Evaluation of dry powder inhalers with a focus on ease of use and user preference in inhaler-naive individuals

被引:14
|
作者
von Schantz, Sofia [1 ]
Katajavuori, Nina [1 ]
Antikainen, Osmo [1 ]
Juppo, Anne [1 ]
机构
[1] Univ Helsinki, Fac Pharm, POB 56, FIN-00014 Helsinki, Finland
关键词
Dry powder inhaler; Preference; Ease of use; Inhalation errors; INHALATION TECHNIQUE; ASTHMA; MANAGEMENT; BURDEN;
D O I
10.1016/j.ijpharm.2016.05.023
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Inhaler errors are common amongst inhaler users. Therefore, in the development work of new inhalation devices, it is important to characterize the ease of use of the inhalers. In this study four dry powder inhalers, Diskus, Easyhaler, Ellipta and Turbuhaler, were evaluated, focusing on ease of use and patient preference. The study used a triangular methodology. The sample consisted of 31 inhaler naive individuals. Educational videos for all inhalers were watched, and afterwards, the use of all four inhalers was demonstrated in a random order. The demonstrations were videotaped. Thereafter they were checked against a predefined checklist and all mistakes were recorded. Only 33% of inhaler demonstrations were completed without the participants making any mistakes. The proportions of subjects who used the devices correctly were as follows: Diskus 48%, Easyhaler 19%, Ellipta 55% and Turbuhaler 16%. When comparing correct and incorrect inhaler technique for each inhaler pair the following differences were statistically significant: Diskus vs. Easyhaler (p < 0.05), Ellipta vs. Easyhaler (p < 0.01), Diskus vs. Turbuhaler (p < 0.01), Ellipta vs. Turbuhaler (p < 0.01). In the participants' ranking, the inhalers Ellipta, followed by Turbuhaler, were most often ranked as most preferred. Participants' preference of Ellipta over Easyhaler (p < 0.01) and over Diskus (p < 0.001) were statistically significant. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:50 / 58
页数:9
相关论文
共 31 条
  • [21] Correct use and ease-of-use of placebo ELLIPTA dry-powder inhaler in adult patients with chronic obstructive pulmonary disease
    Siler, Thomas
    Jain, Renu
    Collison, Kathryn
    Sharma, Raj I.
    Sutton, Laura
    Rees, Jamie
    Bernstein, David
    PLOS ONE, 2022, 17 (08):
  • [22] Evaluation of a novel salbutamol dry powder inhaler: Clinical equivalence to a standard metered dose inhaler and in-use stability
    Morice, AH
    Peake, MD
    Allen, MB
    Campbell, JH
    ParryBillings, M
    BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1996, 42 (05) : P658 - P658
  • [23] Prevalence of serious post-training inhaler technique errors made by device-naive patients using three different dry powder inhalers (DPIs)
    Chrystyn, Henry
    Bosnic-Anticevich, Sinthia
    Roche, Nicolas
    Molimard, Mathieu
    Haughney, John
    Lavorini, Federico
    Shan, Dawn
    Sims, Erika
    Burden, Anne
    Ashton, Valerie L.
    Efthimiou, John
    Ferkl, Marek
    Price, David B.
    EUROPEAN RESPIRATORY JOURNAL, 2015, 46
  • [24] Concurrent use of metered-dose and dry powder inhalers by children with persistent asthma does not adversely affect spacer/inhaler technique
    Chan, Debora S.
    Callahan, Charles W.
    Hatch-Pigott, Virginia B.
    Lawless, Annette
    Proffitt, H. Lorraine
    Manning, Nola E.
    Schweikert, Mary P.
    ANNALS OF PHARMACOTHERAPY, 2006, 40 (10) : 1743 - 1746
  • [25] Inhalation devices for long-acting β2-agonists:: efficiency and ease of use of dry powder formoterol inhalers for use by patients with asthma and COPD
    Molimard, Mathieu
    Till, Denise
    Stenglein, Stephan
    Singh, Dilraj
    Krummen, Markus
    CURRENT MEDICAL RESEARCH AND OPINION, 2007, 23 (10) : 2405 - 2413
  • [26] Evaluation of inhaler technique, adherence to therapy and their effect on disease control among children with asthma using metered dose or dry powder inhalers
    Capanoglu, Murat
    Misirlioglu, Emine Dibek
    Toyran, Muge
    Civelek, Ersoy
    Kocabas, Can Naci
    JOURNAL OF ASTHMA, 2015, 52 (08) : 838 - 845
  • [27] Inhalation characteristics and their effects on in vitro drug delivery from dry powder inhalers .1. Inhalation characteristics, work of breathing and volunteers' preference in dependence of the inhaler resistance
    deBoer, AH
    Winter, HMI
    Lerk, CF
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1996, 130 (02) : 231 - 244
  • [28] Ease of use of a two-strip dry powder inhaler (DPI) to deliver fluticasone furoate/vilanterol (FF/VI) and FF alone in asthma
    Svedsater, Henrik
    Jacques, Loretta
    Goldfrad, Caroline
    Bleecker, Eugene R.
    O'Byrne, Paul M.
    Woodcock, Ashley
    EUROPEAN RESPIRATORY JOURNAL, 2013, 42
  • [29] Comparison of serious inhaler technique errors made by device-naive patients using three different dry powder inhalers: a randomised, crossover, open-label study
    Chrystyn, Henry
    Price, David B.
    Molimard, Mathieu
    Haughney, John
    Bosnic-Anticevich, Sinthia
    Lavorini, Federico
    Efthimiou, John
    Shan, Dawn
    Sims, Erika
    Burden, Anne
    Hutton, Catherine
    Roche, Nicolas
    BMC PULMONARY MEDICINE, 2016, 16
  • [30] ASSESSMENT OF NAIVE INHALER USER INHALATION PROFILES AND THE IMPACT ON SUBSEQUENT AEROSOL PERFORMANCE : COMPARISON OF A DRY POWDER INHALER (DPI) AND A PRESSURIZED METERED DOSE INHALER (PMDI) WITH VALVED HOLDING CHAMBER (VHC) USING THE SAME ACTIVE PHARMACEUTICAL INGREDIENTS (APIS)
    Suggett, Jason A.
    Nagel, Mark W.
    Mitchell, Jolyon P.
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2020, 33 (02) : A3 - A3