Considerations in Selecting a Fitting a Function for Film Dosimetry in Radiation Therapy

被引:0
|
作者
Hwang, Ui-Jung [1 ]
Kim, Ki Hwan [1 ]
Lim, Young Kyung [2 ]
Kim, Haksoo [2 ]
Choi, Sang Hyoun [3 ]
Min, Byung Jun [4 ]
Jeong, Hojin [5 ]
Rah, Jeong Eun [6 ]
机构
[1] Chungnam Natl Univ Hosp, Dept Radiat Oncol, Daejeon 35015, South Korea
[2] Natl Canc Ctr, Proton Therapy Ctr, Goyang 10408, South Korea
[3] Korea Inst Radiol & Med Sci, Res Ctr Radiotherapy, Seoul 01812, South Korea
[4] Chungbuk Natl Univ Hosp, Dept Radiat Oncol, Cheongju 28644, South Korea
[5] Gyeongsang Natl Univ, Inst Hlth Sci, Dept Radiat Oncol, Jinju 52727, South Korea
[6] Myongji Hosp, Dept Radiat Oncol, Goyang 10475, South Korea
基金
新加坡国家研究基金会;
关键词
Radiation therapy; Quality assurance; EBT3; film; OD to dose calibration; Curve fitting; PORTAL IMAGING DEVICE; FLAT-BED SCANNER; RADIOCHROMIC FILM; EBT FILM; SENSITOMETRIC CURVES; PHOTON; MODEL; CALIBRATION; ELECTRON; PROTOCOL;
D O I
10.3938/jkps.77.94
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Although many technologies, such as an electronic portal imaging device, a software-based quality assurance system, an ion chamber array,etc., have been developed for the purpose of quality assurance of radiotherapy planning, radiochromic film is still one of the most reliable technologies and is being widely used. In particular, compared with other technologies, EBT3 film has unique advantages, including its unlimited spatial resolution, relatively small volume and waterproof quality, which means it can be used inside the patient's body. The main obstacle to using EBT3 film is that various equations have to be used in the calibration process, which makes it very sensitive to the results. The manufacturer provides a fitting function, but many software programs and researchers use their own functions. In this study, some functions normally used for calibrating EBT3 films are compared by analyzing a couple of characteristics in the course of fitting the function to the measured values. Although our study shows the lack of a suitable fitting function that works well in every situation, it can provide valuable insight into selecting suitable functions for fitting in a clinical setting.
引用
收藏
页码:94 / 101
页数:8
相关论文
共 50 条
  • [11] Considerations in selecting protease inhibitor therapy
    Boyle, BA
    Elion, RA
    Moyle, GJ
    Cohen, CJ
    AIDS REVIEWS, 2004, 6 (04) : 218 - 225
  • [12] DOSIMETRY AND COGNITIVE FUNCTION AFTER RADIATION THERAPY IN LOW GRADE GLIOMAS
    Asha, Wafa
    Koro, Sami
    Asha, Wafa
    Wei, Wei
    Chen, Yanwen
    Murray, Eric
    Murphy, Erin
    Suh, John
    Chao, Samuel
    Angelov, Lilyana
    Peereboom, David
    NEURO-ONCOLOGY, 2022, 24 : 205 - 205
  • [13] Film dosimetry studies for patient specific quality assurance in microbeam radiation therapy
    Ocadiz, Alexandre
    Livingstone, Jayde
    Donzelli, Mattia
    Bartzsch, Stefan
    Nemoz, Christian
    Kefs, Samy
    Pellicioli, Paolo
    Giraud, Jean-Yves
    Balosso, Jacques
    Krisch, Michael
    Brauer-Krisch, Elke
    Serduc, Raphael
    Adam, Jean-Francois
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 65 : 227 - 237
  • [14] Micrometer-resolved film dosimetry using a microscope in microbeam radiation therapy
    Bartzsch, Stefan
    Lott, Johanna
    Welsch, Katrin
    Braeuer-Krisch, Elke
    Oelfke, Uwe
    MEDICAL PHYSICS, 2015, 42 (07) : 4069 - 4079
  • [15] SIMPLIFIED FILM PROCESSING FOR RADIATION DOSIMETRY
    MAGILL, RJ
    NUCLEONICS, 1954, 12 (08): : 43 - 44
  • [16] DYE FILM DOSIMETRY FOR RADIATION PROCESSING
    HUMPHREYS, JC
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1980, 25 (07): : 766 - 766
  • [17] Radiochromic film for medical radiation dosimetry
    Butson, MJ
    Yu, PKN
    Cheung, T
    Metcalfe, P
    MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5): : 61 - 120
  • [18] Ultraviolet radiation dosimetry with radiochromic film
    Butson, MJ
    Cheung, T
    Yu, PKN
    Abbati, D
    Greenoak, GE
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (07): : 1863 - 1868
  • [19] POLAROID FILM IN RADIATION DOSIMETRY AND SURVEY
    LEVAN, H
    STEFANI, S
    HAGE, SJ
    RADIOLOGY, 1971, 98 (01) : 182 - &
  • [20] A NEW FILM PACKET FOR RADIATION DOSIMETRY
    HERMAN, WC
    NUCLEONICS, 1951, 9 (04): : 74 - 76