Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners

被引:27
作者
Bowden, AE [1 ]
Kurtz, SM
Edidin, AA
机构
[1] Drexel Univ, Implant Res Ctr, Philadelphia, PA 19104 USA
[2] Exponent Inc, Philadelphia, PA USA
关键词
polyethylene; wear; retrieval; microCT; registration;
D O I
10.1002/jbm.b.30318
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, a novel micro-CT-based technique for evaluating wear in retrieved acetabular liners was introduced and validated. Six UHMWPE acetabular components ranging in implantation time from 2.7 to 14.4 years were collected and evaluated with the use of a high-resolution micro-CT scanner. The components were scanned with a uniform volumetric resolution of 74 microns (16-bit precision) with the use of a 1024 X 1024 in-plane image matrix. Manual rigid 3D image registration of the interior hemispherical portion of the acetabular cup with geometric primitives by trained observers allowed for isolation, visualization, and measurement of the wear volume. Results for these six components indicated an average wear rate of 65 mm(3)/year. Overall scanner error was quantified gravimetrically and associated with a maximum uncertainty of 0.6%. Intra- and interobserver uncertainty analysis showed the method to be both accurate and repeatable. (c) 2005 Wiley Periodicals, Inc.
引用
收藏
页码:205 / 209
页数:5
相关论文
共 28 条
[1]   The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: role of the third-body wear process [J].
Affatato, S ;
Bersaglia, G ;
Foltran, I ;
Taddei, P ;
Fini, G ;
Toni, A .
BIOMATERIALS, 2002, 23 (24) :4839-4846
[2]   Radiographic wear measurements in a cementless metal-backed modular cobalt-chromium acetabular component [J].
Barrack, RL ;
Lavernia, C ;
Szuszczewicz, ES ;
Sawhney, J .
JOURNAL OF ARTHROPLASTY, 2001, 16 (07) :820-828
[3]   Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners [J].
Bradford, L ;
Baker, DA ;
Graham, J ;
Chawan, A ;
Ries, MD ;
Pruitt, LA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) :1271-1282
[4]   Comparison of two digital radio stereometric analysis methods in the determination of femoral head penetration in a total hip replacement phantom [J].
Bragdon, CR ;
Estok, DM ;
Malchau, H ;
Kärrholm, J ;
Yuan, XH ;
Bourne, R ;
Veldhoven, J ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (03) :659-664
[5]  
CHARNLEY J, 1975, CLIN ORTHOP RELAT R, P170
[6]   Evaluation of contemporary software methods used to quantify polyethylene wear after total hip arthroplasty [J].
Collier, MB ;
Kraay, MJ ;
Rimnac, CM ;
Goldberg, VM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (12) :2410-2418
[7]   Polyethylene cross-linking by two different methods reduces acetabular liner wear in a hip joint wear simulator [J].
D'Lima, DD ;
Hermida, JC ;
Chen, PC ;
Colwell, CW .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2003, 21 (05) :761-766
[8]  
DEVANE PA, 1995, CLIN ORTHOP RELAT R, P317
[9]   Comparison of polyethylene wear with femoral heads of 22 mm and 32 mm - A prospective randomised study [J].
Eggli, S ;
z'Brun, S ;
Gerber, C ;
Ganz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (03) :447-451
[10]  
Hall R M, 1995, Proc Inst Mech Eng H, V209, P233, DOI 10.1243/PIME_PROC_1995_209_350_02