite restorations: a micro-CT analysis
    Oglakci, Burcu
    Halacoglu, Derya Merve
    Ozduman, Zumrut Ceren
    Dalkilic, Evrim Eliguzeloglu
    JOURNAL OF ADHESION SCIENCE AND TECHNOLOGY, 2020, 34 (24) : 2725 - 2742