Liquid paraffin as new dilution medium for the analysis of high boiling point residual solvents with static headspace-gas chromatography

被引:19
作者
D'Autry, Ward [1 ]
Zheng, Chao [1 ]
Bugalama, John [1 ]
Wolfs, Kris [1 ]
Hoogmartens, Jos [1 ]
Adams, Erwin [1 ]
Wang, Bochu [1 ]
Van Schepdael, Ann [1 ]
机构
[1] Katholieke Univ Leuven, Lab Pharmaceut Anal, Fac Pharmaceut Sci, B-3000 Louvain, Belgium
关键词
Static headspace; Gas chromatography; Residual solvents; Dilution media; Validation; PHARMACEUTICAL PRODUCTS; PHASE MICROEXTRACTION; IONIC LIQUIDS; LIMITS; VAPOR;
D O I
10.1016/j.jpba.2011.04.004
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Residual solvents are volatile organic compounds which can be present in pharmaceutical substances. A generic static headspace-gas chromatography analysis method for the identification and control of residual solvents is described in the European Pharmacopoeia. Although this method is proved to be suitable for the majority of samples and residual solvents, the method may lack sensitivity for high boiling point residual solvents such as N,N-dimethylformamide, N,N-dimethylacetamide, dimethyl sulfoxide and benzyl alcohol. In this study, liquid paraffin was investigated as new dilution medium for the analysis of these residual solvents. The headspace-gas chromatography method was developed and optimized taking the official Pharmacopoeia method as a starting point. The optimized method was validated according to ICH criteria. It was found that the detection limits were below 1 mu g/vial for each compound, indicating a drastically increased sensitivity compared to the Pharmacopoeia method, which failed to detect the compounds at their respective limit concentrations. Linearity was evaluated based on the R(2) values, which were above 0.997 for all compounds, and inspection of residual plots. Instrument and method precision were examined by calculating the relative standard deviations (RSD) of repeated analyses within the linearity and accuracy experiments, respectively. It was found that all RSD values were below 10%. Accuracy was checked by a recovery experiment at three different levels. Mean recovery values were all in the range 95-105%. Finally, the optimized method was applied to residual DMSO analysis in four different Kollicoat (R) sample batches. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:1017 / 1023
页数:7
相关论文
共 35 条
  • [11] Static headspace gas chromatographic analysis of the residual solvents in gel extrusion module tablet formulations
    Li, Z
    Han, YH
    Martin, GP
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 28 (3-4) : 673 - 682
  • [12] A generic static headspace gas chromatography method for determination of residual solvents in drug substance
    Cheng, Chang
    Liu, Shaorong
    Mueller, Bradford J.
    Yan, Zimeng
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (41) : 6413 - 6421
  • [13] Matrix media selection for the determination of residual solvents in pharmaceuticals by static headspace gas chromatography
    Urakami, K
    Higashi, A
    Umemoto, K
    Godo, M
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1057 (1-2) : 203 - 210
  • [14] Use of ionic liquids as headspace gas chromatography diluents for the analysis of residual solvents in pharmaceuticals
    Nacham, Omprakash
    Ho, Tien D.
    Anderson, Jared L.
    Webster, Gregory K.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2017, 145 : 879 - 886
  • [15] Determination of Residual Solvents in Pharmaceuticals by Static Headspace Gas Chromatography Using Natural Deep Eutectic Solvents as Mediums: A Partition Coefficients Study
    Zhang, Wei
    Fang, Sheng
    Liang, Xianrui
    CHROMATOGRAPHIA, 2019, 82 (10) : 1523 - 1529
  • [16] Full evaporation headspace gas chromatography for sensitive determination of high boiling point volatile organic compounds in low boiling matrices
    Kialengila, Didi Mana
    Wolfs, Kris
    Bugalama, John
    Van Schepdael, Ann
    Adams, Erwin
    JOURNAL OF CHROMATOGRAPHY A, 2013, 1315 : 167 - 175
  • [17] Determination of Residual Solvents in Pharmaceuticals by Static Headspace Gas Chromatography Using Natural Deep Eutectic Solvents as Mediums: A Partition Coefficients Study
    Wei Zhang
    Sheng Fang
    Xianrui Liang
    Chromatographia, 2019, 82 : 1523 - 1529
  • [18] Determination of the residual ethanol in hydroalcoholic sealed hard gelatin capsules by static headspace gas chromatography with immiscible binary solvents
    Li, Jingyi
    Shao, Shan
    Solorzano, Mauro
    Allmaier, Gary J.
    Kurtulik, Paul T.
    JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (15) : 3328 - 3336
  • [19] Simultaneous Determination of Twelve Kinds of Residual Solvents in A New Drug CBT108 By Headspace Gas Chromatography
    Hu Mei-Chen
    Huang Jin-Wen
    Huang Lei-Lei
    Xie Da
    Zhu Yang-Bin
    Wang Lin
    Wang Tian-Xing
    Wu Fan-Hong
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2018, 46 (05) : 710 - 715
  • [20] Rapid analysis of the volatile compounds in the rhizomes of Rhodiold sachalinensis and Rhodiola sacra by static headspace-gas chromatography-tandem mass spectrometry
    Wang, HY
    Guo, YL
    ANALYTICAL LETTERS, 2004, 37 (10) : 2151 - 2161