Modeling blurring effects due to continuous gantry rotation: Application to region of interest tomography

被引:9
作者
Cant, Jeroen [1 ,2 ]
Palenstijn, Willem Jan [1 ]
Behiels, Gert [2 ]
Sijbers, Jan [1 ]
机构
[1] Univ Antwerp, iMinds Vis Lab, B-2610 Antwerp, Belgium
[2] Agfa Healthcare NV, B-2640 Mortsel, Belgium
关键词
continous scanning; azimuthal blurring; ROI tomography; motion blur; continuous gantry rotation; IMAGE-RECONSTRUCTION; COMPUTED-TOMOGRAPHY; CT; QUALITY;
D O I
10.1118/1.4914422
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Projections acquired with continuous gantry rotation may suffer from blurring effects, depending on the rotation speed and the exposure time of each projection. This leads to blurred reconstructions if conventional reconstruction algorithms are applied. In this paper, the authors propose a reconstruction method for fast acquisitions based on a continuously moving and continuously emitting x-ray source. They study the trade-off between total acquisition time and reconstruction quality and compare with conventional reconstructions using projections acquired with a stepwise moving x-ray source. Methods: The authors introduce the algebraic reconstruction technique with angular integration concept, which models the angular integration due to the relative motion of the x-ray source during the projection. Results: Compared to conventional reconstruction from projections acquired with pulsed x-ray emission, the proposed method results in substantially improved reconstruction quality around the center of rotation. Outside this region, the proposed method results in improved radial resolution and a decreased tangential resolution. For a fixed reconstruction quality of this region of interest, the proposed method enables a lower number of projections and thus a faster acquisition. Conclusions: The modeling of the continuous gantry rotation in the proposed method substantially improves the reconstruction quality in a region of interest around the rotation center. The proposed method shows potential for fast region of interest tomography. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:2709 / 2717
页数:9
相关论文
共 27 条
  • [1] Effect of detector lag on CT noise power spectra
    Baek, Jongduk
    Pelc, Norbert J.
    [J]. MEDICAL PHYSICS, 2011, 38 (06) : 2995 - 3005
  • [2] Bracewell R., 1956, AUST J PHYS, V9, P198, DOI 10.1071/PH560198
  • [3] Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly undersampled projection data sets
    Chen, Guang-Hong
    Tang, Jie
    Leng, Shuai
    [J]. MEDICAL PHYSICS, 2008, 35 (02) : 660 - 663
  • [4] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [5] Computational analysis and improvement of SIRT
    Gregor, Jens
    Benson, Thomas
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (07) : 918 - 924
  • [6] Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography
    Hsieh, Jiang
    Londt, John
    Vass, Melissa
    Li, Jay
    Tang, Xiangyang
    Okerlund, Darin
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4236 - 4248
  • [7] Hubbell J. H., 1995, 5632 NISTIR
  • [8] Technique factors and image quality as functions of patient weight at abdominal CT
    Huda, W
    Scalzetti, EM
    Levin, G
    [J]. RADIOLOGY, 2000, 217 (02) : 430 - 435
  • [9] Cone-beam computed tomography with a flat-panel imager: Initial performance characterization
    Jaffray, DA
    Siewerdsen, JH
    [J]. MEDICAL PHYSICS, 2000, 27 (06) : 1311 - 1323
  • [10] SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION
    KALENDER, WA
    SEISSLER, W
    KLOTZ, E
    VOCK, P
    [J]. RADIOLOGY, 1990, 176 (01) : 181 - 183