Reliability of peak VO2 and maximal cardiac output assessed using thoracic bioimpedance in children

被引:0
作者
Welsman, J [1 ]
Bywater, K [1 ]
Farr, C [1 ]
Welford, D [1 ]
Armstrong, N [1 ]
机构
[1] Univ Exeter, Childrens Hlth & Exercise Res Ctr, Exeter EX1 2LU, Devon, England
关键词
stroke volume; exercise; PhysioFlow;
D O I
10.1007/s00421-004-1300-5
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The purpose of this study was to evaluate the reliability of a thoracic electrical bioimpedance based device (PhysioFlow) for the determination of cardiac output and stroke volume during exercise at peak oxygen uptake (peak VO2 in children. The reliability of peak VO2 is also reported. Eleven boys and nine girls aged 10-11 years completed a cycle ergometer test to voluntary exhaustion on three occasions each 1 week apart. Peak VO2 was determined and cardiac output and stroke volume at peak VO2 were measured using a thoracic bioelectrical impedance device (PhysioFlow). The reliability of peak VO2, cardiac output and stroke volume were determined initially from pairwise comparisons and subsequently across all three trials analysed together through calculation of typical error and intraclass correlation. The pairwise comparisons revealed no consistent bias across tests for all three measures and there was no evidence of non-uniform errors (heteroscedasticity). When three trials were analysed together typical error expressed as a coefficient of variation was 4.1% for peak VO2, 9.3% for cardiac output and 9.3% for stroke volume. Results analysed by sex revealed no consistent differences. The PhysioFlow method allows non-invasive, beat-to-beat determination of cardiac output and stroke volume which is feasible for measurements during maximal exercise in children. The reliability of the PhysioFlow falls between that demonstrated for Doppler echocardiography (5%) and CO2 rebreathing (12%) at maximal exercise but combines the significant advantages of portability, lower expense and requires less technical expertise to obtain reliable results.
引用
收藏
页码:228 / 234
页数:7
相关论文
共 37 条
[1]  
Armstrong N, 1994, Exerc Sport Sci Rev, V22, P435
[2]  
ARMSTRONG N, 2000, PAEDIAT EXERCISE SCI, P173
[3]   Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine [J].
Atkinson, G ;
Nevill, AM .
SPORTS MEDICINE, 1998, 26 (04) :217-238
[4]   Typical error versus limits of agreement [J].
Atkinson, G ;
Nevill, A .
SPORTS MEDICINE, 2000, 30 (05) :375-377
[5]  
Berstein DP, 1986, CRIT CARE MED, V14, P904
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]   A new impedance cardiograph device for the non-invasive evaluation of cardiac output at rest and during exercise: comparison with the "direct" Fick method [J].
Charloux, A ;
Lonsdorfer-Wolf, E ;
Richard, R ;
Lampert, E ;
Oswald-Mammosser, M ;
Mettauer, B ;
Geny, B ;
Lonsdorfer, J .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2000, 82 (04) :313-320
[8]  
Driscoll David J, 1989, Pediatr Exerc Sci, V1, P102, DOI 10.1123/pes.1.2.102
[9]   COMPARISON OF CARDIAC-OUTPUT MEASUREMENT TECHNIQUES - THERMODILUTION, DOPPLER, CO2-REBREATHING AND THE DIRECT FICK METHOD [J].
ESPERSEN, K ;
JENSEN, EW ;
ROSENBORG, D ;
THOMSEN, JK ;
ELIASEN, K ;
OLSEN, NV ;
KANSTRUP, IL .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 (02) :245-251
[10]   COMPARISON OF CARDIAC OUTPUT DETERMINED BY CO2 REBREATHING AND DYE-DILUTION METHODS [J].
FERGUSON, RJ ;
FAULKNER, JA ;
JULIUS, S ;
CONWAY, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1968, 25 (04) :450-+