Continuous wave cavity ring-down spectroscopy applied to in situ detection of dinitrogen pentoxide (N2O5)

被引:59
作者
Simpson, WR [1 ]
机构
[1] Univ Alaska Fairbanks, Dept Chem, Fairbanks, AK 99775 USA
[2] Univ Alaska Fairbanks, Inst Geophys, Fairbanks, AK 99775 USA
关键词
D O I
10.1063/1.1578705
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe an instrument using cavity ring-down spectroscopy (CRDS), an ultrasensitive detection method, to detect NO3 and N2O5 (via thermal dissociation to NO3 at 80 degreesC). We use continuous-wave (cw) cavity ring-down spectroscopy, allowing a highly reliable diode laser to act as the light source. This instrument uses all solid-state laser and optical components, and it is compact, portable, and power efficient. Advantages and disadvantages of cw CRDS compared to pulsed CRDS are discussed. Inlet losses and possible interferences were extensively investigated. We show field observations of N2O5 mixing ratios in ambient air. Dinitrogen pentoxide was measured because low ambient temperatures shift the equilibrium between NO3, NO2, and N2O5 strongly towards N2O5. Therefore, only N2O5 is assumed to be present in significant quantities in Fairbanks, AK, during winter. From these data, we identify times where the N2O5 mixing ratio must be zero (due to the absence of ozone indicating presence of NO) and use these time periods to measure the system's operational detection limit. The 2sigma detection limit for N2O5 is 2.4 parts per trillion by volume (pptv) in a 25 s average. Prediction of the N2O5 detection limit from estimates of ring-down signal noise indicates a detection limit of 1.6 pptv, in the same averaging period. The observed detection limit is about 50% larger than the predicted detection limit, indicating that other noise sources affect the true detection limit. A similar instrument using pulsed CRDS was described in the last year, and we compare our instrument to this instrument for detection of N2O5. (C) 2003 American Institute of Physics.
引用
收藏
页码:3442 / 3452
页数:11
相关论文
共 36 条
  • [1] Measurements of tropospheric NO3 at midlatitude
    Aliwell, SR
    Jones, RL
    [J]. JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 1998, 103 (D5) : 5719 - 5727
  • [2] MIRROR REFLECTOMETER BASED ON OPTICAL CAVITY DECAY TIME
    ANDERSON, DZ
    FRISCH, JC
    MASSER, CS
    [J]. APPLIED OPTICS, 1984, 23 (08): : 1238 - 1245
  • [3] LASER MEASUREMENTS OF OZONE ABSORPTION CROSS-SECTIONS IN THE CHAPPUIS BAND
    ANDERSON, SM
    MAUERSBERGER, K
    [J]. GEOPHYSICAL RESEARCH LETTERS, 1992, 19 (09) : 933 - 936
  • [4] Solving chemical problems of environmental importance using cavity ring-down spectroscopy
    Atkinson, DB
    [J]. ANALYST, 2003, 128 (02) : 117 - 125
  • [5] Broadband cavity ringdown spectroscopy of the NO3 radical
    Ball, SM
    Povey, IM
    Norton, EG
    Jones, RL
    [J]. CHEMICAL PHYSICS LETTERS, 2001, 342 (1-2) : 113 - 120
  • [6] High-latitude springtime photochemistry .1. NOx, PAN and ozone relationships
    Beine, HJ
    Jaffe, DA
    Herring, JA
    Kelley, JA
    Krognes, T
    Stordal, F
    [J]. JOURNAL OF ATMOSPHERIC CHEMISTRY, 1997, 27 (02) : 127 - 153
  • [7] BEINE HJ, 1996, THESIS U ALASKA FAIR
  • [8] Cavity ring-down spectroscopy: Experimental schemes and applications
    Berden, G
    Peeters, R
    Meijer, G
    [J]. INTERNATIONAL REVIEWS IN PHYSICAL CHEMISTRY, 2000, 19 (04) : 565 - 607
  • [9] Simultaneous in situ detection of atmospheric NO3 and N2O5 via cavity ring-down spectroscopy
    Brown, SS
    Stark, H
    Ciciora, SJ
    McLaughlin, RJ
    Ravishankara, AR
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (09) : 3291 - 3301
  • [10] Cavity ring-down spectroscopy for atmospheric trace gas detection:: application to the nitrate radical (NO3)
    Brown, SS
    Stark, H
    Ravishankara, AR
    [J]. APPLIED PHYSICS B-LASERS AND OPTICS, 2002, 75 (2-3): : 173 - 182