Optimisation of movement detection and artifact removal during laser speckle contrast imaging

被引:32
作者
Omarjee, Loukman [1 ]
Signolet, Isabelle [1 ]
Humeau-Heutier, Anne [2 ]
Martin, Ludovic [3 ,4 ]
Henrion, Daniel [3 ]
Abraham, Pierre [3 ]
机构
[1] Univ Hosp, Lab Physiol Vasc Invest, F-49933 Angers 09, France
[2] Univ Angers, LUNAM Univ, LISA, F-49045 Angers, France
[3] Univ Angers, INSERM, CNR5 UMR 6214, UFR Med,U1083, F-49045 Angers, France
[4] Univ Hosp, Dept Dermatol, F-49933 Angers 09, France
关键词
Laser speckle contrast imaging; Noise; Microcirculation; Blood flow; Signal processing; MICROVASCULAR REACTIVITY; TIME;
D O I
10.1016/j.mvr.2014.09.005
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Introduction: Laser speckle contrast imaging (LSCI) allows an easy non-contact monitoring of the cutaneous blood flow (CBF), but is highly sensitive to movement artifacts (ARTm). Subtraction of a signal recorded on an adhesive opaque surface (AOS) close to the area of interest was reported as a mean of reducing noise from the raw skin LSCI (LSCIsk) signal, provided an individual calibration was performed. Assuming that AOS = a . CBF + b . ARTm, an ideal patch should completely block the light reflection due to CBF and thus be insensitive to skin blood flow changes ("a" similar to 0), while keeping a reflection signal amplitude similar to the one from the skin in case of artifact ("b" similar to 1). This ideal AOS has not been determined and may discriminate flow from movements during LSCI recordings. Materials and methods: We tested different AOSs to determine their "a" and "b" parameters in 35 and 34 healthy volunteers, respectively. The AOS surface providing results as close as possible to an ideal AOS, was used for a point-by-point de-noising of post occlusive reactive hyperemia (PORH) on two different days in 15 new subjects. Correlation of raw, smoothed (average smoothing over 1 s intervals) and denoised signals was tested through a cross-correlation analysis of the two POHR tests. Results: The optimal "a" and "b" values were obtained with a homemade bilayer adhesive patch (a = 0.06 +/- 0.05 and b = 1.03 +/- 0.17) whereas other tested AOS had "a" values ranging from 0.05 to 023 and "b" values ranging from 2.69 to 3.82. Using the bilayer adhesive patch the cross-correlation between the two tests of POHR increased from 0330 +/- 0.128 for raw, to 0.461 +/- 0.168 for smoothed and 0.649 +/- 0.128 for denoised signals respectively (p < 0.05 from raw coefficients). Conclusion: The home-made bilayer adhesive seems the optimal AOS for the removal of ARTm from the LSCIsk signal while respecting CBF signal. This specific AOS allows for an efficient de-noising of LSCI measurements without the need for individual calibration. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:75 / 80
页数:6
相关论文
共 13 条
[1]   Time course of forearm arterial compliance changes during reactive hyperemia [J].
Baldassarre, D ;
Amato, M ;
Palombo, C ;
Morizzo, C ;
Pustina, L ;
Sirtori, CR .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2001, 281 (03) :H1093-H1103
[2]   Laser speckle contrast imaging: theoretical and practical limitations [J].
Briers, David ;
Duncan, Donald D. ;
Hirst, Evan ;
Kirkpatrick, Sean J. ;
Larsson, Marcus ;
Steenbergen, Wiendelt ;
Stromberg, Tomas ;
Thompson, Oliver B. .
JOURNAL OF BIOMEDICAL OPTICS, 2013, 18 (06)
[3]   Evaluation of systemic microvascular endothelial function using laser speckle contrast imaging [J].
Cordovil, Ivan ;
Huguenin, Grazielle ;
Rosa, Glorimar ;
Bello, Annie ;
Koehler, Onilmar ;
de Moraes, Roger ;
Tibirica, Eduardo .
MICROVASCULAR RESEARCH, 2012, 83 (03) :376-379
[4]   Skin microdialysis coupled with Laser Speckle Contrast Imaging to assess microvascular reactivity [J].
Cracowski, J. L. ;
Gaillard-Bigot, F. ;
Cracowski, C. ;
Roustit, M. ;
Millet, C. .
MICROVASCULAR RESEARCH, 2011, 82 (03) :333-338
[5]   Cutaneous microvascular functional assessment during exercise: a novel approach using laser speckle contrast imaging [J].
Mahe, G. ;
Abraham, P. ;
Le Faucheur, A. ;
Bruneau, A. ;
Humeau-Heurtier, A. ;
Durand, S. .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 2013, 465 (04) :451-458
[6]   Air movements interfere with laser speckle contrast imaging recordings [J].
Mahe, Guillaume ;
Durand, Sylvain ;
Humeau, Anne ;
Leftheriotis, Georges ;
Rousseau, Pascal ;
Abraham, Pierre .
LASERS IN MEDICAL SCIENCE, 2012, 27 (05) :1073-1076
[7]   Distance between laser head and skin does not influence skin blood flow values recorded by laser speckle imaging [J].
Mahe, Guillaume ;
Haj-Yassin, Firas ;
Rousseau, Pascal ;
Humeau, Anne ;
Durand, Sylvain ;
Leftheriotis, Georges ;
Abraham, Pierre .
MICROVASCULAR RESEARCH, 2011, 82 (03) :439-442
[8]   Laser speckle contrast imaging accurately measures blood flow over moving skin surfaces [J].
Mahe, Guillaume ;
Rousseau, Pascal ;
Durand, Sylvain ;
Bricq, Stephanie ;
Leftheriotis, Georges ;
Abraham, Pierre .
MICROVASCULAR RESEARCH, 2011, 81 (02) :183-188
[9]   Laser speckle contrast imaging of cerebral blood flow in humans during neurosurgery: a pilot clinical study [J].
Parthasarathy, Ashwin B. ;
Weber, Erica L. ;
Richards, Lisa M. ;
Fox, Douglas J. ;
Dunn, Andrew K. .
JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (06)
[10]   Comparison of infrared thermography and laser speckle contrast imaging for the dynamic assessment of digital microvascular function [J].
Pauling, J. D. ;
Shipley, J. A. ;
Raper, S. ;
Watson, M. L. ;
Ward, S. G. ;
Harris, N. D. ;
McHugh, N. J. .
MICROVASCULAR RESEARCH, 2012, 83 (02) :162-167