Objective: The aim of this study was to investigate whether a real-time noise reduction algorithm provided speech perception benefit for Cochlear (TM) Nucleus (R) cochlear implant recipients in the laboratory. Design: The noise reduction algorithm attenuated masker-dominated channels. It estimated the signal-to-noise ratio of each channel on a short-term basis from a single microphone input, using a recursive minimum statistics method. In this clinical evaluation, the algorithm was implemented in two programs (noise reduction programs 1 [NR1] and 2 [NR2]), which differed in their level of noise reduction. These programs used advanced combination encoder (ACE T) channel selection and were compared with ACE without noise reduction in 13 experienced cochlear implant subjects. An adaptive speech reception threshold (SRT) test provided the signal-to-noise ratio for 50% sentence intelligibility in three different types of noises: speech-weighted, cocktail party, and street-side city noise. Results: In all three noise types, mean SRTs for both NR programs were significantly better than those for ACE. The greatest improvement occurred for speech-weighted noise; the SRT benefit over ACE was 1.77 dB for NR1 and 2.14 dB for NR2. There were no significant differences in speech perception scores between the two NR programs. Subjects reported no degradation in sound quality with the experimental programs. Conclusions: The noise reduction algorithm was successful in improving sentence perception in speech-weighted noise, as well as in more dynamic types of background noise. The algorithm is currently being trialed in a behind-the-ear processor for take-home use.
引用
收藏
页码:382 / 390
页数:9
相关论文
共 50 条
  • [31] Improving Localization and Speech Reception in Noise for Bilateral Cochlear Implant Recipients
    Potts, Wendy B.
    Ramanna, Lakshmish
    Perry, Trevor
    Long, Christopher J.
    TRENDS IN HEARING, 2019, 23
  • [32] PREDICTING THE EFFECT OF AGC ON SPEECH INTELLIGIBILITY OF COCHLEAR IMPLANT RECIPIENTS IN NOISE
    Khing, Phyu P.
    Ambikairajah, Eliathamby
    Swanson, Brett A.
    2013 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING (ICASSP), 2013, : 8061 - 8065
  • [33] Spatial dissociation of changes of level and signal-to-noise ratio in auditory cortex for tones in noise
    Ernst, Stephan M. A.
    Verhey, Jesko L.
    Uppenkamp, Stefan
    NEUROIMAGE, 2008, 43 (02) : 321 - 328
  • [34] Signal-to-Noise Ratio Analysis Based on Different Space Remote Sensing Instruments
    Zhang, Lei
    Li, Bo
    Li, Hanshuang
    Gu, Guochao
    Wang, Xiaoxu
    IEEE PHOTONICS JOURNAL, 2024, 16 (02): : 1 - 9
  • [35] Speech Understanding Performance of Cochlear Implant Subjects Using Time-Frequency Masking-Based Noise Reduction
    Qazi, Obaid Ur Rehman
    van Dijk, Bas
    Moonen, Marc
    Wouters, Jan
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (05) : 1364 - 1373
  • [36] Signal-to-noise ratio for source determination and for a comodulated masker in goldfish, Carassius auratus
    Fay, Richard R.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2011, 129 (05) : 3367 - 3372
  • [37] Influence of wavefront distortion on the measurement of pulse signal-to-noise ratio
    Xing, Dingding
    Yuan, Suochao
    Kou, Jingwei
    Da, Zhengshang
    OPTICS COMMUNICATIONS, 2024, 554
  • [38] Microrobotic Swarms for Intracellular Measurement with Enhanced Signal-to-Noise Ratio
    Wang, Xian
    Wang, Tiancong
    Chen, Xin
    Law, Junhui
    Shan, Guanqiao
    Tang, Wentian
    Gong, Zheyuan
    Pan, Peng
    Liu, Xinyu
    Yu, Jiangfan
    Ru, Changhai
    Huang, Xi
    Sun, Yu
    ACS NANO, 2022, 16 (07) : 10824 - 10839
  • [39] Optimal Linear Control for Channels with Signal-to-Noise Ratio Constraints
    Johannesson, Erik
    Rantzer, Anders
    Bernhardsson, Bo
    2011 AMERICAN CONTROL CONFERENCE, 2011, : 521 - 526
  • [40] Improved resolution and signal-to-noise ratio in MRI via enhanced signal digitization
    Elliott, MA
    Insko, EK
    Greenman, RL
    Leigh, JS
    JOURNAL OF MAGNETIC RESONANCE, 1998, 130 (02) : 300 - 304