Optimized FIB silicon samples suitable for lattice parameters measurements by convergent beam electron diffraction

被引:8
作者
Alexandre, L. [1 ]
Rousseau, K. [1 ]
Alfonso, C. [1 ]
Saikaly, W. [2 ]
Fares, L. [3 ]
Grosjean, C. [3 ]
Charai, A. [1 ]
机构
[1] Univ Paul Cezanne, TECSEN UMR 6122, Fac Sci St Jerome, F-13397 Marseille 20, France
[2] Univ Paul Cezanne, CP2M, Fac Sci St Jerome, F-13397 Marseille 20, France
[3] Zone Ind Rousset, RCCAL, St Microelect Rousset, F-13106 Rousset, France
关键词
convergent beam electron diffraction; electron energy loss spectroscopy; focused ion beam; damaged layer; amorphized layer thickness; specimen thickness measurement; strain measurement;
D O I
10.1016/j.micron.2007.01.005
中图分类号
TH742 [显微镜];
学科分类号
摘要
The aim of this paper is to check the effect of artefacts introduced by focused ion beam (FIB) milling on the strain measurement by convergent beam electron diffraction (CBED). We show that on optimized silicon FIB samples, the strain measurement can be performed with a sensitivity of about 2.5 x 10(-4) which is very close to the theoretical one and we conclude that FIB preparation can be suitable for such measurements in microelectronic devices. To achieve this, we first used CBED and electron energy loss spectroscopy (EELS) which provide a procedure permitting an exact knowledge of the sample geometry, i.e. the thickness of both amorphous and crystalline layers. This procedure was used in order to measure the FIB-amorphized sidewall layer. It was found that if the FIB preparation is optimized one can reduce this amorphous layer down to around 7 nm on each side. Secondly different preparation techniques (cleavage, Tripod (TM) and FIB) permit to check if the surface damaged layer introduced by FIB influences the strain state of the sample. Finally, it was found that the damaged layer does not introduce measurable strain in pure silicon but reduces appreciably the quality of the CBED patterns. (C) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:294 / 301
页数:8
相关论文
共 32 条
[1]  
ALLEN SM, 2000, PHILOS MAG, V43, P325
[2]   Measuring the thickness of aluminium alloy thin foils using electron energy loss spectroscopy [J].
Bardal, A ;
Lie, I .
MATERIALS CHARACTERIZATION, 2000, 44 (03) :329-343
[3]  
BENEDICT J, 1990, MATER RES SOC S P, V115, P265
[4]   Volume fraction measurement of dispersoids in a thin foil by parallel energy-loss spectroscopy: Development and assessment of the technique [J].
Botton, GA ;
LEsperance, G ;
Gallerneault, CE ;
Ball, MD .
JOURNAL OF MICROSCOPY-OXFORD, 1995, 180 :217-229
[5]   Crystal thickness and extinction distance determination using energy filtered CBED pattern intensity measurement and dynamical diffraction theory fitting [J].
Delille, D ;
Pantel, R ;
Van Cappellen, E .
ULTRAMICROSCOPY, 2001, 87 (1-2) :5-18
[6]  
DELILLE D, 1999, MICROSC MICROANAL, P898
[7]   RELATIVISTIC HARTREE-FOCK X-RAY AND ELECTRON SCATTERING FACTORS [J].
DOYLE, PA ;
TURNER, PS .
ACTA CRYSTALLOGRAPHICA SECTION A-CRYSTAL PHYSICS DIFFRACTION THEORETICAL AND GENERAL CRYSTALLOGRAPHY, 1968, A 24 :390-&
[8]  
EGERTON, 1996, ELECT ENERGY LOSS SP
[9]   A green electroluminescent device made from an yttrium complex [J].
Gao, DQ ;
Bian, ZQ ;
Guan, M ;
Li, FY ;
Huang, CH .
RESEARCH ON CHEMICAL INTERMEDIATES, 2004, 30 (4-5) :563-568
[10]  
*GATAN, 5933 GATAN