Stable isotope-dilution liquid chromatography/tandem mass spectrometry method for determination of thyroxine in saliva

被引:16
作者
Higashi, Tatsuya [1 ,2 ]
Ichikawa, Takuya [1 ,2 ]
Shimizu, Chikara [3 ]
Nagai, So [4 ]
Inagaki, Shinsuke [1 ,2 ]
Min, Jun Zhe [1 ,2 ]
Chiba, Hitoshi [5 ]
Ikegawa, Shigeo [6 ]
Toyo'oka, Toshimasa [1 ,2 ]
机构
[1] Univ Shizuoka, Sch Pharmaceut Sci, Lab Analyt & Bioanalyt Chem, Suruga Ku, Shizuoka 4228526, Japan
[2] Univ Shizuoka, Global COE Program, Suruga Ku, Shizuoka 4228526, Japan
[3] Hokkaido Univ Hosp, Div Lab & Transfus Med, Kita Ku, Sapporo, Hokkaido 0608648, Japan
[4] Hokkaido Univ, Grad Sch Med, Dept Med 2, Kita Ku, Sapporo, Hokkaido 0608638, Japan
[5] Hokkaido Univ, Grad Sch Hlth Sci, Fac Hlth Sci, Kita Ku, Sapporo, Hokkaido 0600812, Japan
[6] Kinki Univ, Fac Pharmaceut Sci, Lab Bioanalyt Sci, Higashiosaka, Osaka 5778502, Japan
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2011年 / 879卷 / 13-14期
基金
日本学术振兴会;
关键词
Thyroxine; Saliva; Liquid chromatography/electrospray; ionization-tandem mass spectrometry; Isotope dilution; Graves disease; HUMAN SERUM; HORMONE; TOOL;
D O I
10.1016/j.jchromb.2011.02.048
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A liquid chromatography/electrospray ionization-tandem mass spectrometry (LC/ESI-MS/MS) method for the determination of thyroxine (T-4) in human saliva has been developed and validated. The saliva was deproteinized with methanol, purified using a Strata-X (TM) cartridge, and subjected to LC/ESI-MS/MS. Quantification was based on selected reaction monitoring, and [C-13(6)]-T-4 was used as the internal standard. This method allowed the reproducible (intra- and inter-assay relative standard deviations, < 4.8%) and accurate (analytical recovery, 96.5-99.6%) quantification of the salivary T-4 using a 400 mu l sample, and the limit of quantification was 25.0 pg/ml. A preliminary study using the developed method found that there is a diagnosable difference in the salivary T-4 concentration between the euthyroid subjects and the patients with Graves disease. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:1013 / 1017
页数:5
相关论文
共 15 条
  • [11] Simultaneous determination of salivary testosterone and dehydroepiandrosterone using LC-MS/MS: Method development and evaluation of applicability for diagnosis and medication for late-onset hypogonadism
    Shibayama, Yujin
    Higashi, Tatsuya
    Shimada, Kazutake
    Odani, Akira
    Mizokami, Atsushi
    Konaka, Hiroyuki
    Koh, Eitetsu
    Namiki, Mikio
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (25): : 2615 - 2623
  • [12] Thyroid hormone testing by tandem mass spectrometry
    Soldin, Offie P.
    Soldin, Steven J.
    [J]. CLINICAL BIOCHEMISTRY, 2011, 44 (01) : 89 - 94
  • [13] The measurement of free thyroxine by isotope dilution tandem mass spectrometry
    Soldin, SJ
    Soukhova, N
    Janicic, N
    Jonklaas, J
    Soldin, OP
    [J]. CLINICA CHIMICA ACTA, 2005, 358 (1-2) : 113 - 118
  • [14] Development and evaluation of a reference measurement procedure for the determination of total 3,3′,5-trilodothyronine in human serum using isotope-dilution liquid chromatography-tandem mass spectrometry
    Tai, SSC
    Bunk, DM
    White, EV
    Welch, MJ
    [J]. ANALYTICAL CHEMISTRY, 2004, 76 (17) : 5092 - 5096
  • [15] Analysis of thyroid hormones in serum by liquid chromatography-tandem mass spectrometry
    Wang, Dongli
    Stapleton, Heather M.