Size-specific dose estimates for pediatric head CT protocols based on the AAPM report TG-293

被引:3
|
作者
Sapignoli, Sonia [1 ]
Roggio, Antonella [1 ]
Boschini, Alessio [2 ]
Guida, Federica [1 ]
Merlo, Carlo [1 ]
Paiusco, Marta [1 ]
Zorz, Alessandra [1 ]
De Monte, Francesca [1 ,3 ]
机构
[1] IRCCS, Veneto Inst Oncol IOV, Med Phys Dept, Padua, Italy
[2] Med Phys Dept, Azienda ULSS2 Marca Trevigiana, Treviso, Italy
[3] IRCCS, Veneto Inst Oncol IOV, Med Phys Dept, Via Gattamelata 64, I-35128 Padua, Italy
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2022年 / 100卷
关键词
SSDE; Head CT; Pediatric; Radiation dose tracking system; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; SUBSEQUENT RISK; PATTERNS; CHILDREN; TRENDS; SCAN;
D O I
10.1016/j.ejmp.2022.06.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To establish the Size Specific Dose Estimate (SSDE) typical values for pediatric head CT examinations based on the AAPM report TG-293; to provide a new stratification based on the water-equivalent diameter (Dw ), given that SSDE is related to the head size. Methods: 296 Head CT scans of pediatric patients collected using a dose monitoring software were retrospectively analysed. Typical values were derived stratifying data by age in three methods: the first proposed by the European Guidelines on Diagnostic Reference Levels for Pediatric Imaging (RP185), the second by the National Istisan Report 20/22 and a local one related to the clinical protocols (LStrata). For each scan, a self-developed Matlab routine calculated the water-equivalent diameter (Dw) and related SSDE values with the conversion factors fH16and fB16provided by the AAPM reports TG-293 and TG-204, respectively. Eventually, a Dwstratification was introduced starting from a measure of the lateral dimension of the head. Results: SSDE based on TG-204 overestimatesthe dose up to 12%. Four Dwgroups were identified thanks to the good correlation between the head lateral dimension andDw: Dw < 14 cm, 14 < Dw < 16 cm, 16 < Dw < 17 cm, Dw >= 17 cm. The Dw -stratified dosimetric indices present greater variability than those grouped by age because of the large variability of the size of the infant's head. Conclusions: The variability of the SSDE metric underlines that age-optimized protocols are not when size is considered.
引用
收藏
页码:26 / 30
页数:5
相关论文
共 39 条
  • [31] Multi-slice CT examinations of adult patients at Sudanese hospitals: radiation exposure based on size-specific dose estimates (SSDE)
    Bashier, Einas H.
    Suliman, I. I.
    RADIOLOGIA MEDICA, 2018, 123 (06): : 424 - 431
  • [32] Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in Water-equivalent Diameter
    Leng, Shuai
    Shiung, Maria
    Duan, Xinhui
    Yu, Lifeng
    Zhang, Yi
    McCollough, Cynthia H.
    RADIOLOGY, 2015, 276 (01) : 184 - 190
  • [33] PATIENT DOSE SURVEY BASED ON SIZE-SPECIFIC DOSE ESTIMATE AND ACCEPTABLE QUALITY DOSE IN CHEST AND ABDOMEN/PELVIS CT EXAMINATIONS
    Mehdipour, Ali
    Parsi, Masoumeh
    Khorram, Faezeh-Sadat
    RADIATION PROTECTION DOSIMETRY, 2019, 185 (02) : 176 - 182
  • [34] Estimating Patient Dose from X-ray Tube Output Metrics: Automated Measurement of Patient Size from CT Images Enables Large-scale Size-specific Dose Estimates
    Ikuta, Ichiro
    Warden, Graham I.
    Andriole, Katherine P.
    Khorasani, Ramin
    Sodickson, Aaron
    RADIOLOGY, 2014, 270 (02) : 472 - 480
  • [35] Radiation doses and size-specific dose estimate from CT brain examinations according to head sizes in a tertiary hospital in Malaysia
    Kayun, Zunaide
    Karim, Muhammad Khalis Abdul
    Harun, Hanif Haspi
    Shaari, Abdul Halim
    Mahmud, Rozi
    Hamid, Hamzaini Abdul
    Sabarudin, Akmal
    Chew, Ming Tsuey
    RADIATION PHYSICS AND CHEMISTRY, 2021, 189
  • [36] An Age-Based Size-Specific Dose Estimate for Pediatric Computed Tomography Head Examinations Performed at Songklanagarind Hospital, Thailand, from 2017 to 2019
    Poosiri, Saowapark
    Chuboonlap, Kanokkwan
    Kaewlaied, Nuttita
    APPLIED SCIENCES-BASEL, 2024, 14 (17):
  • [37] Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE)
    Masafumi Kidoh
    Daisuke Utsunomiya
    Seitaro Oda
    Takeshi Nakaura
    Yoshinori Funama
    Hideaki Yuki
    Kenichiro Hirata
    Tomohiro Namimoto
    Daisuke Sakabe
    Masahiro Hatemura
    Yasuyuki Yamashita
    European Radiology, 2017, 27 : 2267 - 2274
  • [38] Size-Specific Dose Estimates of Radiation Based on Body Weight and Body Mass Index for Chest and Abdomen-Pelvic CTs
    Xu, Jian
    Wang, Xiangquan
    Yang, Panfeng
    Luo, Kuangnan
    He, Xiaolong
    BIOMED RESEARCH INTERNATIONAL, 2020, 2020
  • [39] Conversion factor for size specific dose estimation of head CT scans based on age, for individuals from 0 up to 18 years old
    da Silva, Edilaine Honorio
    Baffa, Oswaldo
    Elias Jr, Jorge
    Buls, Nico
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (08):