Sequential versus automated cutting guides in computer-assisted total knee arthroplasty

被引:40
作者
Koulalis, Dimitrios [2 ]
O'Loughlin, Padhraig F. [1 ]
Plaskos, Christopher [1 ]
Kendoff, Daniel [1 ]
Cross, Michael B. [1 ]
Pearle, Andrew D. [1 ]
机构
[1] Hosp Special Surg, New York, NY 10021 USA
[2] Univ Athens, Dept Orthopaed Surg, Univ Hosp Attikon, Athens 12462, Greece
关键词
Computer-assisted TKA; Robotic knee; Cutting guides; Precision; Accuracy; Operative time; LOAD DISTRIBUTION; TIBIAL COMPONENT; REPLACEMENT; ALIGNMENT; SURGERY; ACCURACY; POSITION; FLEXION; ERRORS;
D O I
10.1016/j.knee.2010.08.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The accuracy and efficiency of automated cutting guides in CAS systems have not been previously compared with conventional CAS techniques. Therefore, it is not yet clear if these more advanced technologies are warranted. We hypothesized that a novel automated cutting guide with CAS for total knee arthroplasty would be more efficient and more accurate than conventional navigation with sequential cutting blocks. Twelve cadaver legs were used in total. Each leg was randomly assigned to either an automated guide positioning or a conventional freehand computer-navigated guide positioning. The guide positions postosseous fixation and the final bone-cut surfaces were digitized and compared to the targeted cutting planes. The final location of the impacted trial implant was also digitized and compared to the planned implant location. The time for each step and the total time taken to prepare the femur were measured for both groups. The mean femoral preparation time was shorter with the automated cutting guide than the conventional method (5.5 min versus 13.8 min, p<0.001). The average deviation in the final bone resections from the planned resections was significantly lower for the automated cutting guide in the frontal/rotational plane (0.55 degrees versus 1.1 degrees), sagittal plane (0.75 degrees versus 2.0 degrees), and cut height direction (0.56 mm versus 1.6 mm). Therefore, based on these results, we concluded that automated cutting-guide positioning resulted in more efficient and more accurate femoral cuts in comparison to the conventional navigation method in a cadaveric model. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:436 / 442
页数:7
相关论文
共 28 条
  • [1] Fixation of a hydroxyapatite-tricalcium phosphate-coated cementless knee prosthesis - Clinical and radiographic evaluation seven years after surgery
    Akizuki, S
    Takizawa, T
    Horiuchi, H
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2003, 85B (08): : 1123 - 1127
  • [2] In vivo patellar kinematics during total knee arthroplasty
    Anglin, Carolyn
    Ho, Karen C. T.
    Briard, Jean-Louis
    De Lambilly, Charles
    Plaskos, Christopher
    Nodwell, Eric
    Stindel, Eric
    [J]. COMPUTER AIDED SURGERY, 2008, 13 (06) : 377 - 391
  • [3] Towards a standard in assessment of bone cutting for total knee replacement
    Barrera, O. A.
    Haider, H.
    Garvin, K. L.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H1) : 63 - 74
  • [4] Intraoperative cutting errors in total knee arthroplasty
    Bäthis, H
    Perlick, L
    Tingart, M
    Perlick, C
    Lüring, C
    Grifka, J
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (01) : 16 - 20
  • [5] Navigated total knee replacement - A meta-analysis
    Bauwens, Kai
    Matthes, Gerrit
    Wich, Michael
    Gebhard, Florian
    Hanson, Beate
    Ekkernkamp, Axel
    Stengel, Dirk
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (02) : 261 - 269
  • [6] Robot-assisted total knee arthroplasty
    Bellemans, Johan
    Vandenneucker, Hilde
    Vanlauwe, Johan
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (464) : 111 - 116
  • [7] Tibial component failure mechanisms in total knee arthroplasty
    Berend, ME
    Ritter, MA
    Meding, JB
    Faris, PM
    Keating, EM
    Redelman, R
    Faris, GW
    Davis, KE
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) : 26 - 34
  • [8] The accuracy of bone resections made during computer navigated total knee replacement. Do we resect what the computer plans we resect?
    Biant, Leela C.
    Yeoh, Kwan
    Walker, Peter M.
    Bruce, Warwick J. M.
    Walsh, William R.
    [J]. KNEE, 2008, 15 (03) : 238 - 241
  • [9] Alignment deviation between bone resection and final implant positioning in computer-navigated total knee arthroplasty
    Catani, Fabio
    Biasca, Nicola
    Ensini, Andrea
    Leardini, Alberto
    Bianchi, Luca
    Digennaro, Vitantonio
    Giannini, Sandro
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (04) : 765 - 771
  • [10] Coller JP, 1991, CONTROVERSIES TOTAL, P95