Reproducibility of arterial elasticity parameters derived from radial artery diastolic pulse contour analysis: the multi-ethnic study of atherosclerosis

被引:16
作者
Brumback, Lyndia C. [1 ]
Jacobs, David R., Jr. [2 ]
Dermond, Norma [1 ]
Chen, Haiying [4 ]
Duprez, Daniel A. [3 ]
机构
[1] Univ Washington, Dept Biostat, Seattle, WA 98195 USA
[2] Univ Minnesota, Div Epidemiol & Community Hlth, Minneapolis, MN 55455 USA
[3] Univ Minnesota, Div Cardiovasc, Minneapolis, MN 55455 USA
[4] Wake Forest Univ, Dept Biostat Sci, Winston Salem, NC 27109 USA
关键词
arterial compliance; arterial elasticity; blood pressure measurement; reproducibility; REPEATABILITY; POPULATION; SYSTEM;
D O I
10.1097/MBP.0b013e32833fe2a6
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The aim of this report was to describe the reproducibility of arterial elasticity parameters derived from radial artery diastolic pulse contour analysis. Methods Two measurements were taken on the same day by the same technician with the HDI/Pulsewave CR-2000 System on each of a subset of participants selected randomly from the Multi-Ethnic Study of Atherosclerosis. Large and small artery elasticities (C1 and C2), C1 and C2 each multiplied by systemic vascular resistance (C1xR and C2xR, which are functions of the radial artery pulse waveform only), and systolic and diastolic blood pressures (SBP and DBP) were evaluated. Results Two measurements were taken of each of the 131 participants. The mean difference +/- standard deviation of the difference was -0.13 +/- 3.5 mmHg x 10 for C1, -0.03 +/- 1.5 mmHg x 100 for C2, 0.10 +/- 4.6 s x 10 for C1xR, 0.05 +/- 1.8 s x 100 for C2xR, 2.8 +/- 8.2 mmHg for SBP, and 1.5 +/- 5.0 mmHg for DBP. The between-measure correlation was 0.74 for C1, 0.84 for C2, 0.58 for C1xR, 0.74 for C2xR, 0.90 for SBP, and 0.86 for DBP. Conclusion The reproducibility of arterial elasticity measurements is less than that of SBP and DBP. Multiple measurements of arterial elasticity of a participant are recommended. Blood Press Monit 15:312-315 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:312 / 315
页数:4
相关论文
共 7 条
[1]   Comparison of arterial elasticity measured in left and right arms using the HDI/Pulsewave ™ CR-2000 Research System [J].
Alavi, HB ;
Prisant, LM ;
Jupin, D ;
Oracion, A .
BLOOD PRESSURE MONITORING, 2002, 7 (05) :277-280
[2]   Multi-ethnic study of atherosclerosis: Objectives and design [J].
Bild, DE ;
Bluemke, DA ;
Burke, GL ;
Detrano, R ;
Roux, AVD ;
Folsom, AR ;
Greenland, P ;
Jacobs, DR ;
Kronmal, R ;
Liu, K ;
Nelson, JC ;
O'Leary, D ;
Saad, MF ;
Shea, S ;
Szklo, M ;
Tracy, RP .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 2002, 156 (09) :871-881
[3]  
Duprez DA, 2009, J AM COLL CARDIOL, V53, pA382
[4]  
FINKELSTEIN SM, 1992, J HYPERTENS, V10, pS11
[5]   Assessment of repeatability and correlates of arterial compliance [J].
Prisant, LM ;
Pasi, M ;
Jupin, D ;
Prisant, ME .
BLOOD PRESSURE MONITORING, 2002, 7 (04) :231-235
[6]   A comparison between systolic and diastolic pulse contour analysis in the evaluation of arterial stiffness [J].
Rietzschel, ER ;
Boeykens, E ;
De Buyzere, ML ;
Duprez, DA ;
Clement, DL .
HYPERTENSION, 2001, 37 (06) :E15-E22
[7]   Determination of arterial compliance using blood pressure waveform analysis with the CR-2000 system - Reliability, repeatability, and establishment of normal values for healthy European population - The Seven European Sites Study (SESS) [J].
Zimlichman, R ;
Shargorodsky, M ;
Boaz, M ;
Duprez, D ;
Rahn, KH ;
Rizzoni, D ;
Payeras, AC ;
Hamm, C ;
McVeigh, G .
AMERICAN JOURNAL OF HYPERTENSION, 2005, 18 (01) :65-71