A novel method for kinetic measurements of rare cell proliferation using Cellometer image-based cytometry

被引:10
作者
Chan, Leo L. [1 ]
Zhong, Xuemei [2 ]
Pirani, Alnoor [3 ]
Lin, Bo [3 ]
机构
[1] Nexcelom Biosci LLC, Dept Technol R&D, Lawrence, MA 01843 USA
[2] Boston Univ, Med Ctr, Dept Med, Hematol Oncol Sect, Boston, MA 02118 USA
[3] Nexcelom Biosci LLC, Dept Applicat, Lawrence, MA 01843 USA
关键词
Cell proliferation; Cellometer Vision; Image-based Cytometry (IBC); CFSE-labeling; B1 B cell; B2 B cell; FLOW-CYTOMETRY; B-CELLS; DIVISION; DIFFERENTIATION; GROWTH; DNA;
D O I
10.1016/j.jim.2012.01.006
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Cell proliferation is an important assay for pharmaceutical and biomedical research to test the effects of a variety of treatments on cultured primary cells or cell lines. For immunological studies, the ability to perform rapid cell proliferation analysis allows the identification of potential biological reagents for inducing or inhibiting immune cell proliferation. Current cell proliferation analysis methods employ flow cytometry for fluorescence detection of CFSE-labeled cells. However, conventional flow cytometers require a considerable amount of cells per sample, which becomes an issue for kinetic measurements with rare cell population due to the lack of samples for flow cytometric analyses at multiple time points during proliferation period. Here we report the development of a novel cell proliferation kinetic detection method for low cell concentration samples using the new Cellometer Vision system. Since the Cellometer system requires only 20 mu l of sample, cell proliferation can be measured at multiple time points over the entire culturing period, whereas typically, flow cytometry is only performed at the end of the proliferation period. To validate the detection method, B1 and B2 B cells were treated with a B cell mitogen for 6 days, and proliferation was measured using Cellometer on day 1, 3, 5, and 6.10 demonstrate the capability of the system, B1 B cells were treated with a panel of TLR agonists (Pam3Cys, PolyIC, CLO97, and CpG) for 7 days, and proliferation was measured on day 2, 4, 6, and 7. Cellometer image-based cytometry (IBC) was able to obtain proliferation results on each day with the last time point comparable to flow cytometry. This novel method allows for kinetic measurements of the rare cell samples such as B1 B cell, which has the potential to revolutionize kinetic analysis of cell proliferation. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:8 / 14
页数:7
相关论文
共 24 条
  • [1] A rapid method for evaluation of cell number and viability by flow cytometry
    AlRubeai, M
    Welzenbach, K
    Lloyd, DR
    Emery, AN
    [J]. CYTOTECHNOLOGY, 1997, 24 (02) : 161 - 168
  • [2] Liver regeneration: Methods for monitoring and their applications
    Assy, N
    Minuk, GY
    [J]. JOURNAL OF HEPATOLOGY, 1997, 26 (04) : 945 - 952
  • [3] THE ERROR OF THE RED CELL COUNT
    BIGGS, R
    MACMILLAN, RL
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1948, 1 (05) : 288 - 291
  • [4] Callard R, 2007, IMMUNOL REV, V216, P119
  • [5] Direct concentration and viability measurement of yeast in corn mash using a novel imaging cytometry method
    Chan, Leo L.
    Lyettefi, Emily J.
    Pirani, Alnoor
    Smith, Tim
    Qiu, Jean
    Lin, Bo
    [J]. JOURNAL OF INDUSTRIAL MICROBIOLOGY & BIOTECHNOLOGY, 2011, 38 (08) : 1109 - 1115
  • [6] Cellometer Vision as an Alternative to Flow Cytometry for Cell Cycle Analysis, Mitochondrial Potential, and Immunophenotyping
    Chan, Leo L.
    Zhong, Xuemei
    Qiu, Jean
    Li, Peter Y.
    Lin, Bo
    [J]. CYTOMETRY PART A, 2011, 79A (07) : 507 - 517
  • [7] A rapid detection method for apoptosis and necrosis measurement using the Cellometer imaging cytometry
    Chan, Leo Li-Ying
    Lai, Ning
    Wang, Elena
    Smith, Tim
    Yang, Xifeng
    Lin, Bo
    [J]. APOPTOSIS, 2011, 16 (12) : 1295 - 1303
  • [8] Proliferative markers as prognostic and predictive tools in early breast cancer: where are we now?
    Colozza, M
    Azambuja, E
    Cardoso, F
    Sotiriou, C
    Larsimont, D
    Piccart, MJ
    [J]. ANNALS OF ONCOLOGY, 2005, 16 (11) : 1723 - 1739
  • [9] Flow cytometry and cell sorting of heterogeneous microbial populations: The importance of single-cell analyses
    Davey, HM
    Kell, DB
    [J]. MICROBIOLOGICAL REVIEWS, 1996, 60 (04) : 641 - +
  • [10] Blockade of B-cell-activating factor suppresses lupus-like syndrome in autoimmune BXSB mice
    Ding, Hanlu
    Wang, Li
    Wu, Xiongfei
    Yan, Jun
    He, Yani
    Ni, Bing
    Gao, Wenda
    Zhong, Xuemei
    [J]. JOURNAL OF CELLULAR AND MOLECULAR MEDICINE, 2010, 14 (6B) : 1717 - 1725