Optimal design of a flexure hinge-based XYZ atomic force microscopy scanner for minimizing Abbe errors -: art. no. 073706

被引:70
作者
Kim, D [1 ]
Kang, D [1 ]
Shim, J [1 ]
Song, I [1 ]
Gweon, D [1 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Engn Mech, Taejon 305701, South Korea
关键词
D O I
10.1063/1.1978827
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To establish of standard technique of nanolength measurement in a two-dimensional plane, a new (AFM) system has been designed. In this system, measurement uncertainty is dominantly affected by the Abbe error of the XYZ scanning stage. No linear stage is perfectly straight; in other words, every scanning stage is subject to tilting, pitch, and yaw motion. In this article, an AFM system with minimum offset of XYZ sensing is designed. And, the XYZ scanning stage is designed to minimize the rotation angle because Abbe errors occur through the multiply of offset and rotation angle. For XY stage, optimal design is performed to minimize the rotation angle by maximizing the stiffness ratio of motion direction to the parasitic motion direction of each stage. For the Z stage, the optimal design of maximizing the first-resonant frequency is performed. When the resonant frequency increases, the scan speed is improved, thereby reducing errors caused by sensor drift. This article describes the procedures of selecting parameters for the optimal design. The full range of the XYZ scanner is 100 mu m x 100 mu m x 10 mu m. Based on the solution of the optimization problem, the XYZ scanner is fabricated. And tilting, pitch, and yaw motion are measured by autocollimator to evaluate the performance of XY stage. (c) 2005 American Institute of Physics.
引用
收藏
页数:7
相关论文
共 9 条
[1]  
Bienias M, 1997, SURF INTERFACE ANAL, V25, P606, DOI 10.1002/(SICI)1096-9918(199706)25:7/8<606::AID-SIA285>3.0.CO
[2]  
2-S
[3]  
DIXON RG, 1999, P SOC PHOTO-OPT INS, V3677, P20
[4]   Calibration of step heights and roughness measurements with atomic force microscopes [J].
Garnaes, J ;
Kofod, N ;
Kühle, A ;
Nielsen, C ;
Dirscherl, K ;
Blunt, L .
PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2003, 27 (01) :91-98
[5]   Real-time, interferometrically measuring atomic force microscope for direct calibration of standards [J].
Gonda, S ;
Doi, T ;
Kurosawa, T ;
Tanimura, Y ;
Hisata, N ;
Yamagishi, T ;
Fujimoto, H ;
Yukawa, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (08) :3362-3368
[6]   Calibrated scanning force microscope with capabilities in the subnanometre range [J].
Hasche, K ;
Herrmann, K ;
Mirandé, W ;
Seemann, R ;
Vitushkin, L ;
Xu, M ;
Yu, G .
SURFACE AND INTERFACE ANALYSIS, 2002, 33 (02) :71-74
[7]   Long-range AFM profiler used for accurate pitch measurements [J].
Meli, F ;
Thalmann, R .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (07) :1087-1092
[8]  
Paros J M., 1965, Mach. Des, V37, P151
[9]   Optimal design of a flexure hinge based XYθ wafer stage [J].
Ryu, JW ;
Gweon, DG ;
Moon, KS .
PRECISION ENGINEERING-JOURNAL OF THE AMERICAN SOCIETY FOR PRECISION ENGINEERING, 1997, 21 (01) :18-28