Typical Accuracy and Quality Control of a Process for Creating CT-Based Virtual Bone Models

被引:17
作者
Noser, Hansrudi [1 ]
Heldstab, Thomas [1 ]
Schmutz, Beat [2 ]
Kamer, Lukas [1 ]
机构
[1] AO Fdn, CH-7270 Davos, Switzerland
[2] Queensland Univ Technol, Kelvin Grove, Qld 4059, Australia
关键词
Computed tomography; virtual bone models; size accuracy;
D O I
10.1007/s10278-010-9287-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A pragmatic method for assessing the accuracy and precision of a given processing pipeline required for converting computed tomography (CT) image data of bones into representative three dimensional (3D) models of bone shapes is proposed. The method is based on coprocessing a control object with known geometry which enables the assessment of the quality of resulting 3D models. At three stages of the conversion process, distance measurements were obtained and statistically evaluated. For this study, 31 CT datasets were processed. The final 3D model of the control object contained an average deviation from reference values of -1.07 +/- 0.52 mm standard deviation (SD) for edge distances and -0.647 +/- 0.43 mm SD for parallel side distances of the control object. Coprocessing a reference object enables the assessment of the accuracy and precision of a given processing pipeline for creating CT-based 3D bone models and is suitable for detecting most systematic or human errors when processing a CT-scan. Typical errors have about the same size as the scan resolution.
引用
收藏
页码:437 / 445
页数:9
相关论文
共 10 条
  • [1] Bade R., 2006, SIMULATION VISUALISI, V1, P289
  • [2] Metro:: Measuring error on simplified surfaces
    Cignoni, P
    Rocchini, C
    Scopigno, R
    [J]. COMPUTER GRAPHICS FORUM, 1998, 17 (02) : 167 - 174
  • [3] Surface modeling - Uncertainty estimation and visualization
    Drapikowski, Pawel
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2008, 32 (02) : 134 - 139
  • [4] A CT database for research, development and education: Concept and potential
    Messmer, Peter
    Matthews, Felix
    Jacob, Augustinus Ludwig
    Kikinis, Ron
    Regazzoni, Pietro
    Noser, Hansruedi
    [J]. JOURNAL OF DIGITAL IMAGING, 2007, 20 (01) : 17 - 22
  • [5] Piegl L, 1996, The Nurbs Book
  • [6] Schmutz Beat, 2007, 6 AUSTR BIOM C, P93
  • [7] The quality of reconstructed 3D images in multidetector-row helical CT: Experimental study involving scan parameters
    Shin, JH
    Lee, HK
    Choi, CG
    Suh, DC
    Lim, TH
    Kang, W
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2002, 3 (01) : 49 - 56
  • [8] Volume rendering in the presence of partial volume effects
    Souza, A
    Udupa, JK
    Saha, PK
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (02) : 223 - 235
  • [9] STALLING D, 1998, CAR, V98, P137
  • [10] Recommendations for thresholds for cortical bone geometry and density measurement by peripheral quantitative computed tomography
    Ward, KA
    Adams, JE
    Hangartner, TN
    [J]. CALCIFIED TISSUE INTERNATIONAL, 2005, 77 (05) : 275 - 280