The forgotten fragment: additional lesser tuberosity fixation of 4-part proximal humeral fractures - a biomechanical investigation

被引:6
作者
Katthagen, J. Christoph [1 ]
Michel, Philipp [1 ]
Raschke, Michael J. [1 ]
Sussiek, Julia [1 ]
Frank, Andre [1 ]
Wermers, Jens [1 ]
Dyrna, Felix [1 ]
Schliemann, Benedikt [1 ]
机构
[1] Univ Hosp Munster, Dept Trauma Hand & Reconstruct Surg, Waldeyer Str 1, D-48149 Munster, Germany
关键词
Proximal humerus; 4-part fracture; double plate; fracture; biomechanics; LOCKING PLATE FIXATION; BONE; AUGMENTATION; FORCES; SCREWS;
D O I
10.1016/j.jse.2021.04.041
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Fractures of the proximal humerus are common. The most frequent surgical treatment option is open reduction and locking plate fixation. Multifragmentary fractures, including 3- and 4-part fractures, are especially challenging to teat because they correlate with an increased risk of fixation failure. In the past, several mechanisms of additional fixation were investigated, but none directly addressed the lesser tuberosity (LT). The goal of this study was to investigate the biomechanical impact of additional anterior fracture fixation in lateral locked plating (LLP) of 4-part proximal humeral fractures (PHFs). Methods: Twenty-seven fresh frozen human shoulder specimens (mean age, 80 years) with intact rotator cuffs (RCs) were randomized into 4 groups: 3-part PELF with LLP and RC cerclage (n = 6); 4-part PHF with LLP and RC cerclage as standard of care (n = 7); 4-part PHF with LLP, RC cerclage, and 2 anterior 3.5-mm cortical screws (n = 7); and 4-part PHF with LLP, RC cerclage, and additional anterior one-third tubular plate (additional anterior plating [AAP], n = 7). Static load of the RC was simulated with weights. A force-controlled cyclic loading test was performed with a servo-hydraulic testing machine, followed by load-to-failure testing. An optical motion capture system recorded humeral head range of motion. Results: LLP of a 4-part PHF showed more humeral head motion than LLP of a 3-part PIT without fracture of the LT (P < .001). Fixing the LT to the humeral head with two 35-mm screws significantly reduced humeral head motion compared with LLP with RC cerclage alone (P < .006). Using AAP significantly increased the construct stiffness compared with the standard of care (P = .03). Conclusion: LLP of a 4-part PHF is biomechanically less stable than LLP of a 3-part PHF without fracture of the LT. Additional screw fixation of the LT in 4-part PHFs improves stability compared with LLP alone. In case of metaphyseal comminution, AAP is favorable from a biomechanical perspective. (C) 2021 Journal of Shoulder and Elbow Surgery Board of Trustees. All rights reserved.
引用
收藏
页码:2852 / 2861
页数:10
相关论文
共 39 条
  • [1] Biomechanical Contribution of Tension-Reducing Rotator Cuff Sutures in 3-Part Proximal Humerus Fractures
    Arvesen, John E.
    Gill, Stephen W.
    Sinatra, Philip M.
    Eng, Michael
    Bledsoe, Gary
    Kaar, Scott G.
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2016, 30 (08) : E262 - E266
  • [2] Locking plate fixation of proximal humerus fractures in patients older than 60 years continues to be associated with a high complication rate
    Barlow, Jonathan D.
    Logli, Anthony L.
    Steinmann, Scott P.
    Sems, Stephen A.
    Cross, William W.
    Yuan, Brandon J.
    Torchia, Michael E.
    Sanchez-Sotelo, Joaquin
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2020, 29 (08) : 1689 - 1694
  • [3] Augmentation of plate osteosynthesis for proximal humeral fractures: a systematic review of current biomechanical and clinical studies
    Biermann, Niklas
    Prall, Wolf Christian
    Boecker, Wolfgang
    Mayr, Hermann Otto
    Haasters, Florian
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2019, 139 (08) : 1075 - 1099
  • [4] "90/90" Plating of proximal humerus fracturea technical note
    Cassidy, John Tristan
    Coveney, Eamonn
    Molony, Diarmoud
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (1)
  • [5] Technical tips: Dualplate fixation technique for comminuted proximal humerus fractures
    Choi, Sungwook
    Kang, Hyunseong
    Bang, Hyeongsig
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2014, 45 (08): : 1280 - 1282
  • [6] Anatomical and biomechanical evaluation of an intramedullary nail for fractures of proximal humerus fractures based on tuberosity fixation
    Clavert, Ph.
    Hatzidalds, A.
    Boileau, P.
    [J]. CLINICAL BIOMECHANICS, 2016, 32 : 108 - 112
  • [7] BIOMECHANICS OF FOUR TECHNIQUES FOR FIXATION OF THE FOUR-PART HUMERAL HEAD FRACTURE
    da Graca, Elpidio
    Okubo, Rodrigo
    Shimano, Antonio Carlos
    Mazzer, Nilton
    Barbieri, Claudio Henrique
    [J]. ACTA ORTOPEDICA BRASILEIRA, 2013, 21 (01): : 34 - 39
  • [8] Biomechanical behavior of three types of fixation in the two-part proximal humerus fracture without medial cortical support
    di Tullio, Paulo Ottoni
    Giordano, Vincenzo
    Souto, Eder
    Assed, Hugo
    Chequer, Joao Paulo
    Belangero, William
    Mariolani, Jose Ricardo L.
    Koch, Hilton A.
    [J]. PLOS ONE, 2019, 14 (07):
  • [9] Foruria AM, 2015, J ORTHOP TRAUMA, V29, pE24, DOI 10.1097/BOT.0000000000000244
  • [10] Surgical and Non-Surgical Treatment for 3-Part and 4-Part Fractures of the Proximal Humerus: A Systematic Review of Overlapping Meta-Analyses
    Fu, Bai-sheng
    Jia, Hong-lei
    Zhou, Dong-sheng
    Liu, Fan-xiao
    [J]. ORTHOPAEDIC SURGERY, 2019, 11 (03) : 356 - 365