Wavemeter improvements for laser diode calibration

被引:0
|
作者
Outumuro, I. [1 ]
Diz-Bugarin, J. [2 ]
Valencia, J. L. [1 ]
Blanco, J. [2 ]
Dorrio, B. V. [2 ]
机构
[1] Parque Tecnolox Galicia, LOMG, Orense 32901, Spain
[2] Univ Vigo, Appl Phys Dept, Campus Univ, Vigo 36310, Spain
来源
THIRD INTERNATIONAL CONFERENCE ON APPLICATIONS OF OPTICS AND PHOTONICS | 2017年 / 10453卷
关键词
wavemeter; Michelson interferometer; ECDL; metrology; calibration; UPDATED EDLEN EQUATION; WAVELENGTH METER; REFRACTIVE-INDEX;
D O I
10.1117/12.2272010
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
This paper shows the progress made in the wavemeter developed to give traceability to the wavelength of lasers and ECDLs (External Cavity Laser Diode). The improvements are: duplication of the optical path of the laser beams due to a double pass through the interferometer arms [1], the electronic fringe counter [2], the measurement of the refractive index of air and the uncertainty calculations of the wavelength for the case of lasers with frequencies that differs more than 10 THz from laser reference. The new measurements improve the previous results [3].
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Beam-combining improvements optimize laser-diode pump sources
    Wilson, D.
    Duesterberg, R.
    Xu, L.
    Skidmore, J.
    LASER FOCUS WORLD, 2011, 47 (08): : 45 - +
  • [42] LASER WAVELENGTH MEASUREMENT WITH A FOURIER-TRANSFORM WAVEMETER
    JUNTTILA, ML
    STAHLBERG, B
    APPLIED OPTICS, 1990, 29 (24): : 3510 - 3516
  • [43] Calibration of an InGaAs photodiode at 1300 nm with a cryogenic radiometer and a diode laser
    Atkinson, E.G.
    Butler, D.J.
    Metrologia, 35 (04): : 241 - 245
  • [44] The development of the Icarus T600 laser diode calibration system
    Bonesini, M.
    Benocci, R.
    Bertoni, R.
    Cervi, T.
    Mazza, R.
    Menegolli, A.
    Raselli, G.
    Rossella, M.
    Torti, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 936 : 261 - 262
  • [45] Calibration of single-mode content of semiconductor diode laser emission
    Dinklage, A
    Solyman, S
    Wilke, C
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (10): : 3832 - 3833
  • [46] Calibration of an InGaAs photodiode at 1300 nm with a cryogenic radiometer and a diode laser
    Atkinson, EG
    Butler, DJ
    METROLOGIA, 1998, 35 (04) : 241 - 245
  • [47] INTERMODE CALIBRATION OF DIODE-LASER SPECTRA USING TANDEM ETALONS
    JENNINGS, DE
    APPLIED OPTICS, 1980, 19 (01) : 2 - 4
  • [48] FIZEAU WAVEMETER FOR PULSED LASER WAVELENGTH MEASUREMENT.
    Morris, Mark B.
    McIlrath, Thomas J.
    Snyder, James J.
    Applied Optics, 1983, 23 (21): : 3862 - 3868
  • [49] IMPROVEMENTS TO DIODE ION PUMPS
    PONTVIAN.J
    VACUUM, 1968, 18 (03) : 137 - &
  • [50] Compact diffraction grating laser wavemeter for cold atom experiments
    魏春华
    颜树华
    张甜
    OptoelectronicsLetters, 2017, 13 (05) : 335 - 338