A clinical comparison of variable-damping and mechanically passive prosthetic knee devices

被引:200
作者
Johansson, JL
Sherrill, DM
Riley, PO
Bonato, P
Herr, H
机构
[1] MIT, Media Lab, Cambridge, MA 02139 USA
[2] Harvard Univ, Spaulding Rehabil Hosp, Sch Med, Dept Phys Med & Rehabil, Boston, MA 02115 USA
[3] Harvard Univ, MIT, Div Hlth Sci & Technol, Cambridge, MA 02138 USA
[4] Univ Virginia, Dept Phys Med & Rehabil, Charlottesville, VA USA
关键词
prosthesis; hydraulic knee; magnetorheological knee; walking metabolism;
D O I
10.1097/01.phm.0000174665.74933.0b
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objective: Although variable-damping knee prostheses offer some improvements over mechanically passive prostheses to transfemoral amputees, there is insufficient evidence that such prostheses provide advantages at self-selected walking speeds. In this investigation, we address this question by comparing two variable-damping knees, the hydraulic-based Otto Bock C-leg and the magnetorheological-based Ossur Rheo, with the mechanically passive, hydraulic-based Mauch SNS. Design: For each prosthesis, metabolic data were collected on eight unilateral amputees walking at self-selected speeds across an indoor track. Furthermore, kinetic, kinematic, and electromyographic data were collected while walking at self-selected speeds across a 10-m walkway in a laboratory. Results: When using the Rheo, metabolic rate decreases by 5% compared with the Mauch and by 3% compared with the C-leg. Furthermore, for the C-leg and Rheo knee devices, we observe biomechanical advantages over the mechanically passive Mauch. These advantages include an enhanced smoothness of gait, a decrease in hip work production, a lower peak hip flexion moment at terminal stance, and a reduction in peak hip power generation at toe-off. Conclusion: The results of this study indicate that variable-clamping knee prostheses offer advantages over mechanically passive designs for unilateral transfemoral amputees walking at self-selected ambulatory speeds, and the results further suggest that a magnetorheological-based system may have advantages over hydraulic-based designs.
引用
收藏
页码:563 / 575
页数:13
相关论文
共 27 条
[1]  
[Anonymous], 1999, MED ORTHOP TECHNOL
[2]  
[Anonymous], 1996, CLIN REHABIL
[3]   Advances in wearable technology and applications in physical medicine and rehabilitation [J].
Bonato P. .
Journal of NeuroEngineering and Rehabilitation, 2 (1)
[4]   Energy cost of walking: Comparison of ''Intelligent prosthesis'' with conventional mechanism [J].
Buckley, JG ;
Spence, WD ;
Solomonidis, SE .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1997, 78 (03) :330-333
[5]   Validity of the prosthetic activity monitor to assess the duration and spatio-temporal characteristics of prosthetic walking [J].
Bussmann, JBJ ;
Culhane, KM ;
Horemans, HLD ;
Lyons, GM ;
Stam, HJ .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2004, 12 (04) :379-386
[6]  
Datta D, 1998, PROSTHET ORTHOT INT, V22, P129
[7]  
Dawson B., 2001, BASIC CLIN BIOSTATIS, V3rd
[8]  
FLOWERS WC, 1972, THESIS MIT
[9]   FEASIBILITY OF AN ACTIVE CONTROL SCHEME FOR ABOVE KNEE PROSTHESES [J].
GRIMES, DL ;
FLOWERS, WC ;
DONATH, M .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1977, 99 (04) :215-221
[10]   User-adaptive control of a magnetorheological prosthetic knee [J].
Herr, H ;
Wilkenfeld, A .
INDUSTRIAL ROBOT-AN INTERNATIONAL JOURNAL, 2003, 30 (01) :42-55