Investigation of the "block effect" on spatial resolution in PET detectors

被引:19
作者
Tomic, N [1 ]
Thompson, CJ
Casey, ME
机构
[1] McGill Univ, Jewish Gen Hosp, Dept Radiat Oncol, Montreal, PQ H3T 1E2, Canada
[2] McGill Univ, Dept Phys Med, Montreal, PQ H3T 1E2, Canada
[3] CTI PET Syst Inc, Knoxville, TN 37932 USA
基金
加拿大自然科学与工程研究理事会;
关键词
BGO; block detectors; positron emission tomography (PET); spatial resolution;
D O I
10.1109/TNS.2005.851433
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Previous studies suggest that when block detectors are used in positron emission tomography (PET) scanners, a "block effect" of about 2 mm should be added in quadrature to the other effects blurring spatial resolution (SR). However, no explanation of the cause of the effect has ever been published. If the effect could be identified, isolated, and reduced then the SR of PET scanners should improve. We studied the origin of the "block effect," using experiments in which all other blurring effects on SR were minimized and precisely determined. Thin, 1 mm wide bismuth germanate (BGO) crystals and a small (1 mm) (68)Ge source were used to probe the SR of a HR+ block detector. Two precise translation stages were used to move detectors. We compared coincidence aperture functions (CAFs) for crystals in the block with CAFs of single crystals of various widths. From those measurements, done with one block detector, we concluded that the central crystals in the block showed an additional blurring of 0.8 mm whereas the edge ones showed no additional blurring. When the detectors were separated by 21 cm, the full width at half maximum (FWHM) for central crystals was found to be 2.20 +/- 0.04 mm and for edge crystals 1.98 +/- 0.04 mm. The FWHM for 4.4 mm width crystals (as used in the HR + block) was estimated to be 2.1 +/- 0.2 mm. Results from the crystal identification matrix show that the apparent centroids of the crystals are not located at the geometric centers, which would give errors in the reconstruction algorithm assumed uniform sampling. Our results suggest that the additional blurring previously reported in PET scanners with block detectors is not entirely due to the use of block detectors.
引用
收藏
页码:599 / 605
页数:7
相关论文
共 50 条
  • [31] On the assessment of spatial resolution of PET systems with iterative image reconstruction
    Gong, Kuang
    Cherry, Simon R.
    Qi, Jinyi
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (05) : N193 - N202
  • [32] From PET detectors to PET scanners
    John L. Humm
    Anatoly Rosenfeld
    Alberto Del Guerra
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 1574 - 1597
  • [33] Correction of myocardial artifact due to limited spatial resolution in pet
    Zhang, H
    Zhang, ZX
    Yang, L
    Ma, XR
    Proceedings of 2005 International Conference on Machine Learning and Cybernetics, Vols 1-9, 2005, : 5060 - 5064
  • [34] From PET detectors to PET scanners
    Humm, JL
    Rosenfeld, A
    Del Guerra, A
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (11) : 1574 - 1597
  • [35] Design and development of 1 mm resolution PET detectors with position-sensitive PMTs
    Shao, YP
    Cherry, SR
    Chatziioannou, AF
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 477 (1-3) : 486 - 490
  • [36] Detector motion method to increase spatial resolution in photon-counting detectors
    Daehee Lee
    Kyeongjin Park
    Kyung Taek Lim
    Gyuseong Cho
    Journal of the Korean Physical Society, 2017, 70 : 567 - 573
  • [37] Scintillator-photodiode detectors for scanning introscopy with high-spatial resolution
    Ryzhikov, V
    Kozin, D
    Grinyov, B
    Lysetskaya, E
    Katrunov, K
    Kvyatnitskaya, V
    Chernikov, V
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 505 (1-2) : 58 - 62
  • [38] Spatial Resolution Enhancement in Integration-Mode Detectors for Proton Radiography and Tomography
    Gianoli, C.
    Meyer, S.
    Magallanes, L.
    Dedes, G.
    Landry, G.
    Nijhuis, R.
    Ganswindt, U.
    Thieke, C.
    Belka, C.
    Parodi, K.
    RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S46 - S47
  • [39] Detector Motion Method to Increase Spatial Resolution in Photon-Counting Detectors
    Lee, Daehee
    Park, Kyeongjin
    Lim, Kyung Taek
    Cho, Gyuseong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 70 (06) : 567 - 573
  • [40] Use of YAP(Ce) in the development of high spatial resolution radiation imaging detectors
    Yamamoto, Seiichi
    Kamada, Kei
    Yoshikawa, Akira
    RADIATION MEASUREMENTS, 2018, 119 : 184 - 191