Clinical Evaluation of Signal-to-Noise Ratio-Based Noise Reduction in Nucleus® Cochlear Implant Recipients

被引:86
|
作者
Dawson, Pam W. [1 ,2 ]
Mauger, Stefan J. [1 ]
Hersbach, Adam A. [2 ]
机构
[1] HEARing Cooperat Res Ctr, Melbourne, Vic, Australia
[2] Cochlear Ltd, Melbourne, Vic, Australia
关键词
TEST LISN-S; SPEECH RECOGNITION; BACKGROUND-NOISE; DYNAMIC-RANGE; INTELLIGIBILITY; ENHANCEMENT; SUPPRESSION; SENTENCES; HEARING; SYSTEM;
D O I
10.1097/AUD.0b013e318201c200
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: The aim of this study was to investigate whether a real-time noise reduction algorithm provided speech perception benefit for Cochlear (TM) Nucleus (R) cochlear implant recipients in the laboratory. Design: The noise reduction algorithm attenuated masker-dominated channels. It estimated the signal-to-noise ratio of each channel on a short-term basis from a single microphone input, using a recursive minimum statistics method. In this clinical evaluation, the algorithm was implemented in two programs (noise reduction programs 1 [NR1] and 2 [NR2]), which differed in their level of noise reduction. These programs used advanced combination encoder (ACE T) channel selection and were compared with ACE without noise reduction in 13 experienced cochlear implant subjects. An adaptive speech reception threshold (SRT) test provided the signal-to-noise ratio for 50% sentence intelligibility in three different types of noises: speech-weighted, cocktail party, and street-side city noise. Results: In all three noise types, mean SRTs for both NR programs were significantly better than those for ACE. The greatest improvement occurred for speech-weighted noise; the SRT benefit over ACE was 1.77 dB for NR1 and 2.14 dB for NR2. There were no significant differences in speech perception scores between the two NR programs. Subjects reported no degradation in sound quality with the experimental programs. Conclusions: The noise reduction algorithm was successful in improving sentence perception in speech-weighted noise, as well as in more dynamic types of background noise. The algorithm is currently being trialed in a behind-the-ear processor for take-home use.
引用
收藏
页码:382 / 390
页数:9
相关论文
共 50 条
  • [21] Envelope-Based Multichannel Noise Reduction for Cochlear Implant Applications
    de Souza, Luciana M. X.
    Costa, Marcio H.
    Borges, Renata Coelho
    IEEE-ACM TRANSACTIONS ON AUDIO SPEECH AND LANGUAGE PROCESSING, 2024, 32 : 1873 - 1884
  • [22] Contralateral Routing of Signal Yields Significant Speech in Noise Benefit for Unilateral Cochlear Implant Recipients
    Dwyer, Robert T.
    Kessler, David
    Butera, Iliza M.
    Gifford, Rene H.
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2019, 30 (03) : 235 - 242
  • [23] Measurement and study on signal-to-noise ratio of a spaceborne camera
    Chen, Yuheng
    Zhou, Jiankang
    Chen, Xinhua
    Ji, Yiqun
    Shen, Weimin
    2011 INTERNATIONAL CONFERENCE ON OPTICAL INSTRUMENTS AND TECHNOLOGY: OPTOELECTRONIC IMAGING AND PROCESSING TECHNOLOGY, 2011, 8200
  • [24] Prediction of Individual Cochlear Implant Recipient Speech Perception With the Output Signal to Noise Ratio Metric
    Watkins, Greg D.
    Swanson, Brett A.
    Suaning, Gregg J.
    EAR AND HEARING, 2020, 41 (05) : 1270 - 1281
  • [25] A SUPERVISED SIGNAL-TO-NOISE RATIO ESTIMATION OF SPEECH SIGNALS
    Papadopoulos, Pavlos
    Tsiartas, Andreas
    Gibson, James
    Narayanan, Shrikanth
    2014 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING (ICASSP), 2014,
  • [26] An Investigation of the Effect of AGC Gain on the Output Signal to Noise Ratio in Cochlear Implant Sound Processing
    Watkins, Greg D.
    Swanson, Brett A.
    Suaning, Gregg J.
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 1206 - 1209
  • [27] The Noise Reduction Algorithm May Not Compensate for the Degradation in Output Signal-to-Noise Ratio Caused by Wide Dynamic Range Compression
    Yun, Donghyeon
    Lentz, Jennifer
    Shen, Yi
    AMERICAN JOURNAL OF AUDIOLOGY, 2024, 33 (03) : 793 - 809
  • [28] Research on evaluation index of stray light suppression ability of star sensor based on signal-to-noise ratio
    Liu, Shi
    Zhang, Jierui
    Sun, Gaofei
    Zhang, Guoyu
    Chen, Siwen
    Chen, Jiali
    OPTICS COMMUNICATIONS, 2023, 530
  • [29] Effects of Cognitive Functions on Speech Recognition in Noise in Cochlear Implant Recipients
    Gundogdu, Ogulcan
    Serbetcioglu, Mustafa Buelent
    Kara, Eyyup
    Eser, Busra Nur
    ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY HEAD AND NECK SURGERY, 2023, 85 (04): : 208 - 214
  • [30] Clinical Outcomes of the Cochlear™ Nucleus® 5 Cochlear Implant System and SmartSound™ 2 Signal Processing
    Runge, Christina L.
    Henion, Kathryn
    Tarima, Sergey
    Beiter, Anne
    Zwolan, Teresa A.
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2016, 27 (06) : 425 - 440