A Picosecond Image-Converter Dissector with a Crossed Sweep

被引:0
作者
Andreev, S. V. [1 ]
Vorobiev, N. S. [1 ]
Dorokhov, V. L. [2 ]
Zarovsky, A. I. [1 ]
Komelkov, A. S. [1 ]
Kurkin, G. Ya. [2 ]
Meshkov, O. I. [2 ,3 ]
Smirnov, A. V. [1 ]
机构
[1] Russian Acad Sci, Prokhorov Inst Gen Phys, Moscow 119991, Russia
[2] Russian Acad Sci, BINP, Siberian Branch, Novosibirsk 630090, Russia
[3] Novosibirsk State Univ, Novosibirsk 630090, Russia
关键词
Abstract—: A dual-slit dissector operating in the crossed-sweep mode with a picosecond time resolution has been created on the basis of the PIF-01 image converter tube. The dissector is applicable to synchrotron radiation sources and electron-positron colliders when it is necessary to simultaneously record longitudinal profiles of multi-bunch beams that fill two to several hundred adjacent separatrices of an accelerator ring. Two trains of light pulses shifted in time with respect to each other have been recorded with a time resolution of 6.0 ± 0.5 ps during the dynamic tests of the developed dissector with a laser system. © 2020; Pleiades Publishing; Ltd;
D O I
10.1134/S0020441220010017
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A dual-slit dissector operating in the crossed-sweep mode with a picosecond time resolution has been created on the basis of the PIF-01 image converter tube. The dissector is applicable to synchrotron radiation sources and electron-positron colliders when it is necessary to simultaneously record longitudinal profiles of multi-bunch beams that fill two to several hundred adjacent separatrices of an accelerator ring. Two trains of light pulses shifted in time with respect to each other have been recorded with a time resolution of 6.0 +/- 0.5 ps during the dynamic tests of the developed dissector with a laser system.
引用
收藏
页码:87 / 91
页数:5
相关论文
共 6 条
  • [1] Calibration of an image-tube picosecond dissector
    Anchugov O.V.
    Dorokhov V.L.
    Kurkin G.Y.
    Zinin E.I.
    Molyavin P.A.
    Meshkov O.I.
    Shvedov D.A.
    Petrozhitskii A.V.
    Gornostaev P.B.
    Zarovskii A.I.
    Shashkoff E.V.
    Smirnov A.V.
    Shchelev M.Y.
    [J]. Optoelectronics, Instrumentation and Data Processing, 2016, 52 (3) : 304 - 310
  • [2] A Picosecond Electron-Optical Dissector for Detecting Synchrotron Radiation
    Andreev, S. V.
    Anchugov, O. V.
    Vorob'ev, N. S.
    Dorokhov, V. L.
    Zarovskii, A. I.
    Komel'kov, A. S.
    Krutikhin, S. A.
    Kurkin, G. Ya.
    Malyutin, D. A.
    Matveenko, A. V.
    Meshkov, O. I.
    Smirnov, A. V.
    Shashkov, E. V.
    [J]. INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2019, 62 (02) : 208 - 213
  • [3] Monastyrskiy M.A., 2006, P INT C CHARG PART O, P23
  • [4] Pashkov P.T, 2006, FIZIKA PUCHKA KOLTSE
  • [5] Review of collective beam instabilities in electron-positron storage rings
    Smaluk, V. V.
    [J]. PHYSICS OF PARTICLES AND NUCLEI, 2012, 43 (02) : 204 - 230
  • [6] STROBOSCOPIC METHOD OF ELECTRO-OPTICAL PICOSECOND-RESOLUTION CHRONOGRAPHY AND ITS APPLICATION IN SYNCHROTRON RADIATION EXPERIMENTS
    ZININ, EI