Measuring lower limb circumference and volume - introduction of a novel optical 3D volumetric measurement system

被引:12
作者
Tischer, Thomas [1 ]
Oye, Sebastian [1 ]
Wolf, Andreas [1 ]
Feldhege, Frank [1 ]
Jacksteit, Robert [1 ]
Mittelmeier, Wolfram [1 ]
Bader, Rainer [1 ]
Mau-Moeller, Anett [1 ]
机构
[1] Univ Med Rostock, Dept Orthopaed, Doberaner Str 142, D-18057 Rostock, Germany
来源
BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK | 2020年 / 65卷 / 02期
关键词
cardiovascular disease; edema; swelling; venous disease; volume measurement; RELIABILITY; LYMPHEDEMA; EDEMA;
D O I
10.1515/bmt-2018-0100
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study comprised two experiments: (1) the reliability of a novel optical three-dimensional (3D) volumetric measurement system (BODYTRONIC (R) 600) for the assessment of lower limb circumference and volume; (2) the comparison of data obtained from BODYTRONIC (R) 600 with that provided by computed tomography (CT) for accuracy estimation. Reliability was assessed in 20 healthy subjects. Accuracy was determined by comparing the deviations in the surface topology of two 3D models obtained from BODYTRONIC (R) 600 and CT. Reliability was very high for leg circumference measures [coefficient of variations (CVs) range 0.3%-1.3%] and slightly lower for foot circumference (CVs around 2.0%). Reliability of leg volume was also found to be very high (CVs <= 2.5%). Differences in surface topology between BODYTRONIC (R) 600 and CT were primarily below 1 mm indicating high accuracy. The volumes of the foot were higher (range 0.9% 1.7%) and that of the leg, lower (range 1.0% 1.3%) compared to CT. The BODYTRONIC (R) 600 system provides fast, highly reliable and accurate measures of lower limb circumference and volume and can be considered as a valuable measurement tool for use in various research and clinical applications.
引用
收藏
页码:237 / 241
页数:5
相关论文
共 14 条
[1]   Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine [J].
Atkinson, G ;
Nevill, AM .
SPORTS MEDICINE, 1998, 26 (04) :217-238
[2]  
Brijker F, 2000, CLIN PHYSIOL, V20, P56
[3]   Reproducibility and day time bias correction of optoelectronic leg volumetry: a prospective cohort study [J].
Engelberger, Rolf P. ;
Blazek, Claudia ;
Amsler, Felix ;
Keo, Hong H. ;
Baumann, Frederic ;
Blattler, Werner ;
Baumgartner, Iris ;
Willenberg, Torsten .
BMC MEDICAL RESEARCH METHODOLOGY, 2011, 11
[4]   A comparison of different assessment techniques for measuring foot and ankle volume in healthy adults [J].
Friends, John ;
Augustine, Elizabeth ;
Danoff, Jerome .
JOURNAL OF THE AMERICAN PODIATRIC MEDICAL ASSOCIATION, 2008, 98 (02) :85-94
[5]   3D stereophotogrammetry in upper-extremity lymphedema: An accurate diagnostic method [J].
Hameeteman, Marijn ;
Verhulst, Arico C. ;
Vreeken, Rinaldo D. ;
Maal, Thomas J. J. ;
Ulrich, Dietmar J. O. .
JOURNAL OF PLASTIC RECONSTRUCTIVE AND AESTHETIC SURGERY, 2016, 69 (02) :241-247
[6]   Measures of reliability in sports medicine and science [J].
Hopkins, WG .
SPORTS MEDICINE, 2000, 30 (01) :1-15
[7]  
Hopkins WG, 2015, SPORTSCI, V19, P26
[8]   The validity and reliability of the Perometer in evaluating human knee volume [J].
Man, IOW ;
Markland, KL ;
Morrissey, MC .
CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2004, 24 (06) :352-358
[9]  
Mayrovitz H N, 2000, Adv Skin Wound Care, V13, P272
[10]   Three-Dimensional Camera System for Measuring Arm Volume in Women with Lymphedema Following Breast Cancer Treatment [J].
Ohberg, Fredrik ;
Zachrisson, Adam ;
Holmner-Rocklov, Asa .
LYMPHATIC RESEARCH AND BIOLOGY, 2014, 12 (04) :267-274