BirdcageBuilder: Design of specified-geometry birdcage coils with desired current pattern and resonant frequency

被引:61
作者
Chin, CL
Collins, CM
Li, SH
Dardzinski, BJ
Smith, MB [1 ]
机构
[1] Penn State Univ, Coll Med, Dept Radiol, Ctr NMR Res, Hershey, PA 17033 USA
[2] IGC Med Advances Inc, Milwaukee, WI 53288 USA
[3] Childrens Hosp, Med Ctr, Imaging Res Ctr, Cincinnati, OH 45229 USA
来源
CONCEPTS IN MAGNETIC RESONANCE | 2002年 / 15卷 / 02期
关键词
birdcage coil; radio frequency coil; coil; software;
D O I
10.1002/cmr.10030
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
A simple method for birdcage coil design (high pass, low pass, and band pass) is presented. Rather than iteratively approaching the Larmor resonant frequency with known capacitances and calculated inductances, a more versatile approach to birdcage coil design is developed and validated, in which the necessary capacitances are calculated using a desired current pattern, a calculated inductance, and a predetermined resonant frequency. in order to expedite coil design for experienced and novice coil builders, a computer program (BirdcageBuilder) is also implemented based on this method. Experimental results show that the calculated capacitances and actual capacitances for several existing birdcage coils are in good agreement. (C) 2002 Wiley Periodicals, Inc.
引用
收藏
页码:156 / 163
页数:8
相关论文
共 13 条
[1]  
CHIN C, 1997, P INT SOC MAGN RES M, P1498
[2]  
CHIN CL, 1998, P INT SOC MAGN RES M, P653
[3]  
COLLINS CM, 1998, P INT SOC MAGN RES M, P643
[4]   Practical aspects of birdcage coils [J].
Doty, FD ;
Entzminger, G ;
Hauck, CD ;
Staab, JP .
JOURNAL OF MAGNETIC RESONANCE, 1999, 138 (01) :144-154
[5]  
Grover F.W., 1946, INDUCTANCE CALCULATI
[6]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[7]   A SIMPLE METHOD TO INCORPORATE THE EFFECTS OF AN RF SHIELD INTO RF RESONATOR ANALYSIS FOR MRI APPLICATIONS [J].
JIN, JM ;
MAGIN, RL ;
SHEN, G ;
PERKINS, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (08) :840-843
[8]   A TECHNIQUE FOR DOUBLE RESONANT OPERATION OF BIRDCAGE IMAGING COILS [J].
JOSEPH, PM ;
LU, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (03) :286-294
[9]   Resonant modes of the birdcage coil [J].
Leifer, MC .
JOURNAL OF MAGNETIC RESONANCE, 1997, 124 (01) :51-60
[10]   A method to create an optimum current distribution and homogeneous B-1 field for elliptical birdcage coils [J].
Li, SZ ;
Collins, CM ;
Dardzinski, BJ ;
Chin, CL ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) :600-608