Investigation of the "block effect" on spatial resolution in PET detectors

被引:19
作者
Tomic, N [1 ]
Thompson, CJ
Casey, ME
机构
[1] McGill Univ, Jewish Gen Hosp, Dept Radiat Oncol, Montreal, PQ H3T 1E2, Canada
[2] McGill Univ, Dept Phys Med, Montreal, PQ H3T 1E2, Canada
[3] CTI PET Syst Inc, Knoxville, TN 37932 USA
基金
加拿大自然科学与工程研究理事会;
关键词
BGO; block detectors; positron emission tomography (PET); spatial resolution;
D O I
10.1109/TNS.2005.851433
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Previous studies suggest that when block detectors are used in positron emission tomography (PET) scanners, a "block effect" of about 2 mm should be added in quadrature to the other effects blurring spatial resolution (SR). However, no explanation of the cause of the effect has ever been published. If the effect could be identified, isolated, and reduced then the SR of PET scanners should improve. We studied the origin of the "block effect," using experiments in which all other blurring effects on SR were minimized and precisely determined. Thin, 1 mm wide bismuth germanate (BGO) crystals and a small (1 mm) (68)Ge source were used to probe the SR of a HR+ block detector. Two precise translation stages were used to move detectors. We compared coincidence aperture functions (CAFs) for crystals in the block with CAFs of single crystals of various widths. From those measurements, done with one block detector, we concluded that the central crystals in the block showed an additional blurring of 0.8 mm whereas the edge ones showed no additional blurring. When the detectors were separated by 21 cm, the full width at half maximum (FWHM) for central crystals was found to be 2.20 +/- 0.04 mm and for edge crystals 1.98 +/- 0.04 mm. The FWHM for 4.4 mm width crystals (as used in the HR + block) was estimated to be 2.1 +/- 0.2 mm. Results from the crystal identification matrix show that the apparent centroids of the crystals are not located at the geometric centers, which would give errors in the reconstruction algorithm assumed uniform sampling. Our results suggest that the additional blurring previously reported in PET scanners with block detectors is not entirely due to the use of block detectors.
引用
收藏
页码:599 / 605
页数:7
相关论文
共 50 条
  • [41] Improving depth-of-interaction resolution in pixellated PET detectors using neural networks
    Zatcepin, Artem
    Pizzichemi, Marco
    Polesel, Andrea
    Paganoni, Marco
    Auffray, Etiennette
    Ziegler, Sibylle, I
    Omidvari, Negar
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17)
  • [42] Spatial resolution of double-sided silicon microstrip detectors for the PAMELA apparatus
    Straulino, S
    Adriani, O
    Bonechi, L
    Bongi, M
    Bottai, S
    Castellini, G
    Fedele, D
    Grandi, M
    Papini, R
    Ricciarini, SB
    Spillantini, P
    Taccetti, E
    Taddei, E
    Vannuccini, E
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 556 (01) : 100 - 114
  • [43] Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors
    Krishnamoorthy, Srilalan
    Blankemeyer, Eric
    Mollet, Pieter
    Surti, Suleman
    Van Holen, Roel
    Karp, Joel S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (15)
  • [44] Timing, Energy, and Spatial Characterization of Highly Sampled Monolithic PET Detectors With Different Thicknesses
    Valladares, Celia
    Barrio, John
    Freire, Marta
    Cucarella, Neus
    Lamprou, Efthymios
    Miyaoka, Robert S.
    Hunter, William C. J.
    Harrison, Robert
    Gonzalez, Antonio J.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2023, 7 (07) : 704 - 711
  • [45] Investigation of the spatial resolution of an online dose verification device
    Asuni, G.
    Rickey, D. W.
    McCurdy, B. M. C.
    MEDICAL PHYSICS, 2012, 39 (02) : 697 - 705
  • [46] The effect of attentional spread on spatial resolution
    Balz, GW
    Hock, HS
    VISION RESEARCH, 1997, 37 (11) : 1499 - 1510
  • [47] Estimation of the Spatial Resolution in Acoustic Investigation of Ground Layer
    Zaslavsky, Yu. M.
    Zaslavsky, V. Yu.
    JOURNAL OF APPLIED MECHANICS AND TECHNICAL PHYSICS, 2022, 63 (07) : 1091 - 1100
  • [48] Spatial resolution and depth-of-interaction studies with a PET detector module composed of LSO and an APD array
    Fremout, AAR
    Chen, RR
    Bruyndonckx, P
    Tavernier, SPK
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) : 131 - 138
  • [49] Spatial Resolution and Sensitivity of the Inveon Small-Animal PET Scanner
    Visser, Eric P.
    Disselhorst, Jonathan A.
    Brom, Maarten
    Laverman, Peter
    Gotthardt, Martin
    Oyen, Wim J. G.
    Boerman, Otto C.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (01) : 139 - 147
  • [50] Clinical evaluation of PET image reconstruction using a spatial resolution model
    Andersen, Flemming Littrup
    Klausen, Thomas Levin
    Loft, Annika
    Beyer, Thomas
    Holm, Soren
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (05) : 862 - 869