Determining Significant Elevation of Intraocular Pressure Using Self-tonometry

被引:9
作者
Huang, Jessie [1 ,2 ]
Phu, Jack [1 ,2 ]
Kalloniatis, Michael [1 ,2 ]
Zangerl, Barbara [1 ,2 ]
机构
[1] Univ New South Wales, Ctr Eye Hlth, Kensington, NSW, Australia
[2] Univ New South Wales, Sch Optometry & Vis Sci, Kensington, NSW, Australia
关键词
CENTRAL CORNEAL THICKNESS; OPEN-ANGLE GLAUCOMA; REBOUND TONOMETER; BIOMECHANICAL PROPERTIES; PROSTAGLANDIN ANALOGS; APPLANATION TONOMETRY; ETHNIC-DIFFERENCES; DIURNAL-VARIATION; ACCURACY; EYE;
D O I
10.1097/OPX.0000000000001478
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE The purpose of this study was to use a large patient cohort to develop a practical, analytical tool for interpreting Icare HOME measurements with respect to applanation pressure. METHODS IOP measurements using the Icare HOME and an applanation tonometer were taken prospectively in 498 consecutive patients. Bland-Altman, frequency distribution, and linear regression analysis were applied to determine measurement differences. A novel criterion, Threshold Icare HOME IOP, was developed to assist identification of elevation above target applanation pressure, considering the expected diurnal variation and measurement variability. RESULTS Icare HOME tended to underestimate applanation tonometry (mean bias, -1.7 mmHg; 95% limits of agreement, -7.0 to +3.6). Overall, differences were within +/- 3 mmHg in 71.5% and +/- 5 mmHg in 92% of patients. Based on the novel criterion developed, Icare HOME measurements that exceed target applanation pressure by 6 mmHg or greater are generally outside the 95% limit of expected observations. CONCLUSIONS The Threshold Icare HOME IOP is a novel and practical criterion that can assist clinicians in their interpretation of Icare HOME phasing measurements with respect to target applanation pressures. Elevation above the expected thresholds may prompt closer monitoring or even modifications to glaucoma management.
引用
收藏
页码:86 / 93
页数:8
相关论文
共 50 条
[31]   Rebound Self-tonometry Acquisition Time and Ease of Use Evaluated by Newly Trained Optometry Students and Optometrists [J].
Hyatt, Ruth ;
Furtado, Nadine M. ;
Eberle, Darcy ;
Jensen, Katherine ;
Tsang, Timothy ;
Kwan, Justin .
OPTOMETRY AND VISION SCIENCE, 2020, 97 (02) :94-100
[32]   Influence of Corneal Visualization Scheimpflug Technology Tonometry on Intraocular Pressure [J].
Borroni, Davide ;
Gadhvi, Kunal Ajit ;
Hristova, Rozaliya ;
McLean, Keri ;
de Lossada, Carlos Rocha ;
Romano, Vito ;
Kaye, Stephen .
OPHTHALMOLOGY SCIENCE, 2021, 1 (01)
[33]   Intraocular Pressure: Goldmann Tonometry, Computational Model, and Calibration Equation [J].
Guzman, Andres Fernando ;
Castilla, Alejandro Arciniegas ;
Guarnieri, Fabio Ariel ;
Rodriguez, Fernando Ramirez .
JOURNAL OF GLAUCOMA, 2013, 22 (01) :10-14
[34]   Does the order of intraocular pressure measurement affect tonometry results? [J].
Pe'er, Oren ;
Chiu, Eni ;
Arad, Dikla ;
Lelescu, Cristina ;
Ross, Maya ;
Ofri, Ron .
VETERINARY OPHTHALMOLOGY, 2021, 24 :146-153
[35]   Head Elevation and Intraocular Pressure in Glaucoma [J].
Park, Ji-Hye ;
Nam, Ki Tae ;
Yoo, Chungkwon ;
Kim, Yong Yeon .
OPTOMETRY AND VISION SCIENCE, 2016, 93 (09) :1163-1170
[36]   Effect of Repeated Applanation Tonometry on the Accuracy of Intraocular Pressure Measurements [J].
Gaton, Dan D. ;
Ehrenberg, Miriam ;
Lusky, Moshe ;
Wussuki-Lior, Orly ;
Dotan, Gad ;
Weinberger, Dov ;
Snir, Moshe .
CURRENT EYE RESEARCH, 2010, 35 (06) :475-479
[37]   Non-contact tonometry: predicting intraocular pressure using a material-corneal thickness-independent methodology [J].
Redaelli, Elena ;
Calvo, Begona ;
Rodriguez Matas, Jose Felix ;
Luraghi, Giulia ;
Grasa, Jorge .
FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12
[38]   The influence of intraocular pressure and air jet pressure on corneal contactless tonometry tests [J].
Simonini, Irene ;
Pandolfi, Anna .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 58 :75-89
[39]   The effect of hydrogel and silicone hydrogel contact lenses on the measurement of intraocular pressure with rebound tonometry [J].
Zeri, Fabrizio ;
Calcatelli, Paolo ;
Donini, Bernardo ;
Lupelli, Luigi ;
Zarrilli, Luciana ;
Swann, Peter G. .
CONTACT LENS & ANTERIOR EYE, 2011, 34 (06) :260-265
[40]   Comparison of Scleral Tono-Pen Intraocular Pressure Measurements with Goldmann Applanation Tonometry [J].
Badakere, Swathi Vallabh ;
Choudhari, Nikhil S. ;
Rao, Harsha L. ;
Chary, C. Raghava ;
Garudadri, Chandra Sekhar ;
Senthil, Sirisha .
OPTOMETRY AND VISION SCIENCE, 2018, 95 (02) :129-135